Analysis of Highly Polar Compounds by SFC/Q-TOF MS with Identification using Database and Library Searches

Applications | 2019 | Agilent TechnologiesInstrumentation
LC/TOF, LC/HRMS, LC/MS, LC/MS/MS, SFC
Industries
Environmental
Manufacturer
Agilent Technologies

Importance of the Topic


The analysis of highly polar organic contaminants in environmental waters presents a significant analytical challenge due to poor retention on reversed-phase columns and potential ion suppression from inorganic salts. Supercritical fluid chromatography (SFC) with enhanced fluidity liquid chromatography (EFLC) conditions offers a solution by using high concentrations of organic modifier under elevated pressure. Coupled with accurate-mass quadrupole time-of-flight mass spectrometry (Q-TOF MS) and database/library searches, EFLC-SFC enables sensitive, broad-scope screening of polar analytes in complex matrices such as river water.

Goals and Study Overview


This work aimed to develop and evaluate an SFC-Q-TOF MS screening method for 46 highly polar compounds, spanning ionic buffers, pharmaceuticals, pesticides, artificial sweeteners, and other water contaminants. Key objectives included:
  • Establishing EFLC conditions on multiple stationary phases.
  • Identifying compounds via accurate-mass and fragment-ion library matching.
  • Validating method performance through spike experiments in river water.
  • Assessing limits of detection (LODs), linearity, and compound coverage.

Methodology and Instrumentation


Sample preparation involved spiking river water with 10 µg/L of each target, diluting 1:1 with methanol, and injecting up to 10 µL directly. Chromatographic screening employed four column chemistries under a generic gradient from 5 to 95% methanol (with 3–30 mM ammonium acetate) in liquefied CO₂ at 1.5–2.0 mL/min flow. Q-TOF MS data were acquired in All Ions mode (m/z 50–1000) at collision energies of 0, 10, and 30 eV, in both positive and negative electrospray ionization.

Instrumentation Used


  • Agilent 1260 Infinity II SFC System with Binary Pump, Control Module, Multisampler, High-Pressure Diode Array Detector and 12-position 13-port pump valve.
  • Agilent 1290 Infinity II Multicolumn Thermostat with eight-column selection valve.
  • Agilent 6545 Accurate-Mass Q-TOF LC/MS with Jet Stream ionization.
  • MassHunter Q-TOF Acquisition, Qualitative and Water Screening software with PCDL library.

Main Results and Discussion


Column screening revealed:
  • Cyano (SB-CN) and silica (RxSil) columns provided broad coverage in positive mode, distributing elution across 12 min.
  • Zwitterionic HILIC-Z and HILIC-Plus phases shifted elution windows but showed minor early or late clustering.
  • Negative mode detection yielded complementary coverage, with most analytes eluting mid-run on silica phases and broader windows on HILIC phases.
Database and library searches qualified or confirmed 45 of 46 targets across phases. In river water spikes, a linear signal response was maintained from 1 to 10 µL injection without peak broadening. Twenty-three compounds were detected in negative mode and 22 in positive mode at 10 µg/L, with LODs ranging from 0.03 to >10 µg/L. The majority had LODs below 3 µg/L, and several achieved sub-ppb sensitivity in this screening format.

Benefits and Practical Applications


The EFLC-SFC-Q-TOF workflow offers:
  • Rapid separation of highly polar and ionic analytes in a single run.
  • High throughput with minimal sample cleanup (direct injection after dilution).
  • Comprehensive screening via accurate-mass and fragment-ion matching.
  • Low detection limits suitable for environmental monitoring and QA/QC.

Future Trends and Potential Applications


Advances may include:
  • Extension to broader non-target screening with expanded libraries.
  • Targeted conversion to triple quadrupole MS for enhanced quantitation at lower ppq levels.
  • Automation of sample preparation and data processing for high-volume monitoring.
  • Scaling EFLC-SFC to preparative separations of polar bioactive molecules.

Conclusion


This study demonstrates that EFLC using high modifier concentrations in SFC, combined with accurate-mass Q-TOF detection, effectively separates and identifies a diverse panel of highly polar contaminants in environmental water. The method achieves robust linearity, sub-ppb detection limits, and comprehensive compound coverage, making it a valuable tool for environmental analysis and routine monitoring.

Reference


1. Berset, J. D.; Rennie, E. E.; Glauner, T. Screening and Identification of Emerging Contaminants in Wastewater Treatment Plant Effluents Using UHPLC/Q-TOF MS and an Accurate Mass Database and Library, Agilent Technologies, 2016.
2. Desfontaine, V.; et al. Applicability of Supercritical Fluid Chromatography–Mass Spectrometry to Metabolomics. I – Optimization of Separation Conditions for the Simultaneous Analysis of Hydrophilic and Lipophilic Substances, J. Chrom. A, 2018, 1562, 96–107.
3. Bennett, R.; et al. Enhanced Fluidity Liquid Chromatography: A Guide to Scaling up from Analytical to Preparative Separations, J. Chrom. A, in press, 2019.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Extend Your Application Reach with Achiral and Chiral SFC Columns
Extend Your Application Reach with Achiral and Chiral SFC Columns
2019|Agilent Technologies|Brochures and specifications
APPLICATION NOTEBOOK - UNTARGETED METABOLOMICS  AND LIPIDOMICS
Discovery Metabolomics LC/MS Methods Optimized for Polar Metabolites
Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry