A TMTpro 18plex Proteomics Standard for Assessing Protein Measurement Accuracy and Precision

Posters | 2021 | Thermo Fisher Scientific | ASMSInstrumentation
Consumables, LC/HRMS, LC/MS, LC/MS/MS, LC/Orbitrap
Industries
Proteomics
Manufacturer
Thermo Fisher Scientific

Significance of the Topic


The development of a TMTpro 18plex yeast digest standard addresses critical challenges in multiplexed proteomics by providing a tool to evaluate and optimize quantification accuracy and precision across diverse LC–MS acquisition strategies.

Objectives and Study Overview


  • Design and validate a prototype TMTpro 18plex standard using yeast strains with defined knockout proteins.
  • Compare the performance of high-resolution MS2, SPS-MS3 and real-time search (RTS) MS3 acquisition modes.
  • Assess the impact of FAIMS separation on measurement fidelity.

Used Instrumentation


  • Thermo Scientific Orbitrap Eclipse Tribrid mass spectrometer
  • Thermo Scientific Exploris 480 mass spectrometer
  • FAIMS Pro interface
  • Thermo Scientific EASY-Spray C18 50 cm column
  • Thermo Scientific UltiMate 3000 RSLCnano UHPLC system
  • Thermo Scientific EASY-nLC 1200 HPLC system

Methodology


  • Sample Preparation: A parental Saccharomyces cerevisiae strain (BY4741) and three knockout lines lacking Met6, His4 or Ura2 were grown to OD600 ≈ 3.0. Cells were lysed by bead beating, proteins reduced, alkylated and digested with LysC and trypsin, followed by C18 SPE cleanup.
  • TMTpro Labeling: Peptides from each knockout strain were labeled in triplicate and parental peptides in duplicate with distinct TMTpro 18plex tags. Samples were mixed at fixed ratios (100%, 50%, 25%, 12.5%, 6.25% and 0%).
  • LC Separation: Peptides were separated on a 50 cm EASY-Spray C18 column using gradients (4–28% or 8–32% acetonitrile with 0.1% formic acid) over 50 or 120 minutes.
  • Acquisition Methods: Data were acquired on Orbitrap Eclipse and Exploris 480 instruments using three strategies:
    • High-resolution MS2 (hrMS2)
    • Synchronous precursor selection MS3 (SPS-MS3)
    • Real-time search MS3 (RTS-MS3) with varied close-out thresholds.
    FAIMS Pro was applied in selected experiments to reduce co-isolation interference.
  • Data Analysis: Raw files processed in Proteome Discoverer 2.5/3.0 with SEQUEST HT or COMET search engines, 1% FDR, 10 ppm MS1 and 0.02/0.6 Da MS2 tolerance, and co-isolation filters of 50% (MS2) or 70% (MS3).

Results and Discussion


  • SPS-MS3 methods significantly improved quantification precision relative to hrMS2 by mitigating ratio compression from co-isolated peptides.
  • Inclusion of FAIMS Pro further enhanced measurement accuracy for both MS2 and MS3 workflows by reducing background interference.
  • RTS-MS3 acquisition achieved the highest proteome coverage, with a 26% increase in quantified protein groups over SPS-MS3 due to real-time peptide identification and targeted MS3 scans.
  • Standard curves for Met6, His4 and Ura2 displayed linearity >0.98 across four highest dilutions, with slight decrease at the lowest concentration; RTS-MS3 reduced background in knockout channels to undetectable levels.

Benefits and Practical Applications


  • The TMTpro 18plex yeast standard provides a reproducible benchmark for optimizing multiplex proteomic workflows and instrument performance.
  • Advanced acquisition schemes such as RTS-MS3 enable deeper proteome coverage and more accurate quantification, supporting rigorous QA/QC in research and industrial laboratories.

Future Trends and Potential Applications


  • Integration of real-time search strategies with machine-learning–driven decision making may further enhance dynamic acquisition efficiency.
  • Adoption of multiplexed standards in clinical and large-cohort studies will improve cross-laboratory data comparability and reliability.
  • Extension of standardized mixtures to other organisms or cell types and higher-plex reagents will broaden applicability in diverse proteomic investigations.

Conclusion


The TMTpro 18plex yeast digest standard demonstrates a versatile and robust approach to assess and refine multiplexed proteomic assays. SPS-MS3 and particularly RTS-MS3 workflows deliver superior accuracy, precision and proteome depth, facilitating optimized experimental designs and reliable quantitative results.

Reference


  1. Paulo JA, O’Connell JD, Gygi SP. A triple knockout (TKO) proteomics standard for diagnosing ion interference in isobaric labeling experiments. Journal of the American Society for Mass Spectrometry. 2016;27(10):1620–1625.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

A TMTpro 18plex Proteomics Standard for Assessing Protein Measurement Accuracy and Precision
Implementing Comet search engine into Proteome Discoverer to improve TMT Real-Time Search data processing
Real-Time Search enables a new gold standard for TMT quantitation accuracy on the Orbitrap Eclipse Tribrid mass spectrometer
A Modified Orbitrap Tribrid Mass Spectrometer with Real Time Search and Advanced Spectral Processing Enhances Multiplexed Proteome Coverage and Quantification Accuracy