Rapid Identification and Quantification of Performance-Enhancing Stimulants in Human Urine Using High-Resolution Mass Spectrometry

Applications | 2020 | SCIEXInstrumentation
LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
Industries
Forensics
Manufacturer
SCIEX

Importance of the Topic


Performance-enhancing stimulants are widely abused in competitive sports to reduce fatigue and boost alertness, posing significant health risks to athletes. Rigorous detection methods are essential to enforce anti-doping regulations and safeguard athlete well-being.

Study Objectives and Overview


This study aimed to develop and validate a rapid, high-throughput workflow for the unequivocal identification and quantification of 15 structurally diverse stimulants in human urine. By combining a simple sample preparation protocol with high-resolution mass spectrometry and data-independent acquisition, the method addresses limitations of traditional HPLC-UV and GC-MS approaches.

Methodology and Instrumentation


Sample preparation involved a tenfold dilution of spiked urine with acetonitrile/methanol (80/20, v/v), followed by ultracentrifugation. Calibration curves were built over 100–1000 ng/mL using two replicates per level.

Liquid chromatography was performed on a Phenomenex C18 column (100 × 2.1 mm, 1.7 µm) at 45 °C with water and acetonitrile modifiers, a 0.5 mL/min flow rate, and a 7 min run time.

Mass spectrometry data were acquired in positive mode using SWATH® Acquisition on the SCIEX X500R QTOF System, with one TOF MS scan followed by 12 variable Q1 MS/MS windows across m/z 100–350 and a cycle time of 0.555 s. Data processing utilized SCIEX OS Software 1.5.

Instrumentation Used


  • SCIEX ExionLC™ AC UHPLC system
  • Phenomenex C18 column (100 × 2.1 mm, 1.7 µm)
  • SCIEX X500R QTOF mass spectrometer
  • SCIEX OS Software 1.5 for data acquisition and processing

Main Results and Discussion


The method achieved near-baseline chromatographic separation of all 15 stimulants within 7 min. Calibration curves exhibited excellent linearity (R2 > 0.99) and sensitivity, with limits of detection (LOD) between 1.0–30 ng/mL and limits of quantification (LOQ) between 3.3–100 ng/mL.

Matrix effects were limited (±2.2 % to ±35.7 %), indicating minimal ion suppression/enhancement. Intra-day precision and accuracy were both below 20 % across all tested levels (100, 500, 1000 ng/mL), confirming the robustness and reproducibility of the dilute-and-shoot workflow.

Benefits and Practical Applications


  • High throughput: simple dilute-and-shoot protocol reduces sample preparation time.
  • Comprehensive screening: SWATH acquisition archives full MS/MS spectra for each stimulant in every sample.
  • Regulatory compliance: sensitive detection aligned with WADA Prohibited List requirements.
  • Scalability: potential to expand the panel to additional stimulants with minimal method adjustment.

Future Trends and Applications


Advances in data-independent acquisition and HRMS hardware are expected to further enhance sensitivity and throughput. Integration with automated sample handlers and artificial intelligence–driven data analysis may enable real-time anti-doping monitoring and retrospective screening of large sample cohorts.

Conclusion


The presented workflow leverages rapid sample preparation and SWATH® acquisition on a QTOF platform to deliver reliable, high-sensitivity detection of performance-enhancing stimulants in urine. Its robust performance, combined with digital archiving of full MS/MS datasets, supports routine anti-doping applications and future panel expansions.

References


  • Negri P, Gerace E, Di Corcia D, Vincenti M, Salomone A. Rapid Identification and Quantification of Performance-Enhancing Stimulants in Human Urine Using High-Resolution Mass Spectrometry. Application note, SCIEX X500R QTOF System, 2020.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Rapid Identification and Quantification of Novel Psychoactive Substances in Human Whole Blood Using SWATH® Acquisition
Detection of Fentanyl Analogs and Novel Synthetic Opioids in Hair
Rapid Detection of Growth Hormone-Releasing Peptides in Dried Blood Spots: a Proof of Concept Workflow for Anti-Doping Analysis
Intelligently Designed SWATH® Acquisition for Novel Psychoactive Substances (NPS) Detection in Whole Blood