GC/MSD, GC/TOF
IndustriesEnergy & Chemicals
ManufacturerJEOL
Importance of the Topic
Ionic liquids, composed entirely of ions or ion pairs, combine unique properties such as electrical conductivity, negligible vapor pressure, low combustibility and excellent thermal stability. These features position them as promising functional materials in applications ranging from green solvents to electrochemical devices.
Objectives and Study Overview
The aim of this work was to demonstrate the capability of field desorption (FD) ionization coupled with high-resolution time-of-flight mass spectrometry (JMS-T100GC “AccuTOF GC”) for characterizing a model ionic liquid. Specifically, the study sought to identify both the cationic and anionic components of 1-butyl-3-methylpyridinium bis(trifluoromethylsulfonyl)imide through accurate mass measurements.
Methodology
- Sample: 1-Butyl-3-methylpyridinium bis(trifluoromethylsulfonyl)imide (Aldrich).
- Ionization: Field desorption in positive mode (FD(+)), cathode potential –10 kV, emitter current ramped to 35 mA.
- Mass range: m/z 35–800, with spectral acquisition every 1 s.
Used Instrumentation
- Mass spectrometer: JEOL JMS-T100GC “AccuTOF GC” high-resolution time-of-flight MS.
Main Results and Discussion
The FD mass spectrum exhibited a base peak at m/z 150.1, corresponding to the intact pyridinium cation (C10H16N)+. A secondary peak at m/z 580.2 was assigned to a cluster ion comprising one anion and two cation units. Accurate mass measurements yielded:
- Cation (C): measured m/z 150.1289 (calculated 150.1283), formula C10H16N.
- Cluster (CA+C): measured m/z 580.1766 (calculated 580.1738), formula C22H32F6N3O4S2.
Benefits and Practical Applications
- Minimal fragmentation allows direct detection of intact cations.
- High mass accuracy facilitates unambiguous elemental composition assignments for both ions.
- Applicable to quality control and structure confirmation of novel ionic liquids in research and industry.
Future Trends and Applications
- Integration with chromatographic separation for complex mixtures.
- Automation of FD probe handling for high-throughput screening.
- Extension to diverse ionic liquid chemistries and co-solvent systems.
- Combination with tandem MS for deeper structural elucidation.
Conclusion
Field desorption coupled with high-resolution TOF-MS provides a robust method for analyzing ionic liquids. The technique yields accurate masses of intact cations and cluster ions, enabling determination of both cationic and anionic compositions without extensive fragmentation.
Reference
- J. H. Gross, Mass Spectrometry, Springer-Verlag Berlin Heidelberg, Germany (2004).
- J. H. Gross, J. Am. Soc. Mass Spectrom., 18, 2254 (2007).
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.