Analysis of Carcinogenic Nitrosamines at Ultra-trace Levels Among Terbinafine Using LC/TQ

Posters | 2025 | Agilent Technologies | ASMSInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies

Significance of the Topic


The emergence of genotoxic N-nitroso impurities in pharmaceutical products has prompted regulatory recalls by the FDA, underscoring the need for ultra-trace analysis of nitrosamine drug substance-related impurities (NDSRIs). Terbinafine, an amino-containing antifungal agent, may harbor N-nitroso derivatives that pose health risks even at sub-ppm levels. Reliable detection and quantitation of these impurities ensure product safety and compliance with stringent regulatory limits.

Objectives and Study Overview


This study aimed to develop and validate a highly sensitive LC-MS/MS method for the simultaneous determination of three NDSRIs in terbinafine: N-nitroso-desmethyl terbinafine (NDT), N-nitroso terbinafine impurity A (NTA), and N-nitroso terbinafine degradant (NTD). Key goals included achieving a limit of quantitation (LOQ) below 0.007 ppm, demonstrating linearity, precision, and recovery in both API and tablet formulations.

Methodology and Instrumentation


Sample Preparation
A 37 mg terbinafine powder sample was dissolved in 1 mL ultrapure water, vortex-mixed for 10 minutes, and centrifuged at 10 000 rpm. The clear supernatant was injected directly.

Chromatographic Conditions
  • Instrument: Agilent 1290 Infinity II UHPLC
  • Column: HPH-C18, 3 × 150 mm, 2.7 µm, 30 °C
  • Mobile Phase: (A) 5 mM ammonium acetate in water; (B) methanol
  • Gradient: 2% B (0–5 min), 50% B (5–10 min), 75% B (10–25 min), 90% B (25–38 min), re-equilibration to 2% B
  • Flow Diversion: Terbinafine diverted to waste via an integrated valve to protect the mass spectrometer

Mass Spectrometry
  • Instrument: Agilent 6495C Triple Quadrupole MS
  • Ionization: ESI in positive mode
  • Drying Gas: 250 °C, 15 L/min; Sheath Gas: 350 °C, 9 L/min; Nebulizer: 45 psi
  • Capillary/Nozzle Voltage: 2000 V; Dynamic MRM with narrow resolution
  • UV Detection: 220 nm (for monitoring terbinafine diversion)

Results and Discussion


Chromatographic Separation
Efficient baseline separation of NDT, NTA, and NTD was achieved, while the high-concentration terbinafine peak was diverted to waste, preventing contamination of the MS source.

Sensitivity and Linearity
Calibration was linear over 0.125–2 ng/mL with correlation coefficients (r²) ≥ 0.997 for all analytes. Signal-to-noise ratios at the LOQ (0.25 ng/mL) exceeded 15:1 for all NDSRIs.

Precision and Recovery
  • Precision (RSD) at LOQ: NTD (6.8%), NTA (6.6%), NDT (5.1%)
  • Recovery in API: 96.1–99.2%; in tablet: 93.3–108.1%
The results confirm robustness, reproducibility, and accuracy at the target sub-ppb concentration (0.007 ppm).

Benefits and Practical Applications


  • Regulatory Compliance: Meets stringent guidelines for nitrosamine impurities.
  • High Sensitivity: Enables detection at sub-ppb levels.
  • Operational Robustness: Efficient diversion of matrix-rich drug peak protects MS system.
  • Versatility: Applicable to both API and finished dosage forms.

Future Trends and Opportunities


Advancements may include:
  • Automation of sample preparation and data processing.
  • Integration with high-resolution MS for broader impurity profiling.
  • Adaptation to other nitrogen-containing APIs prone to nitrosamine formation.
  • Real-time monitoring in continuous manufacturing environments.

Conclusion


  • A robust LC-MS/MS method for three terbinafine NDSRIs was established.
  • LOQ of 0.25 ng/mL (0.007 ppm) with excellent linearity (r² ≥ 0.997).
  • Precision at LOQ: RSD < 7%; recovery within 90–110% in API and tablets.
  • Method supports stringent safety testing in pharmaceutical quality control.

Reference


  • Low-Level Quantitation of N-Nitroso Dabigatran Etexilate Impurity in Dabigatran Etexilate Mesylate API Using the Agilent 6495C. Agilent Application Note, 5994-7066EN.
  • Nitrosamine Impurities Application Guide. Agilent Application Guide, 5994-2393EN.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Low-Level Quantitation of N-Nitroso Dabigatran Etexilate Impurity in Dabigatran Etexilate Mesylate API Using the Agilent 6495C LC/TQ
Sensitive and Selective Quantitation of N- Nitroso Dabigatran Etexilate Impurity in Dabigatran Etexilate API using the Agilent 6495C LC/TQ
Highly Sensitive Quantitation of N-Nitroso Timolol Impurity in Timolol API using the Agilent 6495D LC/TQ
Quantitation of N-Nitroso Sitagliptin Impurity (NTTP) in Sitagliptin and Metformin Combination Drug Product Using the Agilent 6475 LC/TQ