Quantitation of N-nitroso propranolol (NOP) in Propranolol formulation and its placebo using LC-MS/MS

Posters | 2023 | Shimadzu | ASMSInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Pharma & Biopharma
Manufacturer
Shimadzu

Significance of the Topic


The presence of N-nitrosamine impurities in pharmaceutical products poses a critical safety concern due to their potent carcinogenic potential. Regulatory agencies have recalled several drug products after detecting low-molecular-weight N,N-dialkylnitrosamines. N-nitroso propranolol (NOP), a genotoxic impurity related to propranolol API, requires highly sensitive analytical methods for trace-level detection to ensure drug safety and compliance.

Objectives and Study Overview


This study aims to develop and validate a robust LC-MS/MS assay for quantifying NOP in propranolol formulations and placebo matrices. The method targets trace concentrations down to single-digit parts per billion, offering reliable quantitation for both quality control and regulatory monitoring.

Methodology


A targeted multiple reaction monitoring (MRM) approach was optimized for NOP on a Shimadzu UHPLC coupled to a triple quadrupole mass spectrometer. Chromatographic separation employed a C8 column with a gradient of 0.1 % formic acid in water and methanol at 0.5 mL/min. Linearity was established over 10–4000 ppb with a coefficient of determination (r2) of 0.999. Detection and quantitation limits were determined as 2 ppb and 10 ppb, respectively, based on signal-to-noise criteria. Sample preparation involved weighing 25 mg of API, formulation or placebo, spiking with NOP standard, dilution, centrifugation and filtration prior to a 30 µL injection.

Použitá instrumentace


  • UHPLC system: Shimadzu Nexera X3
  • Mass spectrometer: Shimadzu LCMS-8060NX triple quadrupole
  • Ionization: APCI; Interface 350 °C; Desolvation line 200 °C; Heater block 200 °C
  • Column: Shim-pack Scepter C8-120, 3 × 150 mm, 5 µm
  • Mobile phases: 0.1 % formic acid in water (A) and methanol (B)
  • Gradient: 50–100 % B over 3–8 min, hold at 100 % until 13 min, re-equilibration to 50 % by 16 min
  • Injection volume: 30 µL

Main Results and Discussion


Calibration yielded a linear equation y = 36769.33 x + 428.53 (r2 = 0.999). Method precision at LOQ showed %RSD below 8 %. Analysis of two formulation lots revealed high NOP levels (3967.9 and 3640.6 ppb), while placebo samples contained 9.0 and 16.3 ppb. Recovery experiments in placebo at 10 ppb returned 104 % and at 20 ppb 119 %, demonstrating method accuracy and robustness.

Benefits and Practical Applications


  • Enables trace-level monitoring of carcinogenic nitrosamines in pharmaceuticals
  • Supports quality assurance and regulatory compliance
  • Applicable to both formulation and placebo matrices for routine screening

Future Trends and Potential Uses


Advancements in detector sensitivity and noise reduction will further lower detection limits. Integration with high-throughput and in-line process analytical technologies is expected. Expansion of this approach to other nitrosamine-related impurities and diverse drug matrices will broaden its applicability.

Conclusion


The validated LC-MS/MS method provides a sensitive, precise and accurate tool for quantifying N-nitroso propranolol at trace levels. Its performance characteristics meet stringent regulatory requirements, making it ideal for pharmaceutical safety surveillance.

Reference


  • [1] King FJ, Searle A, Urquhart MW. Process Res. Dev. 2020, 24(12), 2915–292.
  • [2] Partani P. et al. J Pharm Biomed Anal. 2009, 50(5), 966–976.
  • [3] Robbiano L, Martelli A, Allavena A, Mazzei M, Gazzaniga GM, Brambilla G. Cancer Res. 1991, 51(9), 2273–2279.
  • [4] Brambilla G, Martelli A, Sottofattori E. Carcinogenesis. 1995, 16(5), 1239–1242.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Quantitation of Nitroso-Propranolol in Propranolol HCl Formulation and its Placebo Using LC-MS/MS
Simultaneous quantitation of N-nitrosamines and NDSRI in API and formulation by using a DUIS ionization source in LC-MS/MS
Simultaneous Quantitation of N-nitrosamines and NDSRI in API and Formulation by Using a DUISTM Ionization Source in LC-MS/MS
Highly Sensitive Quantification of Mutagenic NDSRI N-Nitroso Propranolol in Propranolol API and 40 mg Tablets Using LC/MS/MS