Sample Preparation, Consumables, LC/MS, LC/MS/MS, LC/QQQ
IndustriesFood & Agriculture
ManufacturerAgilent Technologies
Significance of the Topic
Monitoring veterinary drug residues in seafood such as salmon is essential for food safety and public health, helping regulators ensure compliance with maximum residue limits in major markets worldwide.
Study Objectives and Overview
The study aimed to develop a streamlined, one-step sample preparation and targeted LC/MS/MS method capable of quantifying 53 veterinary drugs across seven chemical classes in salmon matrices. By integrating extraction and cleanup, the workflow seeks to improve efficiency and reproducibility.
Methodology and Instrumentation
Sample preparation involved:
- Weighing 2 g of homogenized salmon and spiking with analyte standards and isotopically labeled surrogates.
- Extracting with 9 mL of acetonitrile/water/formic acid (80/5/5) using a Geno/Grinder, followed by centrifugation.
- Cleaning the extract via gravity-driven pass-through on a 3 mL Agilent Captiva EMR—Lipid cartridge.
Key Results and Discussion
The method exhibited strong linearity (R² 0.993–0.999) over 0.1–100 ng/g, with matrix-matched calibration. Accuracy and precision were evaluated at low, mid, and high QC levels, yielding recoveries between 70 and 120% for 96% of the 53 drugs and relative standard deviations under 20% in most cases. The one-step extraction with 5% formic acid provided superior recoveries compared to other solvent compositions.
Benefits and Practical Applications
This workflow offers:
- Rapid turnaround with minimal hands-on time.
- Robust cleanup of lipophilic matrix interferences.
- High sensitivity and reproducibility suitable for routine regulatory and quality control analysis.
Future Trends and Potential Applications
Further developments may include adapting the protocol to diverse seafood and animal tissue matrices, integrating automated SPE platforms, and expanding to high-resolution accurate mass detection for untargeted screening of emerging contaminants.
Conclusion
A fast, simple, and reliable sample preparation combined with LC/MS/MS analysis on Agilent platforms enables comprehensive quantification of a broad spectrum of veterinary drug residues in salmon, supporting regulatory monitoring and food safety programs.
Instrumentation
- Agilent 1290 Infinity LC
- Agilent 6495B triple quadrupole LC/MS with Jet Stream ESI
- Agilent InfinityLab Poroshell 120 EC-C18, 3.0×100 mm, 2.7 µm column
- SPEX Geno/Grinder for sample homogenization
- Agilent Captiva EMR—Lipid cartridges (3 mL, 300 mg)
Reference
- Love DC et al. Veterinary drug residues in seafood inspected by the European Union, United States, Canada, and Japan from 2000 to 2009. Environ Sci Technol. 2011;45:7232–7240.