HPLC
IndustriesEnvironmental
ManufacturerAgilent Technologies
Significance of the topic
Per- and polyfluoroalkyl substances (PFAS) are persistent, bioaccumulative environmental contaminants of growing concern. Their strong C–F bonds confer stability but also complicate ultra-trace analysis. Minimizing background PFAS from chromatographic systems is critical for sensitive detection at ng/L levels.
Objectives and overview of the study
This application note outlines the sources of PFAS contamination in Agilent 1260 and 1290 Infinity II liquid chromatography systems and presents a set of modifications to achieve a fluoropolymer-free flow path. The goal is to reduce background signals and enable reliable quantification of PFAS following EPA Method 537 guidelines.
Methodology
Key system modifications:
- Replace solvent bottle caps and inlet lines: switch from PTFE to PEEK tubing and polypropylene caps.
- Remove PTFE filter frits in pump heads and purge valves; bypass degasser to direct PEEK lines to the pump.
- Install a delay column (ZORBAX Eclipse Plus C18, 4.6×50 mm, 3.5 µm) between pump and injector for background separation.
- Modify autosampler components: use PEEK rotor seals, replace PTFE vent lines, and select PTFE-free vials, caps, and septa.
- Implement extended multiwash cycles (15–20 s) to mitigate PFAS carryover on the needle and seat.
Instrumentation used
• Agilent 1260 Infinity II or 1290 Infinity II LC system with high-speed pump and multisampler/autosampler.
• PEEK tubing (1/8”×0.062” id), polypropylene GL45 bottle caps, stainless steel inlet filters.
• Agilent ZORBAX Eclipse Plus C18 delay column (p/n 959943-902).
• Stainless steel filter sinkers, PEEK threaded fittings, ferrules, and lock rings.
• PTFE-free autosampler vials (1 mL) with polypropylene septa.
Main results and discussion
System baseline analysis showed a marked reduction in PFAS background after fluoropolymer removal. Coupling the modified LC to a Q-TOF MS eliminated contaminant peaks observed in the unmodified system. The delay column effectively separated residual PFAS background from analyte peaks, as demonstrated with PFOS. These modifications enabled trace-level quantitation down to EPA advisory limits (20–400 ng/L) without interfering signals.
Benefits and practical applications
• Enhanced sensitivity and lower detection limits for PFAS in water, soil, sludge, and biota.
• Robust workflow for dedicated trace analysis laboratories and QA/QC environments.
• Compatibility with multi-method LC systems via optional delay column installation.
Future trends and potential applications
Ongoing developments may focus on fully fluoropolymer-free chromatography systems, automated tubing and component changeovers, and integration with high-resolution mass spectrometry workflows. Emerging PFAS analogs will require adaptive background removal strategies and dedicated analytical protocols.
Conclusion
Implementing PEEK and polypropylene replacements, removing PTFE frits, and employing a delay column significantly reduces PFAS background in Agilent Infinity II LC systems. These modifications support ultra-trace PFAS analysis in compliance with regulatory advisories and enhance analytical confidence.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.