Significance of the Topic
The widespread release of pharmaceuticals and personal care products (PPCPs) into aquatic environments poses risks due to their persistence, bioactivity and endocrine-disrupting potential. Developing rapid, sensitive and multi-residue analytical methods is essential for environmental monitoring, regulatory compliance and risk assessment.
Study Objectives and Overview
This work aimed to establish a unified UHPLC-ESI-QqQ method capable of quantifying diverse PPCP classes in surface waters. The key goals were to:
- Implement fast polarity switching to detect both positively and negatively ionizing analytes in a single run.
- Optimize the chromatographic gradient to minimize matrix effects and achieve ng/L detection limits.
- Compare performance against conventional single-polarity analyses and evaluate long-term stability.
Methodology and Used Instrumentation
Surface water samples from Japanese rivers and lakes were spiked with 15 PPCPs at concentrations ranging from 1 to 10 000 ng/L without pretreatment.
Used Instrumentation
- UHPLC System: Nexera (Shimadzu, Japan)
- Column: Shim-pack XR-ODSIII (2.0 mm × 50 mm, 1.6 µm) at 40 °C
- Mobile Phases: A – 0.1% formic acid in water; B – acetonitrile; 21 min gradient (0→80% B in 16 min; 100% B at 18 min; re-equilibration to 0% B by 21 min); flow 0.4 mL/min; injection 40 µL
- MS Detector: LCMS-8080 triple quadrupole (Shimadzu, Japan) with ESI, fast polarity switching (20 ms)
- Gas and Temperature Settings: nebulizing 3.0 L/min; curtain 3.5 L/min; heating 12.0 L/min; probe 450 °C; HSID 300 °C
Main Results and Discussion
Optimized gradient separation effectively focused analytes at early elution and reduced matrix interferences.
All 15 PPCPs showed excellent linearity (R2 > 0.999) and limits of quantification between 1 and 50 ng/L. Recoveries in river and lake water ranged from 70% to 120%.
Fast polarity switching delivered data quality comparable to dedicated positive- or negative-only runs. Over 100 sequential injections (10 h), %RSD for representative compounds remained below 4%, demonstrating high stability and reproducibility.
Benefits and Practical Applications
This single-run, multi-residue approach reduces analysis time, solvent use and instrument workload. Achieving sub-ppb sensitivity for a diverse compound set makes it suitable for routine environmental monitoring, water quality assessment and regulatory screening.
Future Trends and Potential Applications
Advances may include coupling to high-resolution MS for suspect screening, automated on-site sampling, broader analyte libraries and integration with data-driven risk models. Further miniaturization and real-time analysis could enhance field deployability and rapid response capabilities.
Conclusion
The UHPLC-ESI-QqQ method with fast polarity switching provides a robust, sensitive and efficient solution for multi-component quantification of PPCPs in environmental waters. Its high reproducibility and low detection limits support comprehensive monitoring and safeguard ecosystem and human health.