Consumables, HPLC, LC columns
IndustriesProteomics
ManufacturerWaters
Significance of the Topic
Protein quantification in therapeutic formulations is essential for drug product quality control and research. Reversed-phase liquid chromatography can overcome limitations of UV absorbance methods when buffers or excipients interfere with measurements.
Objectives and Study Overview
This study evaluates a BioResolve RP mAb Polyphenyl column for quantifying monoclonal antibodies and fusion proteins in histidine buffer containing recombinant human serum albumin (rHSA). Targets include rituximab, interleukin-12, erythropoietin and abatacept.
Methodology and Instrumentation
Separations were performed on a Waters ACQUITY UPLC H-Class Bio system using a BioResolve RP mAb Polyphenyl column (2.1×50 mm, 2.7 µm, 450 Å) with UV detection at 280 nm. Mobile phases were 0.1% trifluoroacetic acid in water (A) and acetonitrile (B). Gradient profiles, flow rates (0.2–1.8 mL/min) and column temperatures (60–80 °C) were optimized per analyte. Data were acquired with Empower 3 software.
Results and Discussion
Proteins were baseline or well resolved from histidine and rHSA in under 10 minutes with resolutions ≥3.5. Calibration curves and linear dynamic ranges:
- Interleukin-12: 0.01–2 µg at 0.3 mL/min (error <8%); 0.1–5 µg at 1.8 mL/min (error <10%).
- Erythropoietin: 0.5–10 µg (error <3%).
- Abatacept: 0.2–5 µg (error <3%).
- Rituximab: 0.1–5 µg at 80 °C (error <12%).
Injection-to-injection carryover was ≤0.2%, ensuring reliable high-throughput quantification.
Benefits and Practical Applications
This approach delivers rapid, high-resolution separation of therapeutic proteins in complex buffers without interference from formulation excipients. The low carryover and wide dynamic ranges support accurate quantification in QA/QC and research laboratories.
Future Trends and Applications
Emerging directions include coupling with mass spectrometry for enhanced specificity, development of affinity-based stationary phases for selective capture, and further miniaturization to increase throughput and reduce solvent consumption.
Conclusion
The BioResolve RP mAb Polyphenyl column offers a robust, efficient platform for quantifying protein drugs in histidine-based formulations containing rHSA, overcoming the limitations of direct UV absorbance methods.
References
- Nguyen JM et al. Designing a new particle technology for reversed-phase separations of proteins. Waters Application Note 720006168EN, 2018.
- Nguyen JM et al. A Novel Phenyl-Based RPLC Stationary Phase for High Throughput, High Resolution Characterization of Protein Therapeutics. Waters Application Note 720006169EN, 2018.
- Zbacnik TJ et al. Role of buffers in protein formulations. Journal of Pharmaceutical Sciences. 2017;106:713–733.
- Human Serum Albumin as a Pharmaceutical Excipient. Drug Development Web resource.
- Hamza T et al. Interleukin 12, a key immunoregulatory cytokine in infection applications. International Journal of Molecular Sciences. 2010;11:789–806.
- Erythropoietin. Wikipedia entry.
- Abatacept. Wikipedia entry.
- Rituximab. Wikipedia entry.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.