HPLC, LC/MS, LC/SQ
IndustriesPharma & Biopharma
ManufacturerWaters
Importance of the Topic
NDMA is a probable human carcinogen capable of causing DNA mutations; controlling its levels in pharmaceuticals is critical to safeguard patient health and comply with regulatory standards.
Objectives and Study Overview
This work demonstrates a fast and reliable HPLC-UV method for detecting N-nitrosodimethylamine in both ranitidine drug substance and tablet products, with added confirmation via mass spectrometry.
Methodology and Instrumentation
Ranitidine samples (30 mg/mL) were extracted in water, centrifuged, and filtered. NDMA calibration standards (10–300 ng/mL) were prepared from a 250 μg/mL stock. Separation was achieved on an XSelect HSS T3 column (4.6×100 mm, 3.5 μm) at 40 °C, 1.0 mL/min using a gradient from water with 0.02% formic acid to acetonitrile. UV detection was performed at 245 nm; mass confirmation used an ACQUITY QDa detector in ESI+ mode (m/z 50–500).
Used Instrumentation
- ACQUITY Arc System with 2998 PDA and QDa detectors
- XSelect HSS T3 Column (4.6×100 mm, 3.5 μm)
- Mobile Phase A: water + 0.02% formic acid; B: acetonitrile; gradient 95:5 to 5:95
- PDA scan range 210–400 nm, quantitation at 245 nm
- QDa MS: ESI+; m/z range 50–500; divert valve at 2.70 min
Main Results and Discussion
The method achieved baseline separation of NDMA (retention ∼2.2 min) from ranitidine in both substance and tablet matrices. Unspiked samples contained detectable NDMA; spiking at 50 ng/mL yielded an average recovery of 85%, demonstrating method accuracy and robustness. Mass spectral data provided rapid identity confirmation.
Benefits and Practical Applications
- Rapid sample preparation and analysis
- High sensitivity for trace NDMA detection
- Dual detection improves selectivity and confidence
- Applicable to raw API and finished dosage forms
Future Trends and Opportunities
Advances may include high-resolution MS for broader nitrosamine profiling, automated sample processing, inline process monitoring, and application to other pharmaceutical contaminants.
Conclusion
The proposed HPLC-UV method with mass confirmation offers a quick, accurate, and robust approach for controlling NDMA impurities in ranitidine products, supporting quality assurance and regulatory compliance.
References
- ICH M7(R1), Assessment and Control of DNA Reactive (Mutagenic) Impurities in Pharmaceuticals, ICH, 2018.
- FDA statement on NDMA found in ranitidine products, 2019.
- Maziarz M, Rainville P. Reliable HPLC/UV Quantitation of Nitrosamine Impurities in Valsartan and Ranitidine Drug Substances, Waters Technology Brief 720006775EN, 2020.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.