LC/MS, LC/MS/MS, LC/HRMS, LC/TOF
IndustriesPharma & Biopharma
ManufacturerAgilent Technologies
Significance of the Topic
Biopharmaceutical formulations commonly include polysorbate surfactants to stabilize protein therapeutics. Oxidative degradation of these surfactants—driven by residual peroxides, light exposure, and trace metals—poses a significant risk to product stability, efficacy, and safety. Rapid and unambiguous screening methods for polysorbate oxidation by-products are essential for developing mitigation strategies and ensuring long-term formulation integrity.
Objectives and Overview of the Study
This application note describes a comprehensive LC/MS/MS workflow to screen and identify oxidative degradation products of polysorbate 80. Key goals include:
- Developing a targeted screening approach using a custom suspect library and the Find by Formula algorithm.
- Implementing an untargeted suspect screening strategy with PolyMatch software to discover novel oxidation markers.
- Demonstrating the performance of an Agilent AdvanceBio Surfactant Profiling HPLC column coupled to an Agilent 6546 LC/Q-TOF system for high-resolution separations and accurate mass measurements.
Methodology and Instrumentation
Sample Preparation
- Polysorbate 80 diluted to 0.02% w/v in 10 mM ammonium acetate (mobile phase A).
- Oxidation induced by incubating with 3% hydrogen peroxide at 40 °C overnight.
Chromatography
- Agilent AdvanceBio Surfactant Profiling 300 Å, 3.5 μm, 2.1 × 100 mm column with guard.
- Mobile phase A: 10 mM ammonium acetate; B: methanol.
- Gradient from 5% B to 91% B over 10 minutes, total run time 18 minutes; flow rate 0.3 mL/min; column temperature 50 °C; injection volume 2 μL.
Mass Spectrometry
- Agilent 6546 LC/Q-TOF with Dual Agilent Jet Stream source.
- Gas temperature 200 °C; nebulizer 30 psig; sheath gas 350 °C at 11 L/min; capillary voltage 3,500 V; fragmentor 145 V.
- Data acquired in Auto MS/MS (top five precursors per cycle) at 5 spectra/sec for MS and 3 spectra/sec for MS/MS; mass range m/z 50–3,000; collision energies at 10, 20, 40 V.
Data Processing
- Targeted screening with Agilent MassHunter Qualitative Analysis 12.0 using the Find by Formula (FBF) algorithm and a Personal Compound Database and Library (PCDL) of >4,800 known polysorbate degradants.
- Suspect screening and untargeted polymer annotation with Agilent PolyMatch Flow and Visualizer software, leveraging Kendrick mass defect plots, homologous series detection, and MS/MS fragment screening.
Main Results and Discussion
Targeted Analysis
- POE1 monohydroperoxy oleate identified with FBF scores of 82.5 in control and 99.5 in oxidized samples; abundance increased by an order of magnitude upon oxidation.
- Series of POE11–POE15 monohydroperoxy oleates showed high FBF scores (96.3–99.7) and sub-1 ppm mass errors; levels increased consistently in oxidized samples with peak area RSDs below 6%.
Suspect Screening Analysis
- PolyMatch detected 261 high-confidence (A-score) polymer species, including oxidized polysorbate isomers separated chromatographically.
- Kendrick mass defect plots normalized to CH2CH2O and retention time versus m/z projections facilitated clear visualization of homologous oxidation series.
- MS/MS spectra confirmed diagnostic ions (e.g., m/z 323.2576 for keto oleic acid) that guided future library expansion.
Benefits and Practical Applications of the Method
- Rapid screening workflow enabling both targeted identification of known degradants and discovery of novel oxidation products.
- High-resolution separations achieved in under 20 minutes, improving throughput.
- Accurate mass and isotopic fidelity support unambiguous compound identification.
- Software tools streamline data processing: FBF for fast targeted analysis, PolyMatch for comprehensive polymer annotation.
Future Trends and Applications
- Expansion of PCDL and PolyMatch libraries to include newly discovered oxidation markers and fragmentation patterns.
- Integration of iterative exclusion MS/MS strategies to increase coverage of low-abundance species.
- Application of the workflow to other surfactants and complex polymer formulations.
- Use of advanced data analytics and machine learning to predict and mitigate oxidation pathways in real time.
Conclusion
The presented LC/MS/MS screening workflow combines the high-resolution separations of the Agilent AdvanceBio Surfactant Profiling column with targeted and untargeted data analysis strategies. This approach effectively profiles polysorbate oxidation by-products, offering robust tools for formulation monitoring and mitigation strategy development in biopharmaceutical research.
Reference
1. Li X et al. Anal. Chem. 2023;95:9156–9163.
2. Kranz W et al. J. Pharm. Sci. 2019;108:2022–2032.
3. Borisov OV et al. J. Pharm. Sci. 2015;104:1005–1018.
4. Hvattum E et al. J. Pharm. Biomed. Anal. 2012;62:7–16.
5. Liu H et al. J. Pharm. Sci. 2022;111:323–334.
6. Kishore RSK et al. J. Pharm. Sci. 2011;100:721–731.
7. Brovc EV et al. Antioxidants 2020;9:441.
8. Koelmel JP et al. J. Am. Soc. Mass Spectrom. 2024;35:413–420.