LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
IndustriesEnvironmental, Forensics
ManufacturerShimadzu
Significance of the topic
River and wastewater monitoring for contaminants of environmental concern (CECs) is essential for assessing pollutant sources, evaluating wastewater treatment performance, and detecting emerging pollutants that can affect ecosystems and human health. Non-targeted analysis (NTA) using high-resolution LC/QTOF with data-independent acquisition (DIA) enables broad chemical profiling without a priori compound lists, making it well suited for environmental surveillance where unknown or unexpected substances (pharmaceuticals, industrial additives, road runoff markers) may occur.
Objectives and study overview
- Apply direct-injection LC/QTOF with DIA-MS/MS for non-targeted profiling of river water from two London rivers (tidal River Thames and the urban Hogsmill tributary).
- Use automated NTA software to detect, align, statistically compare features across sampling sites, and annotate compounds using multiple spectral libraries.
- Assess spatial patterns related to wastewater treatment facility (WWTF) discharges and identify significant CECs, including pharmaceuticals, industrial additives and road-derived contaminants.
Methodology and instrumentation
- Sampling: Grab samples collected at multiple sites along two river systems, including locations upstream and downstream of a WWTF; pooled QC prepared from aliquots.
- Sample preparation: 500 µL river water filtered by centrifugal PVDF 0.22 µm filters (4 min at 12,000 g); direct injection of 40 µL (autosampler at 4 °C) with minimal further cleanup.
- Chromatography: Reversed-phase separation using Shim-pack Velox Biphenyl column (2.1 × 100 mm, 2.7 µm) at 40 °C; flow 0.3 mL/min; mobile phases A: water + 2 mM ammonium formate + 0.002% formic acid and B: methanol + 2 mM ammonium formate + 0.002% formic acid; gradient from 5% to 100% B with a 13–17 min re-equilibration.
- Mass spectrometry acquisition: LCMS-9050 QTOF with electrospray ionisation (positive and negative modes); interface ~300 °C; heat block 400 °C, DL 250 °C; nebulizing gas 2 L/min, heating/drying gas 10 L/min; TOF-MS m/z 100–1000; DIA-MS/MS m/z 40–1000 with 35 windows (25 ms each), collision energy 5–55 V, ~0.925 s cycle time.
- Data processing: LabSolutions Insight Profiler for non-targeted feature detection, retention alignment, statistical analyses (PCA, t-test, one-way ANOVA, Tukey HSD) and compound annotation via concurrent searches against multiple libraries (MassBank, Wiley, Collective Spectral Library (CSL), Shimadzu HRAM forensic/toxicology libraries and formula-based precursor screening).
Used instrumentation
- UHPLC: Nexera X2 with Shim-pack Velox Biphenyl column (2.1×100 mm, 2.7 µm).
- Mass spectrometer: LCMS-9050 Quadruple Time-of-Flight (QTOF) mass spectrometer (Shimadzu) with DIA-MS/MS capability.
- Software: LabSolutions Insight Profiler for automated NTA workflows and library matching.
- Consumables: Millipore Ultrafree centrifugal filters (PVDF 0.22 µm).
Major results and discussion
- High feature count and library matches: From 7,229 aligned features the workflow returned 1,925 library identifications; initial pairwise statistics yielded 170 library-identified features as statistically significant (t-test p < 0.05).
- Group refinement improves statistical power: Principal component analysis (PCA) of Hogsmill samples revealed intra-group heterogeneity; splitting Hogsmill into downstream (H1) and upstream (H2) subsets increased discriminatory power. A subsequent one-way ANOVA across three groups (River Thames, Hogsmill H1, Hogsmill H2) identified 477 library-annotated features as significant and produced robust intergroup p-values with Tukey HSD comparisons.
- Key contaminants and spatial patterns: Sites downstream of the Hogsmill WWTF exhibited markedly elevated levels of prescription drugs and metabolites (venlafaxine and desmethylvenlafaxine, tramadol and O-desmethyltramadol, gabapentin, ketamine, clarithromycin, metronidazole, carbamazepine, lidocaine, lamotrigine). Classic wastewater tracers acetaminophen and caffeine were found ubiquitously at relatively consistent low levels.
- Non-pharmaceutical CECs: Dimethyldiallylammonium (a flocculant used in wastewater treatment) was among the most significant identifications downstream of the WWTF, indicating treated effluent influence. Hexa-methoxymethyl-melamine (HMMM), linked to road runoff and coatings/plastics, was detected and showed trends of increasing concentration downstream in the River Thames.
- Library-driven identification: DIA MS/MS spectra compared to multiple libraries (Wiley, CSL, MassBank, Shimadzu HRAM) enabled high-confidence matches for numerous compounds (example matches: lamotrigine, dimethyldiallylammonium, acetaminophen, HMMM). The study underlines the importance of comprehensive spectral repositories for NTA success.
- Statistical caveats: High within-group variance can mask true differences (e.g., clarithromycin initially not significant until groups were redefined), emphasising the need for careful grouping and QC-driven variance filtering in NTA studies.
Benefits and practical applications of the method
- Rapid, minimal-preparation workflow: Direct injection of filtered river water reduces sample handling, accelerates throughput, and lowers contamination risk compared to exhaustive extraction strategies.
- Broad screening capability: DIA with HRAM QTOF captures MS/MS data across wide m/z windows, enabling retrospective interrogation and discovery of unexpected CECs.
- Integrated automated data pipeline: LabSolutions Insight Profiler consolidates feature finding, alignment, statistics and library matching into configurable batch workflows suited to environmental monitoring programs.
- Environmental surveillance and source attribution: The approach successfully distinguished WWTF-impacted sites from less-impacted locations and identified compounds that serve as wastewater or urban runoff tracers.
Future trends and possibilities for application
- Expansion of spectral repositories: Growth and curation of community spectral libraries (including collated CSL resources) will improve annotation rates and confidence for environmental NTA.
- Complementary sampling and enrichment: Combining direct injection with periodic targeted enrichment, passive samplers, or solid-phase extraction could lower detection limits for trace contaminants while retaining broad screening capability.
- Advanced data science: Machine learning, suspect-list prioritisation and improved statistical workflows can better handle complex variance patterns and prioritize features of ecological or public-health relevance.
- Integration with wastewater-based epidemiology: Coupling NTA chemical profiles with biological markers (e.g., antibiotic resistance genes) and hydrological models can deepen source attribution and risk assessment.
- Instrumental innovation: Adding ion-mobility separation, higher acquisition speeds, or hybrid fragmentation schemes could enhance separations and spectral specificity for co-eluting isomers in complex environmental matrices.
Conclusion
This study demonstrates that direct-injection LC/QTOF with DIA-MS/MS and automated non-targeted data processing can effectively characterise complex river water chemistry with minimal sample preparation. The workflow detected a broad range of CECs, distinguished WWTF-impacted locations, and highlighted contaminants of emerging concern such as HMMM and treatment-related chemicals like dimethyldiallylammonium. Success depended strongly on grouping strategy, QC management and access to multiple spectral libraries, pointing to opportunities for methodological refinement and enhanced library resources to improve environmental monitoring programs.
References
- Lessmann, K., et al. (2026) Collective Spectral Library (CSL); Federal Institute of Hydrology (Bundesanstalt für Gewässerkunde). DOI: 10.5281/zenodo.16901589