Characterization of Monoclonal Antibody Biosimilar through C-terminal and Disulfide Bond Peptides Sequencing Analysis on Q-TOF Mass Spectrometer

Applications | 2020 | ShimadzuInstrumentation
LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
Industries
Clinical Research
Manufacturer
Shimadzu

Significance of the Topic


The rapid expansion of biopharmaceuticals underscores the need to verify primary structure fidelity in biosimilars. Characterizing C-terminal variants and disulfide bond peptides is essential for quality assessment and ensuring clinical performance.

Objectives and Study Overview


  • Characterize C-terminal peptides with and without lysine truncation in a bevacizumab biosimilar.
  • Detect and sequence cysteine-containing peptides released from disulfide bonds.
  • Validate an EIC-based rapid screening approach with mass accuracy below 2 ppm.

Methodology and Instrumentation


  • Sample preparation: Reduction with DTT, alkylation with IAM and tryptic digestion of bevacizumab in Tris buffer.
  • Liquid chromatography: Shim-pack GISS C18 column (250×4.6 mm, 5 µm), 0.3 mL/min flow, gradient elution with 0.1% formic acid in water (A) and acetonitrile (B).
  • Mass spectrometry: LCMS-9030 Q-TOF with heated ESI (300 °C), MS/MS via CID with argon (230 kPa), mass accuracy <2 ppm.

Results and Discussion


  • TIC analysis revealed distinct peaks for tryptic peptides, enabling extraction of C-terminal and cysteine-linked ions.
  • Three C-terminal peptides and fifteen cysteine-containing peptides were detected and confirmed using extracted ion chromatograms with <2 ppm error.
  • Quantitative data showed predominant removal of C-terminal lysine from the heavy chain.
  • MS/MS de novo sequencing provided complete y-ion series for key peptides such as SLSLSPG and VYACEVTHQGLSSPVTK, including carbamidomethyl modifications.

Benefits and Practical Applications


  • The workflow delivers a fast and reliable method for monitoring C-terminal variants and disulfide mapping in mAb biosimilars.
  • High mass accuracy and EIC-based screening streamline quality control in biopharmaceutical production.
  • The approach supports regulatory compliance and can be adopted for routine batch release testing.

Future Trends and Possibilities


  • Automation of data processing and high-throughput LC-MS workflows for comprehensive mAb characterization.
  • Use of advanced fragmentation techniques and higher-resolution mass analyzers for deeper post-translational modification analysis.
  • Extension of the methodology to other biotherapeutic formats, including Fc-fusion proteins and antibody–drug conjugates.

Conclusion


The combination of targeted tryptic digestion, precise LC-Q-TOF analysis and EIC-based peptide identification offers an efficient, accurate platform for characterizing C-terminal and disulfide bond peptides in bevacizumab biosimilars, reinforcing critical quality attribute assessment.

References


  • L.W. Dick Jr, D. Qiu, D. Mahon, M. Adamo, K. Cheng. C-terminal lysine variants in fully human monoclonal antibodies: Investigation of test methods and possible causes. Biotechnology and Bioengineering, 100:1132–1143 (2008).

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Characterization of C-terminal and Disulfide Bond Peptides of Monoclonal Antibody (mAb) on Q-TOF Mass Spectrometer
IMSC: Full Characterization of Heterogeneous Antibody Samples Under Denaturing and Native Conditions on a Hybrid Quadrupole-Orbitrap Mass Spectrometer
Confident peptide mapping and disulfide bond analysis of an IgG2 monoclonal antibody
Peptide Mapping of Monoclonal Antibody (mAb) Using Nexera Bio with Q-TOF Mass Spectrometer for Full Sequence Confirmation