Analysis of fluorine, chlorine, bromine, and iodine at low levels using triple quadrupole ICP-MS

Applications | 2024 | Thermo Fisher ScientificInstrumentation
ICP/MS, ICP/MS/MS
Industries
Food & Agriculture
Manufacturer
Thermo Fisher Scientific

Significance of the Topic


The accurate measurement of halogens at trace levels is vital in diverse fields such as environmental monitoring, pharmaceutical quality control, oil and gas analysis, Li-ion battery research, renewable fuel evaluation, and investigations of per- and polyfluoroalkyl substances (PFAS). Halogens often exhibit poor ionization efficiency and suffer from severe spectral interferences in conventional ICP-MS, creating a need for advanced analytical approaches to ensure reliable detection and quantification at low concentrations.

Study Objectives and Overview


This application note outlines a workflow for the determination of fluorine, chlorine, bromine, and iodine at sub-µg/L levels using triple quadrupole ICP-MS (ICP-MS/MS) on the Thermo Scientific™ iCAP™ MTX instrument. The goals are to demonstrate interference removal, enhance sensitivity for high ionization potential elements, and validate method performance in terms of detection limits, linearity, and accuracy.

Utilized Instrumentation


  • Thermo Scientific iCAP MTX ICP-MS operated in triple quadrupole mode with TQ-O2 reaction cell.
  • Thermo Scientific Qtegra™ ISDS software for instrument control and data processing.
  • Thermo Scientific iSC-65 Autosampler for unattended analysis.
  • Sample introduction system including quartz cyclonic spray chamber (cooled to ~2.7 °C), PLUS torch, iCAP MX Series nebulizer, nickel sampler and skimmer cones.
  • Pure oxygen and helium cell gases delivered via dedicated mass flow controllers.

Methodology and Sample Preparation


Calibration standards for chlorine and bromine were prepared by diluting certified single‐element solutions in 2 % HNO₃. Iodine standards used 0.5 % tetramethylammonium hydroxide in 0.5 % HNO₃, while fluorine standards were prepared in a 30 mg/L barium solution to promote BaF⁺ adduct formation. All analyses employed the Sensitivity (S) mode of the iCAP MTX to maximize signal yield. For fluorine, optimization of Ba concentration and cell gas flow was critical to generate a stable Ba.F⁺ signal at m/z 157 and suppress background from H₃O⁺, H₂O⁺ and Ba‐based polyatomics. Chlorine, bromine, and iodine were measured as oxide reaction products (e.g., ³⁵Cl→³⁵Cl.¹⁶O at m/z 51) to reduce isobaric and polyatomic interferences.

Main Results and Discussion


Instrument Detection Limits (IDLs) achieved for each analyte were:
  • Chlorine: 2.6 µg/L (m/z 51)
  • Bromine: 0.026 µg/L (m/z 95)
  • Iodine: 0.0016 µg/L (m/z 127)
  • Fluorine (as BaF⁺): 27 µg/L (m/z 157)

Correlation coefficients (R²) exceeded 0.994 for all elements, demonstrating excellent linearity. Spike‐recovery studies in drinking water returned recoveries between 87 % and 108 % with relative standard deviations below 7 %, confirming method accuracy and precision. The use of pure oxygen in the collision/reaction cell allowed effective removal of xenon‐ and oxide‐based interferences, particularly benefiting chlorine and iodine determinations. The triple quadrupole configuration was essential for fluorine, since single‐quadrupole ICP-MS could not achieve sufficient mass resolution or interference rejection at m/z 157.

Benefits and Practical Applications


  • Direct, interference‐free quantification of chlorine, bromine, and iodine in complex matrices.
  • Novel indirect measurement of fluorine via Ba.F⁺ adduct formation, overcoming low ionization efficiency.
  • High throughput and unattended operation with autosampler integration.
  • Compatibility with a range of sample types including environmental, biological, and industrial solutions.
  • Complementary use alongside chromatographic techniques for comprehensive halogen speciation.

Future Trends and Opportunities


Ongoing developments are expected to focus on:
  • Enhanced adduct‐based approaches for other non‐metal analytes with high ionization potentials.
  • Integration with chromatographic separation (e.g., HPLC) for speciation of halogen organic compounds, including PFAS.
  • Automation of sample pre‐treatment to further improve throughput.
  • Expansion of cell gas chemistries and resolution settings to tackle emerging interferences in complex matrices.

Conclusion


The triple quadrupole ICP-MS workflow on the iCAP MTX platform provides a robust solution for trace elemental analysis of challenging halogens. By leveraging a tailored reaction cell strategy and optimized sample introduction, the method delivers low detection limits, high accuracy, and reliable interference removal. This technique supports critical applications across environmental, pharmaceutical, and industrial sectors requiring precise halogen quantification.

Reference


Jamiri, N.L.A. et al. Novel non-target analysis of fluorine compounds using ICP-MS/MS and HPLC-ICP-MS/MS, Journal of Analytical Atomic Spectrometry 2017, 32, 942–950.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

WPC: Sensitive and interference free analysis of halogens using triple quadrupole ICP-MS
Analysis of process water from hydrogen fuel cells using triple quadrupole inductively coupled plasma mass spectrometry (ICP-MS)
Accurate analysis of major components and trace level impurities in cathode active materials used in lithium ion battery production
Measurement of traces of heavy metals in cosmetic raw materials and finished products according to ISO/DIS 21392:2021 using triple quadrupole ICP-MS