Ion Mobility, LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
IndustriesProteomics , Metabolomics, Clinical Research
ManufacturerWaters
Significance of the Topic
Rapid metabolomics and lipidomics profiling addresses the challenge of lengthy analysis times, batch effects and data variability in large-scale studies. By reducing run times by 60–75%, these methods enable high-throughput workflows and more consistent data across large cohorts.
Study Objectives and Overview
The main goal was to develop and validate rapid UPLC-MS assays for biofluid profiling in discovery metabolomics and lipidomics. Two complementary methods were optimized: a hydrophilic interaction (HILIC) metabolomics protocol and a reversed-phase lipidomics protocol, both scaled down to minimize cycle time while maintaining chromatographic performance.
Methodology and Instrumentation
- UPLC System: Waters Acquity I-Class with BEH Amide (HILIC) and BEH C8 (lipid) columns, 50×1.0 mm I.D.
- Mass Spectrometer: Waters Synapt G2-Si operated in positive and negative polarity.
- Chromatographic Conditions: 0.2 µL injection, 0.2 mL/min for HILIC at 50 °C; 0.25 mL/min for lipid at 55 °C.
- Chromatographic Gradient: HILIC separation in 2.33 min (99 % B to 50 % B) plus <1 min equilibration; lipid method in 3.70 min total run.
- Ion Mobility: Travelling wave IMS to resolve co-eluting features via collision cross section (CCS).
- Data Processing: MassLynx for acquisition; Progenesis QI for alignment, peak picking and database searches; EzInfo for PCA, OPLS-DA and S-plots.
Main Results and Discussion
- The rapid HILIC method reduced cycle time from 10 min to 3.33 min without loss of retention stability across target compounds.
- The lipid protocol achieved a ~75 % reduction in run time to 3.70 min while preserving class-specific chromatographic regions.
- Analysis of 134 rat urine samples post-treatment completed in 7.5 hours versus >20 hours conventionally.
- Ion mobility doubled the number of detected features by separating overlapping species based on CCS values.
- In human plasma from breast cancer patients, 15 lipids were significant; five upregulated phosphatidylserine species were identified as potential biomarkers.
Benefits and Practical Applications
- Enables acquisition of ~1 000 samples in under three days per assay.
- Reduces mobile phase consumption and minimizes batch-related variability.
- IMS provides an orthogonal separation dimension, improving specificity and identification confidence.
- Applicable to large-scale discovery studies, QA/QC workflows and clinical biomarker screening.
Future Trends and Potential Applications
- Integration of expanded CCS reference libraries for enhanced compound annotation.
- Automation and AI-driven data analysis to accelerate feature selection and interpretation.
- Real-time quality monitoring and adaptive acquisition strategies for ultra-high throughput.
- Application to multi-omic studies combining proteomics and glycomics.
Conclusion
The rapid UPLC-IMS-MS workflows deliver substantial improvements in throughput and data quality for metabolomic and lipidomic profiling. These methods provide robust platforms for large-scale discovery and translational studies.
Reference
- Want EJ, et al. Global metabolic profiling of animal and human tissues via UPLC-MS. Nat Protoc. 2013;8(1):17-32.
- Habchi B, et al. How to really perform high throughput metabolomic analyses efficiently? TrAC Trends Anal Chem. 2016;85:128-139.
- Gray et al. Rapid microbore metabolic profiling UPLC-MS for high-throughput phenotyping. Anal Chem. 2016;88(11):5742-5751.
- Sharma B, Kanwar SS. Phosphatidylserine: A cancer cell targeting biomarker. Semin Cancer Biol. 2017.