Separation of Free Fatty Acids by Mixed-Mode Anion Exchange Chromatography

Applications | 2020 | WatersInstrumentation
Consumables, LC columns
Industries
Food & Agriculture
Manufacturer
Waters

Importance of the Topic


Fatty acids serve as vital energy sources and structural components in food, biological tissues and microbial fermentation products. Accurate profiling of free fatty acids is essential across disciplines including food quality control, clinical biochemistry and metabolic research.

Objectives and Study Overview


This work aims to demonstrate efficient separation of linear free fatty acids ranging from propionic acid (C3:0) to stearic acid (C18:0) using a mixed-mode reversed-phase/anion exchange chromatography column. The separation performance, retention behavior and injection reproducibility were evaluated.

Methodology


A standard mixture of C3:0–C18:0 free fatty acids was prepared in methanol and chloroform. Chromatographic conditions employed a binary mobile phase: aqueous 1 mM ammonium formate at pH 3.0 (solvent A) and a 50 : 50 acetonitrile/isopropanol solution with 1 mM ammonium formate at pH 3.0 (solvent B). A linear gradient from 0 to 100 percent B over six minutes was applied at 0.6 mL/min. The column temperature was set to 60 °C and injection volume was 3 µL.

Instrumentation Used


  • Waters Atlantis PREMIER BEH C18 AX column, 2.1 mm × 50 mm, 1.7 µm bonded phase with tertiary alkylamine groups on bridged ethyl hybrid silica
  • High performance liquid chromatography system with autosampler and column oven

Key Findings and Discussion


The mixed-mode column achieved baseline resolution of free fatty acids from C3:0 to C18:0 within six minutes. Retention increased progressively with chain length, reflecting stronger hydrophobic interactions on the C18 ligand. A higher surface area (270 m2/g) of the PREMIER BEH C18 AX stationary phase enhanced retention compared to 185 m2/g BEH C18. Peak shapes improved for early eluting acids. Injection reproducibility was excellent, with retention time RSD values between 0.02 and 0.14 percent across five replicate injections.

Benefits and Practical Applications


  • Rapid analysis of a broad free fatty acid range without derivatization
  • High resolution and sensitivity suitable for routine food, clinical and bioprocess laboratories
  • Robust and reproducible retention behavior driven by mixed-mode interactions

Future Trends and Applications


Emerging directions include coupling mixed-mode columns to high resolution mass spectrometry for detailed lipidomics, extending separation to branched or unsaturated acids, and applying the method to complex biological matrices such as plasma or microbial culture extracts.

Conclusion


The mixed-mode Atlantis PREMIER BEH C18 AX column delivers fast, high resolution and reproducible separation of free fatty acids from C3:0 to C18:0. Its combined reversed-phase and anion exchange properties offer a versatile platform for routine fatty acid analysis.

References


  1. Harper’s Illustrated Biochemistry, 28th ed. McGraw-Hill, 2009.
  2. Ichihara K. et al. Preparation of Fatty Acid Methyl Esters for Gas-Liquid Chromatography. J. Lipid Res. 2010;51:635–640.
  3. Nikolova-Damyanova B. Reversed-Phase HPLC in Advances in Lipid Methodology, 1997;193–251.
  4. Yang Y. et al. Advances of Supercritical Fluid Chromatography in Lipid Profiling. J. Pharm. Anal. 2019;9:1–8.
  5. Walter TH. et al. New Mixed-Mode Reversed Phase/Anion Exchange Stationary Phase. Waters Application Note, 2020.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

A New Mixed-Mode Reversed-Phase/Anion-Exchange Stationary Phase Based on Hybrid Particles
LC-MS Analysis of Underivatized Bisphosphonate Drugs Using Mixed-Mode Reversed-Phase/Anion-Exchange Chromatography
IMPROVED PERFORMANCE OF MODERN MS-COMPATIBLE REVERSED-PHASE/ANION-EXCHANGE MIXED-MODE HPLC COLUMNS
EVALUATION OF NEW MS-COMPATIBLE MIXED-MODE RP/AX HPLC COLUMNS