LC/MS, LC/MS/MS, LC/QQQ
IndustriesEnvironmental
ManufacturerAgilent Technologies
Significance of the Topic
Per- and polyfluoroalkyl substances (PFAS), including perfluoro carboxylic acids (PFCAs) and their fluorotelomer alcohol (FTOH) precursors, are widely used for water- and stain-resistant applications. Their persistence in the environment and potential health risks have spurred demand for reliable analytical methods. However, strong organic solvents such as methanol, commonly used for sample dissolution, can lead to severe solvent effects—distorted peak shapes and compromised quantitation—especially for early-eluting, highly polar analytes like PFBA.
Objectives and Study Overview
This study aimed to develop a high-throughput, robust LC-MS/MS method for simultaneous quantification of 50 PFAS (16 PFCAs, 5 FTOHs, and 29 related compounds) by mitigating solvent effects. By integrating Agilent’s Hybrid Multisampler in Feed Injection mode with a triple quadrupole mass spectrometer, the method targets enhanced peak shapes, sensitivity, and reproducibility without reducing injection volume or additional sample dilution steps.
Methodology and Instrumentation
- Sample Preparation: Standards prepared in methanol following BS EN ISO 23702-1:2023 dilution protocols; matrix-matched calibration at 0.1–10 µg/L for PFCAs and 5–100 µg/L for FTOHs.
- Chromatographic Conditions: Agilent UHPLC 1290; ZORBAX RRHD Eclipse Plus C18 (3.0×100 mm, 1.8 µm); column temperature 40 °C; flow rate 0.4 mL/min; gradient from 10 % to 100 % acetonitrile over 19 min; injection volumes optimized at 20 µL.
- Feed Injection Mode: Continuous sample infusion at 5 % pump flow to dilute solvent strength in the mobile phase, reducing distortion for PFBA and other early-eluting analytes.
- Mass Spectrometry: Agilent 6475 triple quadrupole in AJS electrospray negative mode; gas temperature 150 °C; gas flow 9 L/min; sheath gas temperature 250 °C; sheath flow 12 L/min; capillary voltage 3500 V; nozzle voltage 1500 V.
Used Instrumentation
- Agilent 1290 Infinity II UHPLC system
- ZORBAX RRHD Eclipse Plus C18 column, 3.0×100 mm, 1.8 µm
- Agilent Hybrid Multisampler with Feed Injection capability
- Agilent 6475 Triple Quadrupole Mass Spectrometer
Main Results and Discussion
- Pump Flow Optimization: 5 % feed speed yielded the sharpest PFBA peak and minimal solvent effects, with a slight retention time shift.
- Injection Volume: 20 µL delivered optimal peak shapes; 40 µL induced notable distortion for PFBA.
- Throughput: Separation of 50 PFAS completed within 19 min under optimized gradient.
- Linearity and Repeatability: All analytes exhibited r² ≥ 0.99; retention time RSD < 0.1 % and peak area RSD < 10 % (n=6) in matrix-spiked samples.
Benefits and Practical Applications
- Mitigation of solvent effects without lowering injection volume or re-dilution.
- Enhanced sensitivity and reproducibility for polar and early-eluting PFAS.
- High throughput suitable for routine environmental, consumer product, and quality control laboratories.
Future Trends and Applications
- Automation integration for large-scale PFAS monitoring in water, soil, and food matrices.
- Extension of Feed Injection strategies to other polar analytes and emerging contaminants.
- Development of field-deployable sampling devices leveraging miniaturized feed-injection modules.
Conclusion
Employing Feed Injection mode on Agilent’s Hybrid Multisampler effectively addresses solvent effects in PFAS analysis, enabling high-throughput, sensitive, and reproducible quantification of a broad panel of PFCAs, FTOHs, and related compounds within a 19-minute run time.
References
- BS EN ISO 23702-1:2023 Leather—Per- and polyfluoroalkyl substances Part 1: Determination of non-volatile compounds by extraction method using liquid chromatography.