LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
IndustriesPharma & Biopharma
ManufacturerShimadzu
Significance of the Topic
Monoclonal antibody therapeutics are critical for treating chronic diseases, and ensuring their structural integrity is essential for biosimilar quality control. Peptide mapping of C-terminal sequences and disulfide bond linkages provides insight into primary structure and post-translational modifications.
Objectives and Study Overview
This study aimed to characterize a bevacizumab biosimilar by mapping and sequencing its C-terminal and cysteine-containing peptides using a Shimadzu LCMS-9030 Q-TOF mass spectrometer. Key goals included detecting lysine variants, disulfide-linked peptides, and confirming sequence integrity.
Methods and Instrumentation
- Sample preparation: Reduction with dithiothreitol, alkylation with iodoacetamide, tryptic digestion at 37 °C, quenching with trifluoroacetic acid and centrifugation.
- LC conditions: Shim-pack GISS C18 column (250 × 4.6 mm, 5 μm), 0.1 % formic acid in water and acetonitrile, gradient from 0 % to 50 % B over 85 min, flow rate 0.3 mL/min, column temperature 40 °C.
- MS settings: Heated electrospray ionization in positive mode, MS scan range m/z 250–2500, MS/MS fragmentation, nitrogen nebulizing and drying gases.
Main Results and Discussion
- TIC and extract ion chromatograms enabled detection of fifteen cysteine-containing peptides and three C-terminal peptides with mass accuracy < 2 ppm.
- Heavy chain C-terminal peptides exhibited lysine truncation, with the clipped form dominating (> 99 %) over the lysine-retained variant.
- Carbamidomethyl modifications on cysteines were confirmed, and de novo sequencing of selected peptides showed comprehensive b- and y-ion coverage, validating sequence assignments.
Benefits and Practical Applications
The described workflow delivers a fast and reliable approach for monitoring critical quality attributes of mAb biosimilars. The EIC-based method with high mass accuracy simplifies peptide identification and supports routine QC environments.
Future Trends and Opportunities
Advancements may include automated data processing, deeper coverage of complex post-translational modifications, integration of ion mobility for conformational analysis, and machine-learning-driven de novo sequencing to further streamline biosimilar characterization.
Conclusion
The LCMS-9030 Q-TOF system effectively characterized C-terminal and disulfide bond-linked peptides of a bevacizumab biosimilar. The study confirms the instrument’s capability for accurate peptide mapping and sequence confirmation in biosimilar development.
Used Instrumentation
- Shimadzu LCMS-9030 Q-TOF mass spectrometer
- Shim-pack GISS C18 column, 5 μm, 250 × 4.6 mm
- Heated ESI interface and nitrogen gas supplies
References
- Dick LW Jr, Qiu D, Mahon D, Adamo M, Cheng K. C-terminal lysine variants in fully human monoclonal antibodies: Investigation of test methods and possible causes. Biotechnol Bioeng. 2008;100:1132-1143.
- Protein Metrics. Byos software.
- MacLean B, et al. Skyline: an open source document editor for creating and analyzing targeted proteomics experiments. Bioinformatics. 2010;26:966–968.
- Shimadzu Application News AD-0176. Peptide Mapping of Monoclonal Antibody (mAb) Using Nexera Bio with Q-TOF Mass Spectrometer for Full Sequence Confirmation. 2019.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.