Consumables, LC columns, LC/MS
IndustriesFood & Agriculture
ManufacturerShimadzu
Importance of the topic
Accurate and reliable analysis of simple sugars (monosaccharides and common disaccharides) in fruit juice and soft drinks is essential for nutritional labeling, quality control, adulteration detection and process monitoring in the food industry. Hydrophilic interaction liquid chromatography (HILIC) coupled with mass spectrometric detection combines effective retention and separation of highly polar carbohydrates with high sensitivity and selectivity, making it a practical approach for routine food analysis.
Objectives and study overview
This application note documents an LC–MS method using Shim-pack Mix-HILIC for the separation and targeted detection of main sugars in fruit juice samples. The method aims to separate and quantify fructose, glucose and sucrose from commercial juice matrices with MS selective ion monitoring (SIM) detection, providing a workflow that is compatible with routine laboratory throughput and mass-spectrometric analysis.
Methodology
The method employs HILIC separation under a high-organic gradient and SIM-mode LC–MS for targeted detection of sugars. Key method features:
- Stationary phase: Shim-pack Mix-HILIC (150 mm × 2.1 mm I.D., 5 µm).
- Mobile phases: A = 40 mmol/L ammonium bicarbonate (aqueous), B = acetonitrile. Gradient: 75% B at 0 min → 60% B at 15 min → return to 75% B at 15.01–20 min.
- Flow rate: 0.3 mL/min; run time approximately 20 min.
- Sample solvent: 60% acetonitrile (to maintain HILIC retention and prevent peak distortion).
- Column temperature: 45 °C; injection volume: 1 µL.
- Simple sample preparation: filtration through a hydrophilic PTFE 0.22 µm syringe/vial filter and transfer into vials.
Used instrumentation
The method and measurements were carried out on Shimadzu equipment and consumables as specified below:
- LC system: Nexera X3.
- Column: Shim-pack Mix-HILIC, 150 mm × 2.1 mm, 5 µm (Shimadzu Shim-pack series).
- Vials/filters: Shim-vial S; TORAST Disc hydrophilic PTFE filter, 13 mm, 0.22 µm (P/N references given in source).
- MS ionization: Dual ionization (DUIS) ESI/APCI in negative mode, operated in SIM (selected ion monitoring).
- Typical MS parameters: interface voltage −2.0 kV; nebulizing gas 3.0 L/min; drying gas 5.0 L/min; heating gas 7.0 L/min; desolvation temperature 400 °C; DL temperature 150 °C.
Main results and discussion
The chromatograms reported in the application note show clear, baseline-resolved peaks for fructose, glucose and sucrose in two commercial fruit juice samples using the Shim-pack Mix-HILIC column and the described gradient. Run time of 20 minutes provides adequate separation while keeping cycle time suitable for routine analysis. SIM detection afforded sensitive, selective monitoring of the target sugar ions, producing robust signal levels for quantitation in the juice matrices.
Method considerations and interpretation:
- HILIC retention is driven by the high organic mobile phase; maintaining sample solvent at 60% acetonitrile reduces peak broadening and preserves retention behavior.
- Ammonium bicarbonate (40 mmol/L) provides volatile buffer capacity compatible with MS and helps stabilize sugar ionization/adduct formation.
- Negative DUIS detection in SIM mode is effective for targeted sugar analysis; however, sugar ionization can depend on ion-source conditions and possible adducts (e.g., formate, bicarbonate, chloride) — method tuning may be required for different instruments or sample types.
- Filtration through hydrophilic PTFE and minimal sample preparation reduces matrix interferences and instrument contamination risk, but more complex matrices may require dilution or additional cleanup.
Benefits and practical applications
This HILIC–LC–MS protocol offers several practical benefits for food laboratories:
- Good retention and resolution for polar sugars without derivatization.
- MS compatibility via a volatile buffer (ammonium bicarbonate) enabling selective and sensitive SIM or MS/MS quantitation.
- Short, single-run workflow (~20 min) suitable for routine QC and comparative analyses of soft drinks and juices.
- Minimal sample preparation (filtration) accelerates throughput and reduces consumables and reagent use.
Future trends and possibilities
Potential developments and extensions of this approach include:
- Switching from SIM to MS/MS or high-resolution accurate-mass (HRAM) detection for improved specificity, multi-component screening and confirmation of sugar isomers or related oligosaccharides.
- Use of isotope-labeled internal standards or standard addition for improved absolute quantitation and matrix effect compensation.
- Method miniaturization and faster columns (shorter lengths, sub-2 µm particles) to reduce run time while maintaining resolution.
- Integration with automated sample prep (solid-phase extraction, online dilution) for high-throughput labs.
- Application to broader carbohydrate panels (sugar alcohols, oligosaccharides) and coupling with orthogonal techniques (ion chromatography, derivatization-LC) when needed.
Conclusion
The Shim-pack Mix-HILIC LC–MS method provides an effective, MS-compatible workflow for the separation and targeted detection of fructose, glucose and sucrose in fruit juice matrices. Its combination of HILIC retention, volatile buffering and SIM detection yields a practical balance of sensitivity, selectivity and throughput appropriate for routine food-analysis laboratories. Careful control of sample solvent composition, source conditions and potential adduct chemistry is recommended to ensure reproducible quantitation across instruments and sample types.
References
- Shimadzu Corporation, Application note ERAS-1000-0654: Analysis of Monosaccharides in Fruit Juice using Shim-pack Mix-HILIC and LC–MS, First Edition Sep. 2026. (Instrument and consumable part numbers provided in the original document.)
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.