HOW TO DETERMINE OLEFINS IN PYROLYSIS/HTL OILS FROM WASTE PLASTICS AND TIRES USING GC×GC-FID?

Posters | 2026 | REFINE LabInstrumentation
Pyrolysis, GCxGC
Industries
Energy & Chemicals
Manufacturer

Importance of the topic


The accurate determination of olefins in pyrolysis and hydrothermal liquefaction (HTL) oils derived from waste plastics and tires is critical because olefins typically form the dominant hydrocarbon fraction (>70 wt%) and strongly influence thermochemical stability, storage behaviour, and downstream valorization (e.g., refining, hydrotreatment). Reliable quantification of olefins informs process optimisation, catalyst selection, and product upgrading strategies and helps to assess the effectiveness of stabilization treatments such as hydrotreatment.

Objectives and study overview


This work presents a practical, low-cost analytical approach to selectively quantify olefins in complex pyrolysis/HTL oils using silver nitrate-impregnated silica (AgNO3/SiO2) solid-phase extraction (SPE) followed by comprehensive two-dimensional gas chromatography with flame ionization detection (GC×GC-FID). The method aims to provide an alternative to selective but expensive detectors (PI-TOF-MS, VUV), enabling laboratories without those instruments to determine olefin content with acceptable accuracy and precision for routine analysis and quality control.

Methodology


The approach uses selective adsorption of olefins on AgNO3-impregnated silica sorbent to separate olefinic species from other hydrocarbon classes, followed by comparative GC×GC-FID analysis of the sample before and after SPE. Key procedural points:
  • SPE format: 3 mL cartridge packed with ~1.45 ± 0.03 g (≈2.5 mL) freshly activated Ag-SiO2 (particle size 0.040–0.063 mm), sorbent densified and secured with a frit.
  • Sample load: ~15 µL of sample (≈12–14 mg) introduced to the cartridge.
  • Elution: Dichloromethane (DCM) applied in 0.6 mL fractions; the first 1.8 mL effluent collected into a 2 mL GC vial. Elution is performed by gravity only.
  • Analysis: GC×GC-FID chromatograms obtained for the original sample and for the SPE-treated effluent.
  • Quantification principle: Loss of peak area in the chromatographic regions corresponding to olefins (before vs. after SPE) is attributed to olefin adsorption; area changes are normalized using an internal standard (externally added or native).

The method workflow requires only ~50 µL of sample and approximately 15 minutes of sample preparation time per sample.

Used instrumentation


  • SPE cartridges packed with AgNO3-impregnated silica (Ag-SiO2).
  • Dichloromethane (DCM) for elution.
  • Comprehensive two-dimensional gas chromatography with flame ionization detection (GC×GC-FID) for chromatographic separation and quantification.
  • Standard laboratory glassware and 2 mL GC vials; no mass spectrometric detectors are required for routine quantification.

Main results and discussion


The AgNO3/SiO2–GC×GC-FID workflow was validated using model-spiked kerosene (JET-A1) and applied to a range of pyrolysis oils produced from sorted polyolefins (PP/PE), mixed waste plastics (MWP) with contaminants, and tire feedstocks. Principal findings include:
  • Quantitative performance: The method provides reliable quantification for samples with olefin contents above approximately 5 wt%. Accuracy and repeatability across major olefin classes were adequate for routine screening, with typical relative standard deviations (RSDs) in the low single-digit percent range for major components.
  • Validation outcomes: Spike-recovery experiments on kerosene spiked with olefinic model mixtures demonstrated close agreement between prepared and determined olefin mass fractions for mid-to-high concentration levels. Performance deteriorates for very low olefin contents (<5 wt%), where relative errors and variability increase.
  • Application to real oils: Pyrolysis oils from polyolefin-rich feedstocks and mixed wastes were found to contain substantial olefin fractions (tens of weight percent for aliphatic olefins and measurable fractions of styrenic/unsaturated aromatics in tire-derived oils). Reported RSDs for most quantified classes were generally <5%, indicating good method precision for typical pyrolysis matrices.
  • Process monitoring: The method successfully captured olefin depletion trends after hydrotreatment: increasing hydrotreatment temperature corresponded to progressive reduction of olefin content, demonstrating the method’s utility for assessing upgrading severity and conversion efficiency.
  • Limitations: Selectivity depends on the performance and activation state of Ag-SiO2 sorbent; certain highly substituted or sterically hindered olefins may show different adsorption behaviour and require careful calibration. Very low-level olefins (<5 wt%) fall below the reliable application range indicated by validation.

Benefits and practical applications


The main practical advantages of this method are:
  • Cost-effectiveness: Avoids the need for expensive selective detectors (VUV or PI-TOF-MS), making olefin quantification accessible to more laboratories.
  • Speed and sample economy: Minimal sample volume (~50 µL) and short sample prep time (~15 min) enable relatively high throughput for routine monitoring.
  • Simplicity: Straightforward SPE protocol with gravity elution and standard GC×GC-FID analysis; no MS spectra interpretation required.
  • Applicability: Suitable for process development labs, pilot plants, and quality-control environments monitoring olefin content in pyrolysis/HTL oils and evaluating treatment processes such as hydrotreatment.

Future trends and potential applications


Potential directions to expand and improve the method include:
  • Automation and throughput: Automating SPE packing and fraction collection to increase sample throughput and reproducibility for routine QC operations.
  • Sorbent optimisation: Investigating alternative silver-based sorbents, different impregnation levels, or tailored particle sizes to improve selectivity and recovery for a broader range of olefin structures.
  • Hybrid workflows: Combining the Ag-SiO2 SPE pretreatment with selective detectors (VUV, PI-TOF-MS) or targeted GC×GC–MS analyses for structural confirmation when identification of specific olefin isomers is required.
  • Quantitative extension: Developing calibration protocols and certified reference materials for olefin classes in pyrolysis matrices to strengthen traceability and comparability across laboratories.
  • Broader feedstock scope: Validating the approach for other complex liquefaction streams (biomass-derived oils, mixed municipal wastes) and for lower-olefin samples by refining sensitivity.

Conclusions


The AgNO3/SiO2 SPE combined with GC×GC-FID provides a practical, low-cost method for selective determination of olefins in pyrolysis and HTL oils when olefin content exceeds ~5 wt%. The procedure requires minimal sample volume and preparation time, delivers acceptable precision for major olefin fractions, and enables routine monitoring of feedstock and process-induced changes (e.g., hydrotreatment). While not a replacement for detailed structural identification achievable with MS-based detectors, this technique expands access to selective olefin quantification for laboratories constrained by instrumentation budgets.

References


  1. M. Auersvald, M. Šiman, P. Straka. Talanta 281 (2025) 126792. DOI: 10.1016/j.talanta.2024.126792.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Quantitative determination of olefins in pyrolysis oils from waste plastics and tires using selective adsorption by Ag–SiO2 followed by GC×GC-FID
A comparative study of aromatic content in pyrolysis oils from waste plastics and tires: Assessing common refinery methods
A comparative study of aromatic content in pyrolysis oils from waste plastics and tires: Assessing common refinery methods
COUNTING DOUBLE BONDS: GC×GC–FID FOR PLASTIC PYROLYSIS OILS