HILIC/MS METHOD FOR THE QUANTIFICATION OF THE POLAR BIOMARKER ITACONIC ACID

Posters | 2019 | Waters | PittconInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Waters

Importance of the Topic


Itaconic acid is a small polar metabolite identified as a biomarker for conditions including gestational diabetes mellitus and rheumatoid arthritis. Sensitive quantification in human serum supports early detection and improved patient care.

Objectives and Study Overview


This study aimed to establish a reliable hydrophilic interaction chromatography–tandem mass spectrometry (HILIC-MS/MS) method for quantifying endogenous itaconic acid in human serum, eliminating the need for derivatisation and reducing sample preparation time.

Methodology and Instrumentation


Sample preparation combined protein precipitation with 0.1% trifluoroacetic acid (TFA) and a weak cation exchange (WCX) solid‐phase extraction pass‐through on Waters Oasis WCX µElution plates. Chromatographic separation used a 100 × 2.1 mm, sub-2 µm amide HILIC column with a mobile phase of >80% acetonitrile buffered at pH 9 with 5 mM ammonium acetate. Detection was performed by multiple reaction monitoring (MRM) in negative electrospray ionization on a Waters ACQUITY UPLC H-Class coupled to a Xevo TQXS tandem quadrupole mass spectrometer. Ostro™ PPT plates facilitated protein removal.

Main Results and Discussion


Optimization of protein precipitation and SPE conditions achieved recoveries up to 94% and significantly reduced matrix effects compared to mixed‐mode SPE and precipitation alone. Calibration curves were linear over 2.5–500 ng/mL (r² > 0.997) with mean accuracy between 88.7% and 110.8% and an LLOQ near 1 ng/mL. The validated range covers reported endogenous concentrations (5.2–260 ng/mL).

Benefits and Practical Applications


The developed HILIC-MS/MS assay delivers high sensitivity, rapid workflow without derivatisation, and robust quantification suitable for clinical laboratories and research settings monitoring itaconic acid as a disease biomarker.

Future Trends and Potential Applications


Potential improvements include incorporation of isotopically labelled internal standards, further reduction of LC cycle times, exploration of alternative MRM transitions to minimize interferences, and automation of sample handling. The platform could be extended to other polar metabolites in clinical metabolomics studies.

Conclusion


This feasibility study confirms that HILIC-MS/MS with a sub-2 µm amide column and optimized sample preparation provides reliable, reproducible, and sensitive quantification of itaconic acid in human serum, supporting its application in biomarker research.

References


1. Alpert AJ. J Chromatogr. 1990;499:177–196.
2. Michopoulos F, Whalley N, Theodoridis G, Dunkley TPJ, Critchlow SPJ, Wilson ID. Targeted profiling of polar intracellular metabolites using ion-pair HPLC–UHPLC–MS/MS: applications to serum, urine, and tissue extracts.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Bioanalytical LC-MS Quantification of Itaconic Acid:  A Potential Metabolic Biomarker of Inflammation
A DIRECT INJECT APPROACH FOR ANALYSIS OF LEGACY AND EMERGING PERFLUOROALKYL SUBSTANCES (PFAS) IN ENVIRONMENTAL WATER AND SOIL SAMPLES
DEVELOPMENT OF A BIOANALYTICAL SPE LC-MS/MS ASSAY FOR OLIGONUCLEOTIDES
DETERMINATION OF LEGACY AND EMERGING PERFLUOROALKYL SUBSTANCES IN GROUND AND SURFACE WATERS USING LC-MS/MS WITH DIRECT INJECTION