Enhanced sensitivity in multiresidue pesticide analysis with the Orbitrap Exploris GC S mass spectrometer

Applications | 2026 | Thermo Fisher ScientificInstrumentation
GC/HRMS, GC/Orbitrap, GC/MSD, GC/MS/MS
Industries
Food & Agriculture
Manufacturer
Thermo Fisher Scientific

Significance of the topic


The reliable detection and quantification of pesticide residues in fruits and vegetables is a cornerstone of food safety and regulatory compliance. High-resolution, full-scan gas chromatography–mass spectrometry workflows provide both targeted sensitivity and the ability to retrospectively screen for unexpected or emerging contaminants. Improving sensitivity, robustness, and throughput in routine multiresidue pesticide analysis reduces false negatives at trace levels, supports lower reporting limits, and increases laboratory efficiency for high-volume testing facilities.

Goals and overview of the study


This technical evaluation aimed to demonstrate the analytical performance of the Orbitrap Exploris GC S mass spectrometer equipped with the NeverVent Advanced Electron Ionization (AEI) source for multiresidue pesticide analysis in challenging fruit and vegetable matrices. Key objectives were to assess instrument detection limits (IDLs), long-sequence stability, mass accuracy, and practical robustness under routine operating conditions using QuEChERS extracts and matrix-matched standards containing a broad pesticide panel (≈180 analytes).

Methodology


Overview of experimental approach:
  • Sample preparation: QuEChERS extracts in acetonitrile prepared from diverse matrices (spinach, blueberry, kiwi, lemon, onion tops, rocket, radish salad) and matrix-matched calibration standards.
  • Analyte panel: A multiresidue mixture comprising ~180 pesticides; standard spike levels for IDL determination were 2.0 µg/kg for most compounds (cypermethrin and deltamethrin at 5 µg/kg).
  • Injection: 1 µL injection volume to reduce matrix load and enhance robustness.
  • IDL determination: Ten replicate injections (n=10) of matrix-matched standards; IDLs computed using one-tailed Student’s t-test (n-1 degrees of freedom) at 99% confidence (t = 2.821) and maximum acceptable peak area RSD ≤ 20%.
  • Long-term stability: More than 400 injections over ~4 weeks including repeated standards and diverse matrix extracts; routine consumable maintenance recorded (liner and septum changes, column trims) and no ion source cleaning performed during the sequence.

Used instrumentation


Instrument configuration and key acquisition settings used in the study:
  • Mass spectrometer: Thermo Scientific Orbitrap Exploris GC S with NeverVent Advanced Electron Ionization (AEI) source (source temperature reported at 280 °C).
  • GC: Thermo Scientific TRACE 1610 GC with iConnect Programmable Temperature Vaporizing (iConnect-PTV) injector and Thermo Scientific AI/AS 1610 liquid autosampler.
  • Column: TraceGOLD TG-5SILMS, 30 m × 0.25 mm i.d. × 0.25 µm (with 5 m integrated guard).
  • Autosampler and PTV details: 1 µL injection, PTV splitless transfer to 320 °C, helium carrier gas at 1.2 mL/min, autosampler washing cycles and parameters optimized to reduce carryover.
  • Acquisition: Full-scan HRAM, m/z 50–600, resolving power 60,000 (FWHM at m/z 200), emission current 25 µA; multiple lock masses used for high mass accuracy.
  • Data system: Chromeleon CDS for instrument control, processing, and reporting with regulatory-compliant workflows and reprocessing capabilities.

Main results and discussion


Performance highlights observed in the evaluation:
  • Detection capability: Calculated IDLs covered a wide range (reported range ~0.03–2.3 µg/kg). Overall, 94% of the compounds investigated were detectable below 0.5 µg/kg under the test conditions.
  • Example signal quality: Extracted ion chromatograms at calculated IDLs showed acceptable ion ratio deviations and sub-ppm mass errors for quantifier ions, supporting confident quantitation near the detection limit.
  • Mass accuracy and selectivity: Across the four-week study, mass error for representative pesticides remained below ~1.2 ppm, enabling selective component detection in complex matrices and reducing false positives from isobaric interferences.
  • Stability and robustness: More than 400 injections (standards and diverse matrices) produced consistent absolute peak areas with %RSD values generally within ±20% over the sequence for selected analytes. Routine maintenance was limited to replacing the inlet liner and septum and periodic column trimming; no ion source cleaning was required during the study period.
  • Throughput considerations: The combination of high-sensitivity ion source and full-scan HRAM acquisition allowed low injection volumes and extended sequence operation, supporting high-throughput laboratory environments.

Benefits and practical applications


Practical advantages demonstrated by the Orbitrap Exploris GC S with NeverVent AEI in pesticide residue testing:
  • Full-scan HRAM provides both quantitative performance for known target lists and the added flexibility for retrospective and suspect/non-target screening without prior method modification.
  • Enhanced ionization efficiency and a focused ion beam from the AEI source yield improved sensitivity at ultra-trace concentrations while enabling reduction of injected matrix load.
  • High mass accuracy and resolving power increase confidence in compound identification in complex food matrices, supporting compliance with regulatory requirements and avoiding misidentification from matrix interferences.
  • Operational robustness and reduced frequency of source cleaning contribute to higher instrument uptime and lower maintenance burden in routine testing labs.

Future trends and potential applications


Anticipated directions and opportunities for HRAM GC workflows in food safety analysis:
  • Broader adoption of full-scan HRAM GC in routine residue monitoring will enable integrated targeted and non-target workflows, improving surveillance for emerging contaminants and metabolites.
  • Development of comprehensive high-resolution spectral libraries and automated suspect-screening pipelines will increase the value of retrospective data mining.
  • Improved software-driven quality assurance (automated reprocessing, flagging, and regulatory reporting) will streamline compliance with evolving guidelines and speed result turnaround.
  • Method miniaturization and reduced sample consumption combined with enhanced sensitivity may permit more sustainable, higher-throughput workflows across contract and regulatory laboratories.

Conclusions


The study demonstrates that the Orbitrap Exploris GC S mass spectrometer equipped with the NeverVent AEI source delivers sensitive, accurate, and robust multiresidue pesticide analysis in difficult fruit and vegetable matrices. High-resolution full-scan acquisition at 60,000 resolving power provided low IDLs for the majority of a 180-pesticide panel, excellent mass accuracy (sub-ppm to ~1 ppm range), and stable quantitative response across extensive injection sequences. These characteristics support both routine targeted testing at trace levels and retrospective screening for unknowns, making the system well suited for high-throughput food safety laboratories.

References


  1. European Commission. Guidelines – Maximum Residue Levels. European Commission guidance on maximum residue levels for pesticides in food and feed.
  2. SANTE/12682/2019 (implemented 01/01/2020). Analytical quality control and method validation procedures for pesticide residues analysis in Food and Feed; referenced as SANTE 11312/2021 v2026 updates.
  3. Thermo Fisher Scientific. Technical Note 10494 – Practical Determination and Validation of Instrument Detection Limit of ISQ 7000 Single Quadrupole GC-MS with ExtractaBrite EI Ionization Source.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Quantitative performance for pesticide analysis with the Orbitrap Exploris GC S mass spectrometer
Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer
Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer
2026|Thermo Fisher Scientific|Brochures and specifications
Enhanced sensitivity for the quantitative analysis of pesticides using a GC Orbitrap MS
Confident analysis of ultra-trace pesticides residues in baby food using triple quadrupole GC-MS