Significance of the topic
Polychlorinated dibenzo-para-dioxins (PCDDs), dibenzofurans (PCDFs) and dioxin-like polychlorinated biphenyls (DL‑PCBs) are persistent organic pollutants with high toxicity and strong bioaccumulation potential. Dietary intake accounts for the vast majority of human exposure, so routine, reliable analytical monitoring of food and feed is essential for food safety, regulatory compliance and public health risk assessment. Advances in gas‑phase ionization and tandem mass spectrometry seek to deliver HRMS-equivalent quantitative performance with lower operational cost, reduced training burden and improved laboratory throughput.
Objectives and study overview
This application study assessed whether gas chromatography coupled with atmospheric pressure chemical ionization tandem quadrupole mass spectrometry (GC‑APCI‑MS/MS, implemented as APGC™ on a Xevo™ TQ Absolute) can provide routine quantification of the 17 regulated PCDD/F congeners and 12 dioxin‑like PCBs in a broad range of food and environmental sample matrices. Performance was benchmarked against a conventional high‑resolution GC‑EI‑HRMS workflow. Key aims were to evaluate quantitative agreement, method precision, robustness for routine operation, and compliance with Regulation (EU) 2017/771 performance criteria.
Methodology
- Sample set: 68 anonymized samples from official monitoring plans, including animal fats (pig, bovine, sheep), milk, eggs, grass, seafood, poultry muscle, food supplements and soils.
- Sample preparation: moisture removal by freeze‑drying or desiccation; pressurized liquid extraction (PLE); sulfuric acid clean‑up for pigment/grass matrices when required; automated multi‑column purification and fractionation (Miura® Automated System) to separate coplanar dioxins/PCBs from non‑coplanar PCBs. Quantification used isotopic dilution with 13C‑labelled internal standards.
- Calibration and QA: five‑point external calibration, daily mid‑range QC calibration point (QCcal) and an in‑house butter reference (QCb) injected repeatedly; response factor variation maintained within ±20% per EU rule; performance checks on chromatographic resolution.
- Data processing: TargetLynx™ software for automated peak integration and concentration calculation. Non-detect congeners set to zero for APGC workflow; HRMS software applied sample‑specific LOQ estimation from background, resulting in slight systematic differences at the very low end.
Instrumentation used
- GC‑APCI‑MS/MS system: Agilent 8890 GC with 7693A autosampler; APGC source coupled to Waters Xevo TQ Absolute triple quadrupole; MRM acquisition mode.
- GC columns: Optima 5MS Accent 60 m × 0.25 mm × 0.25 µm for dioxins and coplanar PCBs; HT8‑PCB 60 m × 0.25 mm × 0.25 µm for non‑planar PCBs.
- GC and injection: pulsed splitless injection for dioxins/coplanar PCBs (32 psi until 1 min) at 280 °C; splitless injection for non‑planar PCBs; programmed temperature ramps specific to each class.
- APGC source and MS settings (representative): source temperature 150 °C; heated transfer line 280 °C; corona current 2.0 µA; cone voltage 40 V; nitrogen makeup and cone/auxiliary gases (specified flows); nitrogen generator (Calypso DP) supplying ultra‑high purity N2 for makeup gas and lower‑purity N2 for MS. No source cleaning required during two weeks of continuous operation.
- Comparator HRMS: Agilent 7890A GC with 7683B autosampler; JMS‑800D magnetic sector HRMS operating at nominal resolution ~10,000 (10% valley) following established HRMS methods.
Key results and discussion
- Quantitative agreement: results for summed WHO‑TEQ2005 across sample types were highly correlated between GC‑APCI‑MS/MS and GC‑EI‑HRMS (R² = 0.9994) with a regression slope of 1.05, indicating nearly equivalent quantification across the full tested range (~1–50 pg/g WHO‑TEQ2005).
- Precision: QCcal RSD = 0.9% (n = 12); in‑house butter (QCb) RSD = 4.3% (n = 8) across batches and 4.3% over three months; comparable to HRMS precision (reported ~4.7%). Calibration response factors complied with the ±20% requirement of EU Regulation 2017/771.
- Sensitivity and LOQ: APGC produced adequate sensitivity with clear chromatographic peaks at the lowest calibration points for representative congeners (examples shown for PCB‑169 and 1,2,3,7,8,9‑HxCDF). Differences in LOQ treatment between platforms caused a small negative intercept in regression (HRMS software estimated LOQ from background while APGC workflow set undetected congeners to zero).
- Ionization control and humidity: APGC ionization mode (charge transfer vs proton transfer) depends strongly on source humidity. Monitoring of a phenanthrene species ratio was used to track the charge transfer fraction; a target ≥70% charge transfer was achieved and stabilized after column changes when the dedicated nitrogen supply was in continuous use. Initial stabilization time can be reduced from ~1 week to ~2 hours by continuously operating the nitrogen generator. Continuous instrument operation further improved ionization stability over injection series.
- Robustness and maintenance: the APGC/Xevo TQ Absolute system ran for two weeks of routine analyses without source cleaning or liner replacement; no maintenance actions were needed. The workflow proved operable by a relatively inexperienced operator after short training, illustrating reduced operator dependence compared with magnetic sector HRMS.
Benefits and practical applications
- Regulatory monitoring: APGC‑MS/MS provides quantitative performance equivalent to GC‑EI‑HRMS for routine screening and quantification of PCDD/Fs and DL‑PCBs across diverse food matrices, meeting EU regulatory requirements.
- Operational advantages: lower capital and running costs, smaller laboratory footprint, reduced calibration and stabilization demands, and shortened operator training from months to weeks compared with magnetic sector HRMS.
- Laboratory throughput and uptime: demonstrated two‑week continuous operation with stable sensitivity supports sustained monitoring campaigns with reduced downtime for maintenance.
Future trends and potential uses
- Wider adoption of APCI‑MS/MS for POPs: as triple quadrupole platforms and APGC sources mature, more routine laboratories can adopt cost‑effective alternatives to magnetic sector HRMS for regulated dioxin/PCB analysis.
- Automation and routine QA workflows: integration with automated purification (e.g., Miura) and robust nitrogen supply strategies will further reduce stabilization time and operator intervention.
- Method harmonization: continued inter‑laboratory comparisons and standardized LOQ/processing rules will be important to ensure comparable reporting between APGC‑MS/MS and HRMS platforms, particularly at concentrations near regulatory limits.
- Expanded analyte scope: the gentle ionization of APGC may enable improved detection of other non‑polar legacy and emerging contaminants when combined with optimized MRM transitions and chromatographic separations.
Conclusion
This study demonstrates that GC‑APCI‑MS/MS (APGC coupled to a Xevo TQ Absolute) provides robust, precise and accurate quantification of regulated PCDD/Fs and DL‑PCBs in a wide variety of food and environmental matrices, with results equivalent to GC‑EI‑HRMS across the tested concentration range. The approach reduces laboratory footprint, operator training and routine maintenance burden while meeting EU performance criteria, making it a practical alternative for routine monitoring laboratories.
References
- Commission Regulation (EU) 2017/771 amending Regulation (EC) No 152/2009 regarding methods for the determination of dioxins and PCBs. Official Journal of the European Union, 2017.
- AFSCA. Contamination of food and feed by dioxins and PCBs: FAQ. Version 2, July 2018.
- EC Regulation (EU) No 277/2012 on maximum levels and action thresholds for dioxins and PCBs in feed. 2012.
- Commission Regulation (EU) 2023/915 on maximum levels for certain contaminants in food. Official Journal of the European Union, 2023.
- Van Bavel, B.; et al. Atmospheric‑pressure Chemical Ionization Tandem Mass Spectrometry (APGC/MS/MS) — An Alternative to High‑resolution GC/HRMS for the Determination of Dioxins. Analytical Chemistry, 2015, 87 (17), 9047–9053. DOI:10.1021/acs.analchem.5b02264.
- Li, X.; et al. Application of GC‑APCI‑MS/MS in determination of PCBs and PCDD/Fs in Chinese Mitten Crab food webs. Chemosphere, 2021, 265, 129055. DOI:10.1016/j.chemosphere.2020.129055.
- Hagberg, J. Analysis of brominated dioxins and furans by HRGC/HRMS. Journal of Chromatography A, 2009, 1216 (3), 376–384. DOI:10.1016/j.chroma.2008.10.022.
- Cariou, R.; et al. Prediction of PCDD/F and dl‑PCB WHO‑TEQ content based on contribution of six congeners: toward screening approaches for fish. Environmental Pollution, 2010, 158 (3), 941–947. DOI:10.1016/j.envpol.2009.09.010.