LC/HRMS, LC/Orbitrap, LC/MS/MS, LC/MS, Software
IndustriesPharma & Biopharma
ManufacturerThermo Fisher Scientific
Significance of the Topic
Monoclonal antibodies (mAbs) are critical biotherapeutics, yet protein aggregation during production poses risks to efficacy and safety. Rapid, reliable monitoring of aggregates is essential for downstream purification process development and quality control. Native intact mass analysis combined with size-exclusion chromatography (SEC) and UV detection enables direct molecular weight measurement of aggregates, supporting real-time decision-making in mAb manufacturing.
Study Objectives and Overview
The primary aim was to establish an SEC-UV-MS platform for:
- Accurate determination of molecular weights for monomers and high-mass species (HMMS) of mAbs.
- Quantitative monitoring of aggregate levels during purification.
- Evaluation of POROS Caprylate mixed-mode cation exchange resin effectiveness in removing aggregates.
Materials, Methods, and Instrumentation
Sample Preparation:
- NISTmAb standard and tetrameric β-galactosidase (≈466 kDa) for sensitivity and accuracy tests.
- In-house mAb1 samples collected from Protein A pool (high aggregates) and post-polish pool.
- Vanquish Horizon UHPLC system with MAbPac SEC-1 column (4 × 300 mm, 300 Å) at 30 °C, 250 µL/min, 50 mM ammonium acetate buffer.
- UV detection at 280 nm.
- Orbitrap Excedion Pro BioPharma hybrid mass spectrometer in full MS mode.
- Extended high mass range up to m/z 12,000; resolution 60,000 at m/z 200.
- ESI source and MS parameters optimized for high-mass protein analysis.
- BioPharma Finder 5.3 software for deconvolution and average mass determination.
Results and Discussion
Sensitivity and Mass Accuracy:
- β-galactosidase tetramer measured at 466,560 Da, matching the expected 466,310 Da within <100 ppm.
- NISTmAb deconvoluted glycoforms achieved <10 ppm mass accuracy.
- SEC-UV profiles showed a 2.2% HMMS peak in pre-polish sample, absent post-polish.
- Deconvolution of HMMS peaks identified mAb1 plus two light chains (≈192,605 Da) and intact mAb1 dimer (≈294,159 Da).
Benefits and Practical Applications
This method offers:
- Label-free, direct molecular weight measurement of monomers and aggregates.
- High sensitivity for detecting low-level HMMS in early development.
- Quantitative monitoring to guide process optimization and resin selection.
- Compatibility with established UHPLC-SEC workflows.
Future Trends and Applications
Anticipated developments include:
- Expansion of mass range and resolution for multi-subunit complexes above 500 kDa.
- Integration with automated inline sampling and feedback control for continuous processing.
- Advanced algorithms and AI-driven deconvolution to streamline data interpretation.
- Coupling with orthogonal separations (e.g., ion exchange) for comprehensive process analytics.
Conclusion
The developed SEC-UV-native intact mass platform leveraging the Orbitrap Excedion Pro BioPharma MS enables rapid, accurate detection and quantification of mAb aggregates. It provides a robust tool for process development, ensuring control of critical quality attributes and supporting efficient purification strategies.