High-Sensitivity Analysis of Drugs in Ultra-Small Volumes Plasma Samples Using Micro-Flow LCMS/ MS

Posters | 2019 | ShimadzuInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Shimadzu

Importance of the Topic


Accurate quantification of drugs and metabolites in minute plasma volumes is vital for discovery-stage pharmacokinetic studies, especially when sample availability is limited by micro-sampling in small animal models. High-sensitivity analytical methods reduce animal usage, lower costs, and improve data quality for safety and efficacy assessments.

Objectives and Study Overview


This study demonstrates a proof-of-concept micro-flow LC–MS/MS method for quantifying the antiarrhythmic drug verapamil and its metabolite nor-verapamil in ultra-small plasma samples. The goals were to:
  • Develop a trap-and-elute micro-flow configuration to maximize sensitivity.
  • Evaluate signal enhancement versus semi-micro flow.
  • Assess recovery, accuracy, and precision using 2 µL plasma inputs.

Methodology and Instrumentation


The workflow combined micro-sampling and isotope-dilution MRM analysis:
  1. Sample Preparation:
    – 2 µL plasma spike with verapamil, nor-verapamil, and deuterated internal standard.
    – Dilution with acetonitrile/formic acid, centrifugation, and 5 µL injection.
  2. Chromatography:
    – Shimadzu Nexera MIKROS system with trap (Shim-pack MCT LC8) and analytical column (0.2×100 mm, 2.7 µm biphenyl).
    – Mobile phases water/formic acid and acetonitrile/formic acid.
    – Flow at 4 µL/min, 11 min total run time.
  3. Detection:
    – LCMS-8060 triple quadrupole with micro-ESI source.
    – MRM transitions for verapamil (455.0 > 150.25) and nor-verapamil (440.95 > 165).
    – Linear calibration 0.5–185 µg/L, 1/x weighting.

Key Results and Discussion


Signal intensity improved markedly compared to semi-micro flow:
  • Verapamil: >4.5-fold increase in peak area, >10-fold improvement in signal-to-noise.
  • Nor-verapamil: >3.5-fold increase in signal.
Recovery and matrix effects were evaluated at low QC (0.5 µg/L):
  • Overall recoveries >97% for both analytes.
  • Calibration accuracy between 85–115% across seven levels.

Benefits and Practical Applications


The micro-flow trap-and-elute approach offers:
  • High sensitivity without increasing injection volume.
  • Compatibility with ultra-small plasma samples (2 µL).
  • Potential integration with automated micro-sampling devices to adhere to 3R principles.

Future Trends and Applications


Emerging directions include applying micro-flow LC–MS/MS to:
  • High-throughput drug screening in early drug discovery.
  • Clinical microsampling for pediatric and neonatal studies.
  • Integration with novel sampling devices for minimally invasive monitoring.

Conclusion


The presented micro-flow LC–MS/MS method dramatically enhances sensitivity for verapamil and nor-verapamil quantification in ultra-small plasma volumes, enabling reduced sample requirements and supporting ethical micro-sampling strategies in pharmacokinetic research.

Reference


[1] Rapid Communications in Mass Spectrometry 2014, 28, 1293–1302

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

High-Sensitivity Analysis of Drugs in Ultra-Small Volume Plasma Samples Using Microflow LC/MS/MS
Shimadzu Microow Liquid Chromatography Mass Spectrometry System Nexera Mikros
Microfow LC-MS/MS analysis of monoclonal antibody in human plasma at ng/mL level with nSMOL proteolysis
Selective and sensitive quantification of glucagon in human plasma using microflow LC/Q-TOF MS