ACQUITY UPLC M-Class System with HDX Technology

Others | 2016 | WatersInstrumentation
HPLC
Industries
Manufacturer
Waters

Importance of the Topic


The ACQUITY UPLC M-Class System with HDX Technology addresses critical challenges in measuring protein higher-order structure and dynamics. By combining low-temperature UPLC separations, high-resolution mass spectrometry and automated fluidics, it transforms HDX-MS into a robust, routine tool for biopharmaceutical research, quality control and structural biology.

Objectives and Overview of the Article


This whitepaper introduces an integrated HDX-MS platform that unites sample preparation, on-line digestion, chromatographic separation and informatics. It aims to demonstrate how the system boosts chromatographic efficiency at 0 °C, automates complex HDX workflows and accelerates data interpretation with dedicated software.

Methodology and Instrumentation


The platform integrates several modules:
  • UPLC M-Class HDX Manager with temperature-controlled injection and trapping valves at 0 °C, plus a pepsin digestion chamber under regulated back pressure.
  • Auxiliary Solvent Manager (ASM) and µBinary Solvent Manager (µBSM) for quench, digestion and peptide trapping flows.
  • Enzymate BEH Pepsin Column for rapid on-line digestion (<30 s) at user-defined pressure up to 15 000 psi.
  • Waters high-resolution MS with Triwave IMS for precise mass measurement and an orthogonal separation dimension.
  • LEAP HDX-2 Automation Manager with Chronos scheduling to automate labeling, quenching and injections for local or global analyses.
  • DynamX HDX Data Analysis Software for interactive uptake curves, spectral overlays, heat maps, ETD support and PyMOL export.

Main Results and Discussion


The system achieves sub-2 µm UPLC separations at 0 °C with cycle times under 10 min when coupled to IMS, delivering superior peak capacity under quench conditions. Automated fluidics yield highly reproducible peptide maps and deuterium uptake data. The DynamX software reduces days of manual analysis to minutes, enabling confident detection of conformational changes, epitope mapping and ligand interactions.

Benefits and Practical Applications


  • High-throughput screening of protein conformation, folding kinetics and stability for drug development and QA/QC.
  • Detailed mapping of membrane receptor interfaces, biosimilar comparability and epitope identification.
  • Regulatory-grade reproducibility and streamlined data management for analytical laboratories.

Future Trends and Potential Applications


As HDX-MS moves toward analysis of megaDalton complexes and dynamic assemblies, coupling IMS and machine learning will unlock deeper structural insights. Expanded automation, multiplexed workflows and integrated modeling tools will drive faster, more detailed conformational profiling across biotechnology and life-science research.

Conclusion


The ACQUITY UPLC M-Class System with HDX Technology overcomes traditional hurdles in HDX-MS by integrating precise low-temperature separations, automated fluidics and intuitive informatics. It delivers rapid, reproducible and high-confidence protein conformation data, supporting diverse applications from biopharma development to structural biology.

Reference


  1. Hvidt C. and Linderstrøm-Lang K., Biochim. Biophys. Acta 14, 574 (1954).
  2. Jacob R.E. and Engen J.R., J. Am. Soc. Mass Spectrom. 23, 1003–1010 (2012).
  3. Wu Y., Engen J.R. and Hobbins W.B., J. Am. Soc. Mass Spectrom. 17, 163–167 (2006).

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Pressurized Online Pepsin Digestion of mAb IgG2 for  Hydrogen Deuterium Exchange Mass Spectrometry
Enhanced Performance of the SYNAPT XS and Its Impact on Hydrogen Deuterium Exchange Mass Spectrometry (HDX MS) Data Quality
Conformational Characterization of Calmodulin by Hydrogen Deuterium Exchange Mass Spectrometry
LOCALIZED CONFORMATION ANALYSES OF MUTATED HUMAN IgG1  BY HYDROGEN DEUTERIUM EXCHANGE MASS SPECTROMETRY AND DIFFERENTIAL SCANNING CALORIMETRY