Consumables, HPLC, LC columns
IndustriesPharma & Biopharma
ManufacturerWaters
Importance of the Topic
Monoclonal antibodies are critical biotherapeutics with complex structures that require precise analytical characterization to ensure safety and efficacy.
Reversed-phase liquid chromatography (RPLC) offers high separation efficiency but intact mAb analysis often suffers from limited recovery and sample degradation at high temperatures and with strong ion-pairing reagents.
Objectives and Study Overview
This study aimed to demonstrate the improved recovery of intact antibodies across a range of temperatures using a silica-based, solid-core BioResolve RP mAb Polyphenyl Column.
A direct comparison was made against a conventional polymeric column to assess recovery, resolution, and applicability for multiple therapeutic mAbs.
Methodology and Instrumentation
- Instrument: ACQUITY UPLC H-Class Bio System with ACQUITY TUV Detector (λ=280 nm) and Empower 3 Software
- Columns:
- BioResolve RP mAb Polyphenyl, 2.1×50 mm, 2.7 µm, 450 Å solid-core silica
- Polymeric PS-DVB, 2.1×50 mm, 1500 Å porous particles
- Mobile phases:
- Water and acetonitrile with either 0.1% TFA
- Water and acetonitrile with 0.1% DFA and 5% n-butanol
- Temperature range: 40–80 °C
- Sample panel: Waters Intact mAb Mass Check Standard, Rituximab, Tocilizumab, Bevacizumab (2.5 µg load)
Main Results and Discussion
Operation at high temperature (80 °C) with 0.1% TFA yielded good resolution but dramatically reduced recovery below 70 °C due to strong adsorption.
Replacing TFA with 0.1% DFA and adding 5% n-butanol restored elution and significantly increased recovery at lower temperatures, preserving resolution of impurity peaks.
Across four mAb samples, the BioResolve column with DFA/butanol showed higher or comparable recovery compared to the polymeric column, demonstrating robustness and sample-to-sample consistency.
Benefits and Practical Applications
- Enables high-recovery intact mAb separations at reduced temperatures, minimizing on-column degradation.
- Extends column lifetime by avoiding harsh conditions.
- Improves method robustness for QC and biopharmaceutical development workflows.
Future Trends and Potential Applications
- Extension of optimized RPLC methods to other large biomolecules such as fusion proteins and antibody-drug conjugates.
- Integration with mass spectrometry for enhanced structural characterization.
- Development of new stationary phases and mobile-phase chemistries to further reduce adsorption.
Conclusion
The BioResolve RP mAb Polyphenyl Column combined with milder ion-pairing additives achieves improved intact antibody recovery at lower temperatures, offering a reliable platform for detailed mAb analysis while preserving sample integrity and column performance.
References
- Liu JKH. The history of monoclonal antibody development–progress, remaining challenges and future innovations. Ann Med Surg. 2014;3:113–116.
- Dong M, Hamuro L, Mhatre R, et al. HPLC for characterization and quality control of therapeutic monoclonal antibodies. LCGC North Am. 2014;32(10):796–808.
- Fekete D, Guillarme D, Carr PW. Impact of mobile phase temperature on recovery and stability of monoclonal antibodies using recent reversed-phase stationary phases. J Sep Sci. 2012;35:3113–3123.
- Nguyen J, Betancourt P, et al. Designing a new particle technology for reversed phase protein separations through optimization of diffusion properties. Waters Application Note 720006168EN. 2018.
- Nguyen J, Betancourt P, et al. A novel phenyl bonded phase for improved protein reversed phase separation. Waters Application Note 720006169EN. 2018.
- Bondarenko PV, Elliott S, et al. Optimization of a reversed-phase LC/MS method for characterizing recombinant antibody heterogeneity and stability. J Chromatogr A. 2006;1120:112–120.