Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Applications | 2018 | Agilent TechnologiesInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Environmental, Food & Agriculture, Metabolomics
Manufacturer
Agilent Technologies

Significance of the topic


Amino acids serve as fundamental building blocks of proteins and key metabolites in plant physiology. Accurate quantitation of underivatized amino acids in complex matrices such as plant extracts enables insights into nutritional quality, stress responses, and metabolic pathways.

Objectives and overview


This study presents a rapid and sensitive LC/MS method for quantifying 23 underivatized amino acids in cucumber leaf tissue using hydrophilic interaction chromatography (HILIC) coupled with triple quadrupole detection. It evaluates method linearity, sensitivity, reproducibility, and matrix effects.

Methodology and instrumentation


The separation employs an Agilent InfinityLab Poroshell 120 HILIC-Z column with a pH 3 ammonium formate buffer and acetonitrile gradient, achieving complete separation in 15 min. Sample extracts are prepared by acid extraction of frozen cucumber powder followed by dilution with internal standards. Quantitation uses seven-point calibration (50–10000 ng/mL) with isotope-labeled standards.
  • LC system: Agilent 1260 Infinity binary pump, autosampler, thermostatted column compartment, ultralow dispersion kit
  • Column: Poroshell 120 HILIC-Z, 2.1×100 mm, 2.7 µm
  • MS: Agilent 6470 triple quadrupole with Jet Stream ESI in positive mode

Main results and discussion


The method provides baseline resolution of all target analytes including isobaric pairs leucine/isoleucine and threonine/homoserine. Calibration curves display excellent linearity (R2 > 0.995), and limits of detection reach low ng/mL levels. Reproducibility tests (n=15) show relative standard deviations below 3% for response and retention time. Spike recovery in cucumber extract ranges from 77% to 127%, confirming robustness against matrix effects.

Benefits and practical applications


This workflow eliminates derivatization, streamlines analysis time, and ensures high throughput for plant metabolomics, food quality testing, and agricultural chemistry applications.

Future trends and potential applications


Advances may include coupling this HILIC method with high-resolution mass spectrometry for broader metabolite profiling, miniaturized sample preparation, and automation strategies to extend applications to diverse plant species and stress studies.

Conclusion


The described HILIC-LC/MS protocol delivers rapid, sensitive, and reproducible quantitation of underivatized amino acids in plant matrices, demonstrating its suitability for routine analysis in research and quality control.

Used instrumentation


Agilent 1260 Infinity binary pump, Agilent 1260 autosampler, thermostatted column compartment, ultralow dispersion kit, Agilent 6470 triple quadrupole LC/MS with Jet Stream ESI, Poroshell 120 HILIC-Z column.

References


1. A. Kennedy, A. Bivens. Methods for the Analysis of Underivatized Amino Acids by LC/MS. Agilent Technologies Application Note, 5991-8582EN, 2017.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Analysis of Foods Using Nexera™ Post-Column Amino Acid Analysis System
A Validated Method for the Quantification of Amino Acids in Mammalian Urine
Agilent Biocolumns - Amino Acid Analysis - "How-To" Guide
UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research