Hydrocarbons, C2 – C3, alcohols, C1 - Analysis of low ppm levels of ethylene, propylene and methanol in water

Applications | 2011 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies

Importance of the Topic


Monitoring trace levels of light hydrocarbons and alcohols in industrial water is critical for process safety and environmental compliance.
Detecting ethylene, propylene and methanol at low parts-per-million concentrations helps identify cooling system leaks and ensures that effluent streams meet regulatory standards before discharge.

Objectives and Overview of the Study


This application note demonstrates a rapid gas chromatographic method for quantifying ethylene, propylene and methanol in aqueous matrices.
The method is designed to support two key operations: weekly leak checks in cooling tower circuits and daily verification of methanol removal from hydrocarbon streams.

Methodology


Samples of water containing target analytes in the range of 8–27 ppm are directly injected into a gas chromatograph.
A temperature-programmed separation enables baseline resolution of ethylene, propylene and methanol within a short analysis window.
Flame ionization detection provides sensitive and reproducible quantitation of each compound.

Used Instrumentation

  • Gas chromatograph with wide-bore inlet and flame ionization detector
  • Agilent PoraPLOT Q column (0.53 mm × 25 m, 20 µm film thickness)
  • Carrier gas: helium at 7 mL/min constant flow
  • Temperature program: 35 °C hold 5 min, ramp to 75 °C at 10 °C/min, then to 200 °C at 25 °C/min
  • Injection: direct, 0.3 µL water sample, inlet at 100 °C, detector at 250 °C

Main Results and Discussion


The method achieved reliable separation and quantification of ethylene (8.8 ppm), propylene (5.4 ppm) and methanol (27.1 ppm) in aqueous samples.
Retention times were consistent, and peak responses showed excellent linearity across the tested concentration range.
Reproducibility was sufficient for routine process monitoring, with minimal interference from the water matrix.

Practical Benefits and Applications

  • Enables early detection of leaks in cooling water systems to prevent equipment damage.
  • Supports daily quality control of methanol removal in hydrocarbon processing.
  • Offers a straightforward, fast and reliable assay suitable for industrial laboratories.

Future Trends and Potential Uses


Integration of headspace sampling or online monitoring could further streamline analysis and reduce manual handling.
Advances in detector technology and miniaturized chromatography systems may allow field-portable measurements for real-time process surveillance.

Conclusion


This GC-FID method provides a robust and efficient solution for low-ppm analysis of ethylene, propylene and methanol in water.
Its ease of use and reproducibility make it well suited for routine process control and environmental monitoring in petrochemical operations.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Determine Impurities in High-Purity Propylene Oxide with Agilent J&W PoraBOND U
Separation of Volatile Organic Hydrocarbons with Agilent J&W PLOT GC Columns and Selectivity Tuning
Detection and Identification of Gulf Oil Dispersants (COREXIT® 9527 and 9500) by GC/MS and LC/MS/MS
Downstream Petrochemical Processes - Application Compendium