Evaluating BioAccord System Performance in Late Stage Development and QC Environment – Peptide Mapping Analysis

Technical notes | 2020 | WatersInstrumentation
LC/TOF, LC/HRMS, LC/MS
Industries
Pharma & Biopharma
Manufacturer
Waters

Importance of the Topic


Liquid chromatography–mass spectrometry (LC-MS) peptide mapping plays a critical role in verifying protein primary sequences and monitoring product quality in biopharmaceutical development. Its reliability is essential for detecting post-translational modifications and ensuring batch-to-batch consistency under regulated conditions.

Study Objectives and Overview


This study aimed to demonstrate the reproducibility of the Waters BioAccord System for peptide mapping in late-stage development and QC environments. By conducting inter- and intra-system assessments using three identical platforms, the work addresses performance criteria required for traditional method validation.

Methodology


Using the Waters mAb Tryptic Digest Standard (NISTmAb), sample solutions were prepared at 0.2 µg/µL and 1 µg injected on column. Peptide separation was performed on an ACQUITY UPLC BEH C18 1.7 µm (2.1×100 mm) column at 65 °C with a 60-minute gradient from 99% to 60% aqueous mobile phase (0.1% formic acid) at 0.2 mL/min, followed by a total run time of 75 minutes including equilibration. Mass detection employed the ACQUITY RDa high-performance oa-TOF detector in positive ESI full-scan and fragmentation mode (2 Hz) over m/z 50–2000. Cone voltage was set to 30 V with a fragmentation cone ramp of 60–130 V, capillary at 1.5 kV, and desolvation at 500 °C. Data acquisition and processing were automated via waters_connect and the UNIFI Scientific Information System with a 10 ppm mass tolerance and minimum of three confirmatory fragments for peptide assignment.

Instrumentation


  • ACQUITY UPLC I-Class PLUS System with optical detection (TUV/FLR)
  • ACQUITY RDa Mass Detector (oa-TOF)
  • waters_connect platform
  • UNIFI Scientific Information System
  • ACQUITY UPLC BEH C18 column, 1.7 µm, 2.1×100 mm

Main Results and Discussion


Overlay of total ion chromatograms from 18 injections over 22 hours revealed retention time standard deviations ≤0.09 minutes (5.4 seconds) across all monitored peptides. A set of native and modified peptides (including glycopeptides and oxidation/deamidation variants) showed an average relative abundance reproducibility (%RSD) within 8.12% across the three systems. These findings confirm both chromatographic stability and consistent MS response suitable for automated attribute monitoring.

Benefits and Practical Applications


The BioAccord System delivers robust, compliance-ready peptide mapping workflows with minimal manual intervention. Its reproducibility supports routine QC deployment and late-stage development efforts, facilitating reliable monitoring of critical quality attributes in biotherapeutics.

Future Trends and Opportunities


Advances in LC-MS automation, data analytics, and integration with information management systems will further streamline peptide mapping in regulated environments. Emerging techniques such as ion mobility separation and AI-driven data processing are poised to enhance throughput, sensitivity, and confidence in attribute monitoring applications.

Conclusion


Inter- and intra-system evaluations demonstrate that the BioAccord System meets stringent reproducibility requirements for peptide mapping in regulated workflows. Consistent retention times and modification quantification underscore its fitness for deployment in QC laboratories and late-stage biopharmaceutical development.

References


  • 1. Shion H et al. Enabling Routine and Reproducible Intact Mass Analysis When Data Integrity Matters. Waters Application Note, 720006472EN, 2019.
  • 2. Ranbaduge N et al. Routine Peptide Mapping Analysis using the BioAccord System. Waters Tech Brief, 720006466EN, 2019.
  • 3. Zhang X et al. Released N-linked Glycan Analysis Using the BioAccord System. Waters Tech Brief, 720006474EN, 2019.
  • 4. Shion H et al. A Platform Method for the Molecular Mass Analysis of the Light Chains and Heavy Chains of Monoclonal Antibodies using the BioAccord System. Waters Tech Brief, 720006529EN, 2019.
  • 5. Zhang X et al. Increasing Productivity and Confidence for N-linked Glycan Analysis of Biosimilars Using the BioAccord System. Waters Tech Brief, 720006545EN, 2019.
  • 6. Shion H et al. Analysis of Antibody Drug Conjugates by Native Mass Spectrometry on the BioAccord System. Waters Tech Brief, 720006570EN, 2019.
  • 7. Ranbaduge N et al. Establishing an Integrated Peptide Attribute Profiling and Monitoring Workflow for Improved Productivity and Confidence. Waters Application Note, 720006602EN, 2019.
  • 8. Ippoliti S et al. Online IEX-MS of mAb Charge Variants Using a BioResolve SCX mAb Column, IonHance CX-MS pH Concentrates, and BioAccord System. Waters Application Note, 720006672EN, 2019.
  • 9. Shion H et al. Evaluating BioAccord LC-MS System Performance for Deployment of a Compact, SmartMS-Enabled Biopharma Solution in Late Stage Development and QC Environment – Intact Mass Analysis. Waters Tech Brief, 720006722EN, 2020.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Evaluating BioAccord LC-MS System Performance for Deployment of a Compact, SmartMS-Enabled Biopharma Solution in Late Stage Development and QC Environment – Intact Mass Analysis
Routine Peptide Mapping Analysis using the BioAccord System
Analysis of Antibody Drug Conjugates (ADCs) by Native Mass Spectrometry on the BioAccord System
Multi-Attribute Methods for Biopharmaceutical Analysis