Protect and Extend Column Lifetime
- Individually QC tested to ensure the highest level of quality
- Extend column lifetime by preventing stationary phase damage
- Improve separation and peak shapes (especially early elutors)
- Improve sensitivity and accuracy of quantitative results
- Available as individual guard columns or as complete kits with connectors
To ensure that all Z-Guards are the highest possible quality, we individually test each one! The columns are attached to a reference Zebron ZB-5 column and are tested using our specially designed QC mix. We carefully monitor activity, bleed, and stability. This way, we are able to say with confidence that Z-Guards will provide the low activity and high quality your methods require.
Recommended Use
Extends column lifetime by preventing stationary phase damage and improves separation, peak shapes (especially early elutors), sensitivity, and accuracy of quantitative results
Zebron Z-Guard Phase Information
Z-Guard columns are universal GC guard columns, designed for use with virtually any GC capillary column, from virtually any manufacturer.
- Extend column lifetime by preventing stationary phase damage
- Improve separation, peak shapes (especially early elutors), sensitivity, and accuracy of quantitative results
- Alternative to Agilent, Restek, Supelco, and many more
Easy Use Z-Guard Column Kits
Try Z-Guard columns individually or get the convenient, complete kit. Each kit includes:
- 5 or 10 meters of deactivated fused-silica tubing
- 5 universal connectors
- 0.5 mL of high-temperature polyimide resin
Guard Column Usage
Phenomenex: Zebron Z-Guard GC Columns: Guard column usage
Tips: Using Z-Guard Columns
When using a guard at high temperatures, make sure that the press-fit connection is secure.
Phenomenex: Zebron Z-Guard GC Columns: Using Z-guard columns
The most important aspect of a good seal is a good column cut.
- The column end should be square and smooth (like a straw). Any irregularities will cause leaks.
- Columns are long - repeat cutting until a good cut is achieved.
- Lightly scoring the column usually creates a cleaner opening.