This compact device saves up to 90% of mobile phase by redirection of the pure solvent to the solvent reservoir during the isocratic HPLC. The device is powered by 5 V DC from computer USB port or power adapter. Working parameters (Threshold and Delay) are set using user friendly software.
Price: 635 € ex VAT
The UFO Solvent Recycler continuously monitors the output signal of the chromatography detector and displays its current value in the application on computer. A high-resolution analog-to-digital converter provides data for further evaluation in the built-in processor.
If the input signal level exceeds a preset value (Threshold), the UFO recycler will redirect the flow to the waste. When the signal drops, the solvent recycler waits for the transport delay and then switches the mobile phase back to the reservoir. An autosampler injection marker connected to the UFO recycler START input can be used to zero the input signal at the time of injection.
Watrex: Principle and process of mobile phase recycling during isocratic HPLC using the UFO system
FEATURES
- Very compact design
- Simple operation
TECHNICAL SPECIFICATION
- Detector compatibility: compatible with any HPLC detector
- Connection: 1/16″ O.D. tubing, 1/4″-28 flat bottom fittings
- Maximum pressure: 30 p.s.i./0,2 MPa
- Wetted material: PEEK, PTFE
- Input range: +/- 1 V or +/- 100 mV, user selectable
- Computer connection: USB-A
- TTL/contact closure input: configurable (ZERO, start, valve position)
- Power: 5V DC/<100 mA (power supply or computer USB port)
- User interface: PC Windows based software
Ordering information:
PN/Component:
- UFO000-001 UFO™ Solvent Recycler , PEEK version
- UFO000-002 UFO™ Solvent Recycler , PTFE version