This Solvent Recycler saves up to 90% of mobile phase by redirection of the pure solvent to the solvent reservoir during the isocratic HPLC. The stand alone device is powered by 5 V DC from power adapter or from computer USB port. Working parameters (Threshold and Delay) are easily set on the display using a membrane keyboard.
Watrex: Watrex SPR-200 Solvent saver - INputs
The SPR-220™ Solvent Recycler continuously monitors the output signal of the chromatography detector and displays its current value on the built-in display. A highresolution analog-to-digital converter provides data for further evaluation in the built-in processor. If the input signal level exceeds a preset value (Threshold), the SPR-200 will redirect the flow to the waste.
When the signal drops, the solvent recycler waits for the transport delay and then switches the mobile phase back to the reservoir. An autosampler injection marker connected to the SPR-200 ZERO input can be used to zero the input signal at the time of injection.
FEATURES
- Stand alone operation
- Detachable valve
- Possibility of manual Waste/Recycle switching
TECHNICAL SPECIFICATION
- Detector compatibility: compatible with any HPLC detector
- Connection: 1/16″ O.D. tubing, 1/4″-28 flat bottom fittings
- Maximum pressure: 30 p.s.i./0,2 MPa
- Wetted material: PEEK, PTFE
- Input range: +/- 1 V or +/- 100 mV, user selectable
- Computer connection: μUSB
- TTL/contact closure input: configurable (ZERO, start, valve position)
- Power: 5V DC/<100 mA (power supply or computer USB port)
- User interface: large character LCD alphanumeric display + membrane keypad
Ordering information:
PN/Component:
- SPR200-001 SPR-200 Solvent Recycler, PEEK version (Price 1 250 € ex VAT)
- SPR200-002 SPR-200 Solvent Recycler, PTFE version (Price 1 250 € ex VAT)
You can also visit this product on the Watrex website.