GC/HRMS, GC/Orbitrap, GC/MSD, GC/MS/MS
IndustriesFood & Agriculture
ManufacturerThermo Fisher Scientific
Importance of the topic
The reliable detection and quantitation of pesticide residues in food is fundamental for consumer safety and regulatory compliance. Complex, highly pigmented food matrices such as spinach pose analytical challenges due to co-extracted pigments and matrix-induced ion suppression. Analytical approaches that combine low detection limits, high selectivity and the ability to screen for known and unknown contaminants in a single run are increasingly important for modern food-safety laboratories.Goals and overview of the study
This technical note evaluates the quantitative performance of the Thermo Scientific Orbitrap Exploris GC S mass spectrometer for multiresidue pesticide analysis in a highly pigmented matrix (spinach). The study aims to demonstrate sensitivity, mass accuracy, spectral fidelity, linearity and robustness when using HRAM (high-resolution accurate-mass) full-scan GC-MS for trace-level pesticide quantitation and simultaneous screening.Methodology and experimental design
The study used QuEChERS extracts of spinach that were matrix-matched and spiked with a 185‑compound pesticide mix. Key analytical features and workflows included:- GC-MS acquisition in EI full-scan mode (m/z 50–600) with HRAM at 60,000 FWHM (m/z 200) and sub-ppm nominal mass calibration routines.
- Matrix-matched calibration curves covering 0.5–200 µg/kg (some compounds 1–200 µg/kg), with triplicate injections at each level to assess linearity and precision.
- Confirmation criteria based on SANTE guidance: two ions per analyte with mass accuracy ≤5 ppm, ion ratio deviation within ±30% was used as supporting evidence, and retention time tolerance ±0.1 min.
- Use of a TRACE 1610 GC with an iConnect PTV injector (PTV splitless mode) and a TraceGOLD TG-5SilMS column (30 m × 0.25 mm × 0.25 µm with 5 m SafeGuard) to achieve robust chromatography for the complex extract.
- Chromeleon CDS for acquisition, data processing, automated workflows and compliant reporting, enabling streamlined review and retrospective reprocessing.
Instrumentation used
- Thermo Scientific Orbitrap Exploris GC S mass spectrometer with NeverVent Advanced Electron Ionization (AEI) source.
- Thermo Scientific TRACE 1610 GC equipped with iConnect-PTV injector and AI/AS 1610 autosampler.
- TraceGOLD TG-5SilMS analytical column with integrated 5 m SafeGuard.
- Chromeleon Chromatography Data System for acquisition and data management.
Key results and discussion
- Sensitivity and detection limits: Method detection capabilities were at or below 0.5 µg/kg for most pesticides; low-level IDLs were derived from replicate injections and confirmed in a separate application note.
- Linearity and precision: Matrix-matched calibration curves showed excellent linearity with R² > 0.993 for all targets. Response factor %RSD was <20% across calibration ranges and many compounds showed RF %RSD substantially lower than that threshold.
- Trueness and recovery: Calculated concentrations were within ±20% of expected values for system suitability; mean recoveries ranged approximately 96–104% across the compound set.
- Mass accuracy and spectral fidelity: Average mass errors for quantifier and qualifier ions were consistently <3 ppm across the calibration range and over extended acquisition periods. Spectra retained high fidelity from the lowest to the highest calibration levels, supporting robust identification and quantitation even at trace levels.
- Library identification and unknown screening: EI spectra produced high-quality matches when searched against HRAM spectral libraries; forward and reverse (R Match) scores were typically ≥800, enabling confident untargeted identification in addition to targeted quantitation.
- Chromatography and robustness: The PTV injector and high-inertness sample path produced Gaussian peak shapes (asymmetry 1.0–1.5) and minimized thermal degradation of thermolabile pesticides, contributing to reliable quantitation across analyte classes.
Benefits and practical applications of the method
- Single-run quantitation plus screening: HRAM full-scan acquisition enables simultaneous targeted quantitation and retrospective screening for non-target or emerging contaminants without re-injection.
- Improved selectivity: High resolving power and sub-ppm mass accuracy reduce false positives/negatives by discriminating isobaric and coeluting interferences typical of complex food extracts.
- Regulatory suitability: The method met SANTE criteria at the default MRL of 10 µg/kg, and more than 90% of compounds were confirmed at 1 µg/kg, indicating applicability for regulatory residue monitoring.
- Operational robustness: The NeverVent AEI source improves ionization efficiency and uptime, supporting high-throughput routine workflows in QA/QC and food-safety laboratories.
Future trends and potential applications
- Wider adoption of HRAM-GC-MS for routine monitoring as regulatory frameworks and residue lists expand, enabling broader analyte coverage without method re-optimization.
- Enhanced spectral libraries and cross-platform HRAM databases to support even greater confidence in untargeted screening and automated identification.
- Integration of retrospective data mining and machine-learning approaches for trend analysis, suspect screening and prediction of matrix interferences.
- Further optimization of sample introduction (PTV, micro-extraction) and data workflows to increase throughput while maintaining trace-level sensitivity for complex matrices.
- Continued harmonization with regulatory guidance to promote routine use of HRAM full-scan strategies in compliance testing.
Conclusion
The Orbitrap Exploris GC S MS equipped with the NeverVent AEI source demonstrates robust, sensitive and selective performance for multiresidue pesticide analysis in highly pigmented matrices. The combination of HRAM full-scan acquisition, reliable chromatography, strong spectral fidelity and stable mass accuracy supports accurate quantitation down to sub-µg/kg levels for most pesticides, while enabling retrospective screening and confident identification—features well suited to contemporary food-safety and regulatory laboratories.Reference
- European Commission. Guidelines — Maximum Residue Levels. European Commission guidance on pesticide residues (SANTE-related frameworks).
- SANTE/11312/2021 v2 (2020). Analytical quality control and method validation procedures for pesticide residues analysis in Food and Feed.
- Thermo Fisher Scientific. Technical Note 004383 — Enhanced sensitivity in multiresidue pesticide analysis with the Orbitrap Exploris GC S mass spectrometer.