SFC
IndustriesForensics
ManufacturerAgilent Technologies
Importance of the Topic
Polar and ionic drugs and metabolites such as GHB, GBL, ethyl sulfate (ETS), ethyl-β-D-glucuronide (ETG), GHB-β-O-glucuronide, meldonium and γ-butyrobetaine present significant analytical challenges in doping control due to poor retention and peak shape in conventional HPLC or GC methods. Supercritical fluid chromatography (SFC) with ion-exchange stationary phases offers complementary selectivity and robust quantification of these analytes in biological matrices.
Objectives and Study Overview
This study aimed to develop and validate an SFC/MS/MS method using the Agilent 1260 Infinity II SFC System and triple quadrupole mass spectrometry to separate and detect highly polar and ionic compounds relevant to anti-doping analysis. Method development included stationary phase screening, modifier and additive optimization, and evaluation of chromatographic parameters under SFC conditions.
Methodology and Instrumentation
- Sample preparation: protein precipitation of urine with methanol, 1:10 final dilution.
- Instrumentation:
- Agilent 1260 Infinity II SFC System (Control Module, Binary Pump, Multisampler, Diode Array Detector, Multicolumn Thermostat)
- Agilent 6470A Triple Quadrupole LC/MS with Agilent Jet Stream
- Data processing: Agilent MassHunter software with MRM and source optimizers
- Stationary phase: silica-based strong cation exchange (SCX) with phenyl modification (4.6×150 mm, 5 µm)
- Mobile phase: CO2 with methanol/water (95:5 v/v) containing 20 mM ammonium formate and 15 mM formic acid
- Gradient: 15% modifier to 60% in 1 min; flow 2.0 to 2.5 mL/min; column 45 °C; backpressure 170 bar
- MS conditions: ESI+ and ESI−, optimized MRM transitions, drying gas 250 °C, sheath gas 375 °C, nebulizer 15 psi
Main Results and Discussion
The optimized SFC method achieved baseline separation of all seven analytes within 4 minutes. Retention times and peak shapes were controlled by water content and additive concentration; higher ammonium formate improved peak shape, particularly for ETG. Method performance data: detection limits ranged from 0.001 mg/L (ETS, meldonium) to 0.5 mg/L (ETG, GHB-Gluc); LOQs were comparable or superior to conventional HPLC/MS/MS methods. Linearity (R2 > 0.97), intra-day precision (area CV 2.1–13.4%, RT CV < 1%), and matrix factors (0.30–1.57) demonstrated robust quantification in urine. LOQs met or exceeded doping control cut-off levels and MRPL requirements.
Benefits and Practical Applications
- Orthogonal separation to HPLC, retaining polar and ionic analytes without breakthrough
- Rapid analysis (< 5 min) with high resolution and consistent peak shapes
- Robust quantification at sub-µg/L levels for anti-doping screening
- Green solvent usage through CO2 as main mobile phase component
Future Trends and Potential Applications
- Extension to broader panels of polar biomarkers and metabolites
- Integration of high-resolution mass spectrometry for non-targeted screening
- Automation of sample preparation and on-line SPE coupling
- Application in clinical and forensic toxicology for rapid polar compound profiling
Conclusion
A dedicated SFC/MS/MS workflow using a strong cation exchange column and optimized methanol/water modifier has been established for efficient separation and quantification of highly polar and ionic doping control analytes. The method delivers rapid, precise, and sensitive analysis in urine, offering a viable alternative to traditional HPLC and GC techniques for routine anti-doping and forensic applications.
References
- Par, M. K.; et al. Anal. Bioanal. Chem. 2016, 408, 6789–6797.
- Desfontaine, V.; et al. Bioanalysis 2015, 7, 1193–1195.
- Fujito, Y.; et al. J. Chromatogr. A 2017, 1508, 138–147.
- Novakova, L.; et al. Anla. Chim. Acta 2015, 853, 647–659.
- Abanades, S.; et al. Ann. N Y Acad. Sci. 2006, 1074, 559–576.
- Mehling, L.-M.; et al. Forensic Toxicol. 2017, 35, 263–274.
- Xhaferaj, M.; et al. J. Chromatogr. A 2020, 1614, 460726.
- Albertmann, M.; et al. J. Chromatogr. Sci. 2012, 50, 51–56.
- Kim, Y.; et al. Mass Spectrom. Lett. 2017, 8, 39–43.
- WADA Technical Document TD2016EAAS, 2015.
- Brailsford, A. D.; et al. J. Anal. Toxicol. 2012, 36, 88–95.
- WADA TD2018MRPL, 2018.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.