Untargeted Metabolomics Analysis of Acute Respiratory Illnesses in Children using LC/QTOF and LabSolutions Insight Profiler

Applications | 2026 | ShimadzuInstrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS, Software
Industries
Metabolomics
Manufacturer
Shimadzu

Significance of the topic

Metabolomics applied to acute respiratory infections (ARIs) in children addresses a key clinical need: better phenotyping to support accurate diagnosis and improved patient management. Conventional diagnostic tests sometimes lack sensitivity or specificity and do not capture host metabolic responses that differentiate viral etiologies. High-resolution LC/QTOF combined with data-independent acquisition (DIA) MS/MS and dedicated data-processing software can reveal metabolic and lipidomic signatures that distinguish SARS-CoV-2, influenza A, and control pediatric cohorts, supporting biomarker discovery and downstream targeted assay development.

Study objectives and overview

The study aimed to demonstrate an end-to-end untargeted metabolomics workflow using UHPLC-DIA-MS/MS and LabSolutions Insight Profiler to (1) profile whole blood extracts from pediatric patients with ARIs, (2) detect metabolic differences between SARS-CoV-2, influenza A and control groups, and (3) showcase automated feature detection, statistical filtering and MS/MS-based compound annotation for biomarker discovery.

Sample preparation

  • Whole blood (100 µL) from age- and gender-matched pediatric subjects: SARS-CoV-2 (n=21), influenza A (n=28), controls (n=10).
  • Extraction: 300 µL ice-cold methanol added to each sample.
  • Quality control: pooled QC aliquots prepared by mixing equal volumes from each extract following mQACC recommendations to monitor analytical performance and enable filtering criteria.

Used instrumentation

  • UHPLC: Nexera X2 with reversed-phase C18 column (2.1 × 100 mm, 1.7 µm), column oven 50 °C, flow 0.4 mL/min. Mobile phases: A = water + 0.1% formic acid; B = acetonitrile + 0.1% formic acid. Gradient ramp to 90–100% B across 25–31 min; rinse solvents included IPA/MeOH/MeCN/H2O with 1% formic acid. Injection volumes: 0.5 µL (positive) / 1.0 µL (negative).
  • MS: Shimadzu LCMS-9050 QTOF in data-independent acquisition (DIA) mode. ESI voltages: +1.5 kV / -3.0 kV. Interface temp 300 °C; heat block 400 °C; DL 250 °C. Gas flows: nebulizing 3 L/min, heating 10 L/min, drying 15 L/min. CID gas pressure 230 kPa.
  • Acquisition strategy: DIA MS and MS/MS across m/z ranges ~60–1000 (MS) and 40–1000 (MS/MS), implemented as 27 precursor isolation events (35 Da windows) with collision energy spread 5–55 V and total cycle time ≈1 s.
  • Software: LabSolutions Insight Profiler for feature detection, alignment, filtering, multivariate and univariate statistics, and MS/MS spectral matching (integrated access to in-house library, MS-DIAL MSP kit and LipidBlast).

Data processing workflow and acceptance criteria

  • All DIA LC/QTOF data were processed in a single LabSolutions Insight Profiler cascade combining peak detection, alignment, gap filling and statistical evaluation.
  • Gap filling: recursive detection for features absent in some samples (likely below threshold) to reduce missingness prior to statistical analysis.
  • QC and feature-level filters: retain features with QC relative standard deviation (RSD) <20% and presence in ≥75% of at least one biological group. Features with near-constant values across all samples (<10% RSD) were excluded as likely analytically derived or biologically uninformative.
  • Missing values after gap filling were imputed as one-fifth of the minimum observed value for that feature.
  • Statistics: one-way ANOVA for multi-group testing with FDR correction by Benjamini–Hochberg (α = 0.05). Significant ANOVA results were followed by Tukey HSD post hoc pairwise comparisons.
  • Compound annotation: in-house authentic standards (MSI level 1) plus MS-DIAL MSP and LipidBlast libraries for putative (MSI level 2) assignments; MSI level 2 calls were manually reviewed as recommended in current identification guidance.

Main results and discussion

  • Overall findings: untargeted profiling identified 75 annotated metabolites and lipids that differed significantly between groups (ANOVA, FDR-adjusted p < 0.05).
  • Pairwise outcomes (Tukey HSD): 73 annotated features were significantly altered in SARS-CoV-2 vs controls, 35 in influenza A vs controls, and 9 features differed between the two infections. This pattern indicates substantial overlap but also infection-specific differences.
  • Top significant features: six highly significant, putatively annotated lipids (via LipidBlast MS/MS matching) included PC O-16:1_18:2, PC 18:2_18:2, PC 18:2_20:4, PE O-16:1_18:2, PC O-16:0_18:2 and PC 17:0_18:2. All six were decreased in both infection groups versus controls. Additionally, PC O-16:1_18:2 and PC O-16:0_18:2 were lower in SARS-CoV-2 than in influenza A.
  • Biological interpretation: observed reductions in multiple phosphatidylcholines and related ether lipids suggest perturbation of lipid metabolism common to viral ARIs in children, potentially reflecting altered membrane turnover, inflammation, or host immune–lipid interactions. Differences in magnitude between SARS-CoV-2 and influenza A for specific lipids point to nuanced host responses that could serve as differential biomarkers.
  • Visualization and review: LabSolutions Insight Profiler provided linked chromatograms, spectra, box plots and post hoc test results to facilitate interactive validation of candidate features and their MS/MS library matches.

Benefits and practical applications of the method

  • Integrated end-to-end workflow: combining DIA acquisition and a single automated processing cascade reduces manual steps and standardizes analysis, facilitating reproducibility in untargeted metabolomics.
  • Robust QC-driven filtering and gap filling improve confidence in reported features and reduce false discoveries arising from analytical variability.
  • MS/MS-enabled putative identification accelerates biological interpretation and prioritization of candidates for targeted verification and clinical validation.
  • Application potential: discovery of host-response biomarkers to support diagnostic algorithms, stratify disease severity or guide therapeutic decisions in pediatric ARIs; platform also applicable to other clinical metabolomics studies.

Future trends and potential developments

  • Deeper clinical validation: translating putative biomarkers into validated targeted assays (e.g., LC–MS/MS MRM panels) and evaluating performance in larger, diverse pediatric cohorts with known clinical outcomes.
  • DIA library expansion: construction of richer, instrument-specific DIA spectral libraries (including retention time indexing) to raise annotation confidence and reduce manual review load.
  • Standardization and QA/QC adoption: wider implementation of mQACC-compliant QC workflows and harmonized reporting standards to improve comparability across studies and laboratories.
  • Computational advances: integration of machine learning for feature prioritization, automated annotation confidence scoring, and multi-omic data fusion to assemble more robust diagnostic signatures.
  • Clinical translation: development of point-of-care-compatible panels derived from discovery data to support rapid differentiation of viral etiologies in pediatric emergency settings.

Conclusion

This application demonstrates that UHPLC-DIA-MS/MS combined with LabSolutions Insight Profiler can produce a streamlined, automated pipeline for untargeted metabolomics in a clinical biomarker-discovery context. The workflow identified reproducible lipid alterations that distinguish pediatric SARS-CoV-2 and influenza A infections from controls, with overlapping but distinct metabolic signatures. The integrated approach accelerates candidate discovery while preserving data quality through QC-driven filtering and MS/MS-based annotation, establishing a practical route toward validated diagnostic or prognostic assays.

References

  1. Kirwan JA, Gika H, Beger RD, et al. Quality assurance and quality control reporting in untargeted metabolic phenotyping: mQACC recommendations for analytical quality management. Metabolomics. 2022;18(9).
  2. Spicer R, Salek R, Steinbeck C. A decade after the metabolomics standards initiative it's time for a revision. Sci Data. 2017;4:170138.
  3. MS-DIAL. MS-DIAL metabolomics MSP spectral kit. Accessed May 2026.
  4. Theodoridis G, Fiehn O, Holčapek M, et al. What’s in a name? Metabolite identification: challenges and pitfalls in untargeted metabolomics. Metabolomics. 2026;22(22).

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

A non-targeted metabolomics analysis of SARS-CoV-2 and influenza infection in children
LabSolutions Insight Profiler Streamlines Analysis of Untargeted Metabolomics Data from LC/QTOF Analysis
Insight Profiler streamlines analysis of non-targeted metabolomics data from LCMS or direct injection in a single software solution
A Biological Model of the Ageing Metabolome Reveals Potential Clinically Relevant Biomarkers