GC/MSD, GC/MS/MS, GC/QQQ
IndustriesFood & Agriculture
ManufacturerShimadzu
Importance of the Topic
Polychlorinated dibenzo-p-dioxins and dibenzofurans (PCDD/Fs) are persistent organic pollutants with high toxicity and the capacity to bioaccumulate in food chains. Regulatory agencies such as the European Union have established strict maximum levels for PCDD/Fs in foodstuffs and animal feed, making reliable monitoring methods essential for food safety and public health.
Study Objectives and Overview
This report describes the development and validation of a sensitive, selective, and reproducible method for quantifying 17 2,3,7,8-substituted PCDD/F congeners in food and feed. A Shimadzu GCMS-TQ8040 triple quadrupole system coupled with Smart MRM technology was employed to meet performance requirements comparable to high-resolution GC/HRMS methods mandated by EU regulations.
Methodology and Instrumentation
The analytical procedure combines modified US EPA Method 1613B sample preparation with Smart MRM-enhanced GC–MS/MS analysis:
- Accelerated solvent extraction of homogenized samples spiked with 13C-labeled surrogates.
- Multi-stage cleanup using acidic/basic silica, alumina, and carbon chromatography.
- Addition of 13C-labeled internal standards prior to analysis.
- GC–MS/MS analysis in multiple reaction monitoring mode with automated transition and collision energy optimization.
Used Instrumentation
- Shimadzu GCMS-TQ8040 triple quadrupole mass spectrometer with OFF-AXIS ion optics.
- DB-5MS capillary column (60 m × 0.25 mm × 0.25 µm).
- Splitless injector (290 °C) and argon collision gas (150 kPa).
- Smart MRM software for retention-time synchronized MRM and optimized dwell/loop settings.
Main Results and Discussion
The method delivered:
- Full chromatographic separation of all 17 target congeners within 60 min.
- Detection limits as low as 10 fg on-column for 2,3,7,8-TCDD, with linear calibration from 0.1 to 200 ng/mL (R² > 0.999).
- Intra- and interday repeatability with RSD < 15% for low-level standards.
- Real-sample analysis of fish oil and milk yielding TEQ values (29.5 pg WHO-TEQ2005/g fat and 1.38 pg WHO-TEQ2005/g fat, respectively) consistent with HRGC/HRMS results.
Benefits and Practical Applications
This triple quadrupole GC–MS/MS approach offers a rapid, cost-effective alternative to high-resolution mass spectrometry for routine PCDD/F screening and quantitation, fully compliant with EU legislation. Smart MRM significantly reduces method development time and enhances laboratory throughput.
Future Trends and Potential Applications
Emerging developments may include:
- Expansion of Smart MRM databases to cover additional environmental contaminants.
- Integration of automated sample prep and data analysis workflows.
- Development of multi-class methods encompassing PCDD/Fs, dioxin-like PCBs, and other regulated pollutants.
Conclusion
The Shimadzu GCMS-TQ8040 with Smart MRM delivers the sensitivity, linearity, and precision required for robust PCDD/F analysis in food and feed. It represents a practical solution for routine regulatory monitoring in compliance with EU standards.
Reference
- Institute of Medicine. Dioxins and dioxin-like compounds in the food supply: Strategies to decrease exposure. The National Academies Press, Washington DC, 2003.
- Commission Regulation (EU) No. 589/2014 of 2 June 2014 laying down methods of sampling and analysis for the control of levels of dioxins, dioxin-like PCBs and non-dioxin-like PCBs in certain foodstuffs.
- Commission Regulation (EU) No. 709/2014 of 20 June 2014 amending Regulation (EC) No. 152/2009 on determination of dioxins and PCBs.
- US EPA Method 1613B: Tetra- through Octa-Chlorinated Dioxins and Furans by Isotope Dilution HRGC/HRMS, 1997.
- Wang P, Zhang Q, Wang Y et al. Evaluation of extraction techniques for PCBs and PBDEs in soil and fish. Anal. Chim. Acta. 2010;663(1):43–48.