ICP/MS, ICP/MS/MS
IndustriesManufacturerThermo Fisher Scientific
Importance of the Topic
Halogen analysis is critical in many industries including environmental monitoring, pharmaceuticals, oil and gas, battery manufacturing and renewable fuels. Conventional methods such as ion chromatography are reliable but time-consuming, while plasma-based techniques like ICP-MS suffer from poor sensitivity and spectral interferences due to the high ionization potentials of halogens.
Aim and Study Overview
This study demonstrates a comprehensive workflow using the Thermo Scientific iCAP TQe triple quadrupole ICP-MS for accurate quantification of fluorine, chlorine and bromine. By operating in TQ-O2 mode with optimized conditions, the method aims to overcome ionization challenges and interferences inherent to halogen detection.
Methodology and Instrumentation
The approach uses indirect detection of fluorine as a barium–fluoride adduct (BaF+) at m/z 157, alongside direct oxide measurements for chlorine and bromine. Pure oxygen cell gas is introduced via dedicated mass flow controllers to remove interferences. Calibration solutions for fluorine contain 30 mg/L barium in 2% HNO3. Key operating parameters include:
- Plasma power: 1550 W
- Nebulizer gas flow: 0.72–1.08 L/min depending on analyte
- Quartz nebulizer and injector (1.5–2.5 mm ID)
- Cell gas flow: 0.3–0.4 mL/min O2
- Dwell times: 0.2–2 s per isotope
Main Results and Discussion
- Instrument detection limits (IDLs): chlorine 2.36 µg/L, bromine 0.05 µg/L, fluorine 0.034 µg/L
- Linearity: correlation coefficients (R²) >0.996 for all analytes
- Spike recoveries in drinking water: chlorine 87–101% (RSD ≤7.1%), bromine 97% (RSD 1.8%), fluorine 118% (RSD 6.8%)
- Survey scans confirmed barium isotope patterns and exclusive detection of BaF+ at m/z 157
Benefits and Practical Applications
- Delivers interference-free quantification of halogens at trace levels
- Streamlines analysis of chlorine and bromine without complex sample preparation
- Enables accurate fluoride monitoring via indirect BaF+ detection
Future Trends and Potential Applications
The workflow is suitable for coupling with liquid chromatography–ICP-MS for screening emerging contaminants like PFAS, and may be extended to speciation studies of halogenated compounds in various matrices.
Conclusion
Triple quadrupole ICP-MS in TQ-O2 mode with pure oxygen cell gas provides a robust, sensitive and accurate platform for halogen analysis. The validated method achieves low detection limits, high precision and reliable interference removal, broadening ICP-MS utility across multiple sectors.
References
- Jamari, N. L. A. et al. J. Anal. At. Spectrom. 32 (2017), 942-950