Determination of Dioxin in Food by GC-MS/MS Coupled with Boosted Efficiency Ion Source (BEIS)

Posters | 2020 | Shimadzu | ASMSInstrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Industries
Food & Agriculture
Manufacturer
Shimadzu

Importance of Topic


The determination of polychlorinated dibenzo-p-dioxins (PCDDs) and polychlorinated dibenzofurans (PCDFs) in food matrices is critical due to their persistence, bioaccumulation and potential health risks. Regulatory agencies require high-sensitivity and high-selectivity analytical methods to monitor trace-level contaminants and ensure food safety.

Objectives and Study Overview


This study evaluates a gas chromatography–tandem mass spectrometry (GC-MS/MS) method using the Shimadzu GCMS-TQ8050 NX instrument equipped with a boosted efficiency ion source (BEIS). The goals were to develop an optimized multiple reaction monitoring (MRM) method for PCDD/F analysis in food, assess linearity, accuracy and precision using low-level standards, and validate real-sample performance against consensus toxicity equivalent (TEQ) values from the Norwegian Institute of Public Health (NIPH).

Methodology and Instrumentation


Sample Preparation:
  • Standards: Native PCDD/Fs and 13C-labeled internal standards from Wellington Laboratories.
  • Calibration: Seven-point curves from 0.01 to 10 ng/mL (CSLQ to CS3).
  • Food samples: Extracted following EPA-1613B, lipid content determined gravimetrically, cleaned on multilayer silica gel, alumina and carbon columns.
Instrumental Conditions:
  • GC Column: DB-5 MS (60 m × 0.25 mm × 0.25 µm).
  • Oven Program: 120 °C (1 min) ramp to 220 °C at 43 °C/min (15 min), then to 250 °C and 260 °C, final hold at 310 °C.
  • Carrier Flow: Constant at 1.5 mL/min, injection volume 2 µL, EI 70 eV.
  • BEIS Conditions: Source 230 °C, interface 300 °C, CID pressure 150 kPa, detector voltage 1.8 kV.
  • Acquisition: MRM with loop time 1.4 s; transitions defined for each congener and its labeled analogue.

Main Results and Discussion


Calibration and Linearity:
  • All congeners displayed linear response (R² > 0.99) over the calibration range.
  • Average relative response factors (RRFs) ranged from 0.91 to 1.39, with RSDs < 12%.
Accuracy and Precision:
  • Quantification at the lowest standard (CSLQ×2) yielded relative errors below 20% for all components.
  • Method repeatability met regulatory precision criteria at trace levels.
Real Sample Analysis:
The TEQ values obtained from various food types matched NIPH consensus data, demonstrating the method’s robustness and suitability for routine monitoring.

Benefits and Practical Applications


The developed GC-MS/MS method with BEIS offers:
  • Enhanced sensitivity for trace-level PCDD/F detection.
  • Improved ionization efficiency and signal stability.
  • Faster analysis times compared to high-resolution MS alternatives.
  • Applicability to QA/QC laboratories and regulatory compliance testing.

Future Trends and Potential Applications


Emerging developments may include:
  • Integration with automated sample preparation for higher throughput.
  • Coupling with high-resolution mass analyzers for broader screening of dioxin-like compounds.
  • Miniaturized or portable GC-MS systems for on-site food safety monitoring.

Conclusion


The Shimadzu GCMS-TQ8050 NX with BEIS and MRM acquisition provides a reliable, accurate and sensitive approach for quantifying PCDD/Fs in food. The method meets regulatory requirements for linearity, precision and accuracy, and its performance aligns with established consensus TEQ values.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Determination of Dioxin in Food by GC-MS/MS Coupled with Boosted Efficiency Ion Source (BEIS)
Determination of Polychlorinated Dibenzo-p-dioxins and Dibenzofurans (PCDD/Fs) in Foodstuffs and Animal Feed Using a Triple Quadrupole GCMS-TQ8040 System with Smart MRM Transforms Laboratory Analysis
Analysis of Dioxins in Feed and Food Using GC-MS/MS as Confirmatory Method in Complying with EU Regulation
Analysis of Dioxins in Foods and Feeds Using GC-MS/MS