COMPARISON OF THE CONTENT OF VOLATILE SULPHUR-CONTAINING COMPOUNDS IN SLOVAK BEERS BY THE SPME METHOD

Scientific articles | 2008 | Kvasny PrumyslInstrumentation
GC, SPME
Industries
Food & Agriculture
Manufacturer
Agilent Technologies, CTC Analytics

Significance of the Topic


Beer aroma is shaped by over 600 volatile compounds, among which sulphur-containing molecules play a critical role due to their low sensory thresholds and potent odours. Even trace levels can impart off-flavours, affecting consumer perception and product quality.

Objectives and Overview of the Study


The study aimed to develop and optimize a solid-phase microextraction (SPME) method coupled with gas chromatography-flame photometric detection (GC-FPD) for profiling volatile sulphur compounds in Slovak beers. Twenty-three samples, including draft and lager styles from eight breweries, were compared to assess brand and gravity influences.

Methodology


Samples (6 ml) were spiked with internal standards (ethyl methyl sulphide, propyl thioacetate) and salted (0.6 g NaCl). Headspace SPME extraction was conducted at 45 °C for 30 min using a DVB/Carboxen/PDMS fibre. Desorption occurred in splitless mode at 270 °C. The GC temperature program ran from 35 °C to 280 °C, and FPD selectively detected sulphur species.

Instrumentation Used


  • Agilent 6890A GC with pulsed flame photometric detector (FPD)
  • Combi PAL autosampler in SPME mode (CTC Analytics)
  • SPME fibre: 50/30 μm DVB/Carboxen/PDMS Stableflex (Supelco)
  • Column: VF-5MS 30 m × 0.25 mm ID, 1.0 μm film (Varian)

Main Results and Discussion


Optimisation revealed that 6 ml sample volume with 10 % NaCl, 45 °C and 30 min extraction maximized recovery of both low- and high-boiling sulphur compounds. Key analytes identified included methanethiol, dimethyl sulphide, dimethyl disulphide, methyl/ethyl thioacetates, methionol and 3-methylthiopropyl acetate. Quantitative differences were observed across brands and beer gravities, though none exceeded sensory off-flavour thresholds under optimized conditions.

Benefits and Practical Applications


The optimized SPME-GC-FPD method offers rapid, solvent-free preconcentration with high sensitivity for routine quality control in breweries. It enables comparative profiling of sulphur off-flavour precursors, aiding process monitoring and raw material evaluation.

Future Trends and Potential Applications


  • Integration of tandem MS or chemiluminescence detectors for improved selectivity
  • Automation and high-throughput screening in production environments
  • Extension to other fermented beverages (wines, ciders)
  • Data-driven flavour modelling and sensory correlation studies

Conclusion


A robust SPME-GC-FPD protocol was established for sensitive, reproducible analysis of volatile sulphur compounds in beer. The approach supports flavour stability assessment and guides brewery quality assurance efforts.

References


  1. Hill PG, Smith RM. J Chromatogr A. 2000;872:203–211.
  2. Fang Y, Qian MC. J Chromatogr A. 2005;1080:177–185.
  3. Vanderhaegen B et al. Food Chem. 2006;95:357–381.
  4. Vermeulen C et al. J Agric Food Chem. 2006;54:5001–5068.
  5. Pinho O et al. J Chromatogr A. 2006;1121:145–153.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

DETERMINATION OF THE FATTY ACIDS IN BEER BY SPME
DETERMINATION OF THE FATTY ACIDS IN BEER BY SPME
2005|Thermo Fisher Scientific|Scientific articles
APPLICATION OF SOME MODERN SAMPLE PREPARATION PROCEDURES FOR QUANTITATIVE DETERMINATION OF VICINAL DIKETONES IN BEER
DETERMINATION OF METHIONINE IN MALT
DETERMINATION OF METHIONINE IN MALT
2009|Thermo Fisher Scientific|Scientific articles
Optimization of Determination of Dimethyl Sulfide in Wort and Beer
Optimization of Determination of Dimethyl Sulfide in Wort and Beer
2017|Thermo Fisher Scientific|Scientific articles