Multiclass Multiresidue Veterinary Drug Analysis in Beef Using Agilent Captiva EMR—Lipid Cartridge Cleanup and LC/MS/MS

Applications | 2017 | Agilent TechnologiesInstrumentation
Sample Preparation, Consumables, LC/MS, LC/MS/MS, LC/QQQ
Industries
Food & Agriculture
Manufacturer
Agilent Technologies

Importance of Topic


Analysis of veterinary drug residues in beef is essential for food safety, regulatory compliance, and public health. Complex matrices like muscle tissue require effective sample preparation to achieve reliable quantitation of a wide range of drug classes.

Study Objectives and Overview


This study assesses a streamlined workflow combining two-step extraction and Agilent Captiva EMR–Lipid cartridge cleanup for the determination of 39 multiclass veterinary drugs in beef. Key goals include evaluating matrix cleanup efficiency, analyte recovery, reproducibility, and comparison to alternative pass-through cleanup cartridges.

Methodology and Instrumentation


The optimized sample preparation entails:
  • Aqueous extraction with 0.1 M EDTA to stabilize polar and chelating drugs.
  • Organic extraction using acetonitrile fortified with 2 % formic acid and 2 % DMSO to improve recovery of hydrophobic compounds.
  • Combination of extracts and gravity-driven cleanup on Captiva EMR–Lipid cartridges, followed by a secondary elution with 80:20 ACN/water.

Chromatographic separation was achieved on a C18 UHPLC column with gradient elution, coupled to a triple quadrupole MS with JetStream electrospray ionization for sensitive MRM detection.

Instrumentation Used


  • Agilent 1290 Infinity UHPLC system (binary pump, autosampler, column compartment)
  • Agilent G6490 Triple Quadrupole LC/MS with JetStream ESI
  • Agilent InfinityLab Poroshell 120 EC-C18 column (150 × 2.1 mm, 2.7 µm)
  • Sample prep: Geno/Grinder, refrigerated centrifuge, vortex/shaker, Vac Elut manifold, Captiva EMR–Lipid cartridges

Key Results and Discussion


• Gravimetric tests showed 48 % reduction of co-extractive residue with EMR–Lipid cleanup versus 18–23 % for other cartridges.
• Post-column infusion experiments demonstrated markedly reduced matrix suppression after EMR–Lipid cleanup.
• Cartridge recovery studies revealed consistent >60 % recoveries for hydrophobic analytes using EMR–Lipid sorbent, whereas alternative hydrophobic sorbents caused significant losses (often <10 %).
• Method validation yielded linear calibration (R² > 0.98), recoveries within 60–120 % for 94 % of analytes, and RSD < 10 % for most compounds across low, mid, and high levels.

Benefits and Practical Applications


  • Hands-free, gravity-driven cleanup without vacuum or pressure control
  • Highly selective lipid removal preserving analyte recovery, even for very hydrophobic drugs
  • Minimal method development due to pass-through SPE format
  • Improved laboratory throughput and instrument robustness for food safety testing

Future Trends and Opportunities


Emerging directions include integration of EMR sorbents into automated platforms, extension to other complex matrices (dairy, eggs, aquaculture), development of sorbents targeting additional interferences (e.g., phospholipids), and coupling with high-resolution MS for non-target screening.

Conclusion


The two-step extraction combined with Agilent Captiva EMR–Lipid cartridge cleanup delivers superior matrix removal, high recovery, and robust quantitation of multiclass veterinary drugs in beef, meeting stringent regulatory and analytical quality requirements.

References


1. FDA/USDA. A Description of the U.S. Food Safety System. 2000.
2. EC Decision 2002/657, Official Journal L122. 2002.
3. Mastovska K., Lightfield A.R. J. Chromatogr. A 2008, 1202, 118–123.
4. Geis-Asteggiante L. et al. J. Chromatogr. A 2012, 1258, 43–54.
5. Schneider M.J., Lehotay S.J., Lightfield A.R. Anal. Bioanal. Chem. 2015, 407, 4423–4435.
6. Han L. et al. J. Chromatogr. A 2016, 1449, 17–29.
7. López-Blanco R. et al. J. Chromatogr. A 2016, 1456, 89–104.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Complex Matrices: Minimizing Lipids, Maximizing Recovery - Food Testing Application Compendium: Volume 3
Multiresidue Analysis of Veterinary Drugs in Bovine Liver by LC/MS/MS
Analyzing 193 Veterinary Drug Residues in Livestock and Poultry Meat
Determination of 19 Polycyclic Aromatic Hydrocarbon Compounds in Salmon and Beef