Analysis of Halogenated Flame Retardants Using Atmospheric Pressure Chemical Ionization GC-MS/MS

Applications | 2026 | WatersInstrumentation
GC/MSD, GC/MS/MS, GC/API/MS, GC/QQQ
Industries
Energy & Chemicals
Manufacturer
Waters

Significance of the topic


Monitoring halogenated flame retardants (HFRs), including legacy polybrominated diphenyl ethers (PBDEs) and a diverse set of non-PBDE HFRs, remains essential because of their historical high-volume use, environmental persistence, long-range transport and potential human health impacts. Robust, sensitive and sustainable analytical workflows are needed to support environmental monitoring, human biomonitoring and regulatory compliance while reducing operational cost, dependence on scarce gases and the expertise barrier associated with legacy high-resolution magnetic sector instruments.

Objectives and overview of the study


This study evaluated modernization of a helium (He)-based electron ionization (EI) magnetic sector GC-HRMS reference method for 21 HFRs by converting it to a nitrogen (N2) carrier gas workflow using atmospheric pressure chemical ionization gas chromatography (GC-APCI, APGC) coupled to a tandem quadrupole mass spectrometer (Xevo TQ Absolute). Goals were to (1) maintain or improve chromatographic separation and quantification performance, (2) demonstrate compatibility with existing column dimensions (no column rescaling), (3) enable multiple reaction monitoring (MRM) with many transitions for confirmatory power across matrices, and (4) reduce reliance on helium and on operator training time.

Methodology


Key experimental design elements and method adaptations:
  • Analytes: 21 halogenated flame retardants (HFRs), including highly brominated congeners; standards prepared at Environment and Climate Change Canada (ECCC) using a five-point matrix-matched calibration (1–250 ppb).
  • Chromatography: Rtx-1614 column (15 m × 0.25 mm ID × 0.10 µm film). Initial oven program started at 110 °C with multiple ramps culminating in a 5 min final hold, providing a total runtime of 24.5 minutes.
  • Injection: 1 µL pulsed splitless injections on an SSL injector at 260 °C with a 0.5 minute pressure pulse to 20 psi. Pressure pulsing compensates for the lower optimum linear velocity of N2 relative to He to ensure efficient transfer of analytes into the column.
  • Carrier gas & flow strategy: All-nitrogen configuration; initial N2 flow 0.80 mL/min ramped to 4.1 mL/min (0.4 mL/min/min) and held to the end of the run. Flow ramping was essential to improve peak widths when using the same column dimensions as the He method.
  • Ionization and detection: APGC (positive ion, dry charge exchange source) with Xevo TQ Absolute tandem quadrupole MS operated in MRM mode. Four to eight MRM transitions per analyte were optimized and used; summed transitions used for quantification.

Instrumentation


Instruments and consumables used in the study:
  • Mass spectrometer: Xevo TQ Absolute (Waters) with APGC source.
  • Gas chromatograph: Agilent 8890 GC coupled to the MS.
  • Column: Restek Rtx-1614, 15 m × 0.25 mm ID × 0.10 µm film.
  • Injection system: split/splitless (SSL) injector with pulsed pressure capability; conditions adapted to avoid thermal degradation of labile HFRs.

Main results and discussion


Performance highlights and method observations:
  • Run time and separation: Total runtime was 24.5 minutes (N2/APGC-TQMS) vs 28 minutes for the He/EI HRMS reference method. The last-eluting analyte, decabromodiphenylethane (DBDPE), eluted at 20.8 minutes with symmetric peak shape (asymmetry a/b = 1.03) and a peak width of ~9.7 s.
  • Chromatographic adaptation: Direct substitution of N2 at a single fixed flow produced broad peaks; implementing a carrier flow ramp and modifying oven hold times restored acceptable peak shapes and maintained separation of critical near-eluting pairs without changing the column dimensions.
  • Comparison to reference: Separation of near-eluting pair DPTE and HBB was equivalent between N2/APGC-MRM on TQ and He/EI HRMS; small differences across other critical pairs were observed, with some pairs slightly better in one method or the other but overall mass discrimination and MRM specificity mitigated potential coelution concerns.
  • Sensitivity and quantitation: Matrix-matched calibrations (1–250 ppb) yielded linear responses for all 21 analytes with an average coefficient of determination r2 = 0.9951. Summed MRM transitions improved quantitation robustness and provided multiple confirmatory traces and ion ratios.
  • MRM strategy advantages: Using 4–8 transitions per analyte (vs typical 2) permits (i) summation for improved signal-to-noise, (ii) multiple confirmation ratios for higher specificity, and (iii) flexibility to select alternative transitions when matrix or sample preparation interferences affect primary traces.
  • Injection and thermal stability: Optimization of injector temperature (250–300 °C) and splitless duration (0.25–1.5 min) ensured complete transfer of high-boiling DBDPE and minimized thermal decomposition of labile isomers such as TBECH. The reference method’s cool-on-column or PTV approaches were noted as alternative strategies for fragile analytes.

Benefits and practical applications


Practical advantages demonstrated by the N2/APGC-TQMS method:
  • Reduced dependence on helium: Using N2 as carrier, collision and source gas reduces vulnerability to He supply disruptions, lowers recurring gas costs and supports sustainability goals.
  • Operational efficiency: Tandem quadrupole MS offers higher acquisition speed and simpler operation compared with magnetic sector HRMS, shortening operator training from months to weeks and increasing throughput.
  • Method versatility: The APGC source tolerates higher carrier flows and many concurrent MRM transitions, enabling use of the same column and comparable separations across matrices and sample prep methods without extensive method revalidation.
  • Confirmatory power: Multiple MRM transitions and summed traces enhance specificity for regulatory and environmental monitoring contexts where confidence in identification and quantitation is essential.

Future trends and applications


Potential directions and broader applicability:
  • Method consolidation: Combining PBDEs and diverse HFRs into a single, robust N2/APGC-TQMS method could streamline monitoring campaigns and laboratory workflows.
  • Expanded analyte lists: The MRM database development approach supports extension to additional halogenated pollutants and emerging flame retardants without substantial hardware changes.
  • Sustainability and resilience: Wider adoption of all-nitrogen GC-MS workflows supports greener laboratory practice and reduces supply-chain vulnerability for critical gases.
  • Matrix challenges and automation: Future work should evaluate a broader set of matrices (sediment, biota, dust, human matrices) and explore automated sample prep coupled with the high MRM capacity to improve throughput and reproducibility.

Conclusion


The study demonstrates that an all-nitrogen APGC coupled to a tandem quadrupole mass spectrometer can effectively replace a helium/EI magnetic sector GC-HRMS reference method for analysis of 21 halogenated flame retardants. With appropriate adaptations—carrier flow ramping, injection pressure pulsing and multiple optimized MRM transitions—the N2/APGC-TQMS workflow delivered comparable chromatographic separation, shorter runtime, excellent linearity (average r2 = 0.9951 across 1–250 ppb) and improved operational sustainability and usability. This approach offers a practical, scalable alternative for environmental monitoring laboratories seeking high confidence quantitation with reduced cost and training burden.

References


1. Portolés T.; et al. Novel Analytical Approach for Brominated Flame Retardants Based on the Use of Gas Chromatography-Atmospheric Pressure Chemical Ionization-tandem Mass Spectrometry with Emphasis in Highly Brominated Congeners. Analytical Chemistry 2015, 87(19), 9892–9899.
2. Van Bavel B.; et al. Atmospheric Pressure Chemical Ionization tandem Mass Spectrometry (APGC/MS/MS) An Alternative to High-resolution Mass Spectrometry (HRGC/HRMS) for the Determination of Dioxins. Analytical Chemistry 2015, 87(17).
3. Concha-Graña E.; et al. Proposal of a Procedure for the Analysis of Atmospheric Polycyclic Aromatic Hydrocarbons in Mosses. Talanta 2015, 134, 239–246.
4. Bruce-Vanderpuije P.; et al. Background Levels of Dioxin-like Polychlorinated Biphenyls and Halogenated Dibenzo-p-dioxins and Dibenzofurans in Sera of Pregnant Women in Accra, Ghana. Science of The Total Environment 2019, 673.
5. Simpson A.K.; et al. Human Biomonitoring of Dioxins, Furans, and Non-ortho Dioxin-like Polychlorinated Biphenyls in Blood Plasma from Old Crow, Yukon, Canada. Science of The Total Environment 2024, 923.
6. Waters Corporation. Analysis of Polybrominated Diphenyl Ether Flame Retardants in Environmental Matrices Using Atmospheric Pressure Chemical Ionization GC/MS/MS, Document #720009524, 2026.
7. Fan X.; et al. Non-PBDE Halogenated Flame Retardants in Canadian Indoor House Dust: Sampling, Analysis, and Occurrence. Environmental Science and Pollution Research 2016, 23(8), 7998–8007.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Analysis of Polybrominated Diphenyl Ether Flame Retardants in Environmental Matrices Using Atmospheric Pressure Chemical Ionization GC-MS/MS
Improvements to the Analysis of PBDEs in Environmental Matrices: Transitioning from EI Magnetic Sector to GC-APCI TQ MS/MS
Converting Target Analysis of Organochlorine Pesticides in Environmental Matrices from Electron Ionization GC-HRMS Using Helium Carrier Gas to Atmospheric Pressure Ionization GC-MS/MS Using Nitrogen Carrier Gas
Routine Quantification of Polychlorinated Dibenzo-para-dioxins, Polychlorodibenzofurans and Dioxin-like Polychlorinated Biphenyls in Food by Atmospheric Pressure Chemical Ionization GC-MS/MS