Single-shot quantitative plasma proteomics using the PASEF method

Posters | 2020 | BrukerInstrumentation
Ion Mobility, LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
Industries
Proteomics
Manufacturer
Bruker

Significance of the topic


The human plasma proteome is a critical resource for biological and clinical research but presents a vast dynamic range of protein concentrations that challenges in-depth, high-throughput analysis. The PASEF (parallel accumulation–serial fragmentation) approach on the timsTOF Pro platform enhances sequencing speed, ion utilization, and sensitivity, offering a streamlined workflow for large-cohort plasma studies without extensive depletion or fractionation.

Objectives and Study Overview


This study evaluated single-shot, label-free plasma proteomics using the PASEF method. Specific aims were to optimize the ion mobility range, assess different chromatographic gradient lengths, and explore library-based matching strategies to maximize protein identifications in undepleted plasma samples.

Methodology


  • Sample preparation: Pooled human plasma was digested with trypsin and cleaned using PreOmics iST kits.
  • Depletion workflow: Selective removal of abundant proteins via Agilent Human-6 and Sigma IgY14 columns, followed by iST digestion.
  • Chromatography: 200 ng peptide loads on a 25 cm Aurora C18 column for 30–45 min gradients; a Bruker TEN column was used for 20 min gradients.
  • Data acquisition: Data-dependent PASEF in oTOF control software for DDA runs; 25 min dia-PASEF experiments also performed.
  • Data analysis: DDA raw files processed in MaxQuant (1% FDR, matching between runs); Spectronaut (Biognosys) used for dia-PASEF quantification.

Instrumentation Used


  • nanoElute UHPLC system with Aurora 25 cm C18 and Bruker TEN columns
  • timsTOF Pro mass spectrometer (Bruker Daltonics)
  • oTOF control software, MaxQuant, Spectronaut for data processing

Main Results and Discussion


  • Optimized ion mobility range (0.85–1.35 1/K0) yielded the highest protein identifications.
  • In undepleted plasma, single-shot DDA PASEF achieved an average of 345 protein groups with a 45 min gradient, and ~280 with 20 min.
  • Matching 30 min DDA runs to a depletion-based library increased identifications to >550 proteins (default matching) and >460 proteins under stringent retention (±6 s) and mobility (±0.01 1/K0) windows.
  • Library-assisted 25 min dia-PASEF quantified >410 proteins in undepleted samples.
  • Quantitative completeness exceeded 90% for most proteins across replicates; LFQ CV remained low for medium-to-high abundance targets.

Benefits and Practical Applications


  • Enables rapid, deep profiling of large plasma cohorts without costly depletion or extensive fractionation.
  • Streamlined label-free workflow reduces preparation time and cost.
  • Flexible balance between throughput and depth suits both discovery and routine QC applications.

Future Trends and Potential Applications


  • Longer chromatographic gradients may approach identification of ~1000 plasma proteins in single runs.
  • Development of comprehensive spectral libraries or adoption of library-free DIA workflows to further boost coverage.
  • Application to biomarker discovery, clinical diagnostics, and expansive cohort studies for translational research.

Conclusion


Single-shot PASEF acquisition on the timsTOF Pro provides a fast, sensitive, and reproducible strategy for plasma proteome profiling. Library-based matching significantly enhances depth, making this approach ideal for high-throughput biological and clinical investigations.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

High throughput 4D-Proteomics – Application of dia-PASEF® and the Evosep One for short gradients
High throughput proteomics Application of dia PASEF and Evosep One for short gradients
Plasma proteomics goes high throughput – timsTOF Pro with PASEF and 4D feature alignment to quantify 500 plasma proteins in 11.5 min
dia-PASEF® applied on different gradient lengths