Qualitative analysis by comprehensive 2D GC / TOFMS - Comparison of kerosene and diesel oil

Applications | 2005 | JEOLInstrumentation
GCxGC, GC/MSD, GC/TOF
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies, JEOL, ZOEX/JSB

Significance of the Topic


Comprehensive two-dimensional gas chromatography coupled with time-of-flight mass spectrometry (GC×GC-TOFMS) offers exceptional separation power for complex hydrocarbon mixtures. In petrochemical analysis, distinguishing compounds by boiling point and polarity is critical for characterizing fuels, ensuring quality control, and optimizing refining processes.

Objectives and Study Overview


This application note illustrates the qualitative comparison of kerosene and diesel oil using a GC×GC-TOFMS system. The primary goal is to demonstrate the high resolving capability of the JEOL AccuTOF-GC mass spectrometer when paired with a GC×GC setup, enabling rapid and detailed profiling of hydrocarbon distributions in different fuel fractions.

Methodology and Instrumentation


A comprehensive layout of the analytical workflow:
  • GC×GC System
    Agilent 6890 GC with ZOEX KT2004 modulator within a single oven to perform heart-cutting between two columns.
  • Column Configuration
    First dimension: HP-1ms (30 m × 0.25 mm I.D., 0.25 µm)
    Second dimension: DB-17 (2 m × 0.1 mm I.D., 0.1 µm)
  • Temperature Program
    Oven: 50 °C (1 min) → 5 °C/min → 280 °C (6 min)
    Injection: 280 °C, split mode 1:200, 0.5 µL volume
  • Carrier Gas
    Helium at constant pressure of 680 kPa
  • Modulation
  • Trapping interval: 6 s

  • TOFMS Detection (AccuTOF-GC JMS-T100GC)
    Ionization: EI+ at 70 eV, 300 µA
    Mass range: m/z 35–500
    Acquisition rate: 25 Hz (0.04 s per spectrum)

Results and Discussion


GC×GC contour plots reveal distinct patterns for kerosene (C9–C15) and diesel oil (C11–C15). The first dimension separates components by volatility: kerosene shows a greater abundance of lower-boiling homologues, whereas diesel is enriched in higher-boiling species. The second dimension resolves compounds by polarity, differentiating saturated, unsaturated, and aromatic hydrocarbons. Peak intensity mapping (color scale from light blue to red) highlights major constituents and their relative concentrations.

Benefits and Practical Applications


  • Enhanced Selectivity: Two-dimensional separation offers superior discrimination of isomers and structural classes.
  • High Throughput: Fast acquisition by AccuTOF-GC ensures no peak loss despite narrow chromatographic signals.
  • Improved Sensitivity and Resolution: High mass resolving power permits confident qualitative profiling of complex mixtures.
  • Quality Control: Detailed compositional fingerprints support batch consistency, contamination assessment, and product verification.

Future Trends and Potential Applications


Advancements in modulation technology and data processing will further reduce cycle times and enhance peak capacity. Integration with chemometric tools can automate pattern recognition for rapid screening in petroleum refining, environmental monitoring, and forensic investigations. Emerging applications may include real-time process control and remote sensing of volatile organic compounds.

Conclusion


This study demonstrates that GC×GC-TOFMS, when equipped with the AccuTOF-GC detector, provides unparalleled separation and mass spectral acquisition speed for complex fuel analyses. The technique offers a robust platform for detailed hydrocarbon profiling, improving quality assurance and enabling new analytical workflows in petrochemical research.

References


[1] JEOL Ltd. Application Note No. 040/055, 2005.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Qualitative analysis by comprehensive 2D GC / TOFMS - Analysis of polycyclic aromatic hydrocarbons in kerosene
Qualitative analysis by comprehensive 2D GC / TOFMS - Analysis of sulfur-contained substances in diesel oil
AccuTOF GC series Applications Notebook
AccuTOF GC series Applications Notebook
2016|Agilent Technologies|Applications
Analysis of diesel oil by using GC x GC-HRTOFMS (FI) with 2 different sets of column combinations