Analytical testing of trace elements in refinery products using a robust ICP-MS approach

Applications | 2020 | Thermo Fisher ScientificInstrumentation
ICP/MS
Industries
Energy & Chemicals
Manufacturer
Thermo Fisher Scientific

Importance of the Topic


Control of trace metals such as nickel, vanadium, mercury and lead in petroleum streams is essential to prevent catalyst poisoning, reduce toxic emissions, and ensure regulatory compliance and operational efficiency.

Objectives and Study Overview


This application note presents the development of a robust ICP-MS method capable of quantifying trace elements across diverse refinery products—from high-viscosity crude and fuel oils to light and heavy naphtha—using a unified analytical approach.

Materials, Methods, and Instrumentation


Samples were prepared by xylene dilution (or direct aspiration for heavy naphtha), spiked with yttrium and indium internal standards (20 µg/L). Calibration covered 0.1–50 µg/L for most elements and 5–100 mg/L for sulfur.
  • ICP-MS systems: Thermo Scientific iCAP RQ (single quad) and iCAP TQ (triple quad)
  • Autosampler: Teledyne CETAC ASX-560
  • Sample introduction: Quartz cyclonic spray chamber at –5 °C, PFA microflow nebulizer, quartz injector
  • Auxiliary oxygen flow (20%) to control carbon deposition
  • Collision/reaction gases: He (KED), H₂, O₂ and NH₃ (TQ mode)


Main Results and Discussion


Single-quadrupole analysis with He KED and H₂ achieved detection limits from sub-µg/L to low mg/L. Triple-quadrupole operation with reactive gases further suppressed polyatomic interferences, lowering sulfur detection to ~13 µg/L and vanadium to 0.001 µg/L. Accuracy was confirmed by NIST SRM 1634c (recoveries 98–105%, RSD < 5%) and spike tests in naphtha (88–113% recoveries). A 12-hour mixed-matrix batch demonstrated stable internal standard responses (90–110%) and QC recoveries within ±10%.

Benefits and Practical Applications


  • One method for crude oil, fuel oils and naphtha
  • Low detection limits with effective interference control
  • Consistent performance over extended runs
  • Simplified preparation via xylene dilution
  • Supports process and quality control in petrochemical labs


Future Trends and Opportunities


Advanced triple-quadrupole ICP-MS with customizable reactive gases will drive further improvements in selectivity and sensitivity. Integration of automated sample handling, inline dilution and AI-based data correction could streamline high-throughput workflows and enhance analytical reliability.

Conclusion


A unified ICP-MS workflow combining optimized sample introduction, dynamic cell gas strategies and triple-quadrupole filtering delivers low detection limits, high accuracy and long-term stability for trace element analysis across refinery products, meeting stringent industry requirements.

Reference


  • NIST SRM 1634c, Trace Elements in Fuel Oil (NIST, Gaithersburg, MD, USA)

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Highly sensitive and reliable analysis of distillate products as per ASTM Method D8110 using single quadrupole ICP-MS
iCAP TQ ICP-MS Applications Compendium
iCAP TQ ICP-MS Applications Compendium
2019|Thermo Fisher Scientific|Guides
Effectively increase sample turnover and boost data confidence using triple quadrupole ICP-MS
Elemental and isotopic analysis: solutions for food authenticity, quality and safety