Analysis of Tap Water Using ICDS™-40A Suppressor Unit

Applications | 2020 | ShimadzuInstrumentation
Ion chromatography
Industries
Environmental
Manufacturer
Shimadzu

Importance of the Topic


Ion chromatography provides high-sensitivity detection of ionic species and is widely used for environmental, pharmaceutical, and food analysis. Accurate measurement of anions in drinking water is essential to ensure compliance with regulatory limits and to protect public health.

Objectives and Study Overview


This study demonstrates the analysis of five anions (fluoride, chloride, nitrite, chlorate, nitrate) in tap water according to Japanese MHLW regulations. Special focus is placed on achieving reliable quantitation of nitrite nitrogen at the low standard limit of 0.004 mg/L.

Methodology and Instrumentation


Anion separation was performed on a Shim-pack IC-SA4 column (150 mm × 4.6 mm I.D.) using a mobile phase of 1.7 mmol/L sodium carbonate and 5.0 mmol/L sodium hydrogen carbonate at 0.8 mL/min and 50 °C. A 50 µL sample volume was injected. Detection combined electroconductivity for most anions and UV-VIS at 210 nm for nitrite to avoid chloride interference.

Used Instrumentation


  • Shimadzu ICDS-40A electrodialysis suppressor unit for continuous suppression and regeneration
  • Shim-pack IC-SA4 analytical column
  • Electroconductivity detector
  • UV-VIS detector (210 nm)

Main Results and Discussion


Chromatograms of standard solutions confirmed baseline separation of five anions. Calibration for nitrite nitrogen over 0.004–0.040 mg/L yielded r² ≥ 0.999. Six replicate injections at 0.004 mg/L showed retention time RSD ≤ 0.12% and area RSD ≤ 3.07% (conductivity) and ≤ 1.10% (UV-VIS). Relative errors for standards were within ±3.87%. Analysis of tap water spiked at 0.004 mg/L demonstrated all analytes below regulatory limits.

Benefits and Practical Applications


  • The ICDS-40A suppressor’s simultaneous suppression and regeneration shortens cycle time.
  • High sensitivity enables quantitation at low regulatory thresholds.
  • Dual detection ensures accuracy for both common anions and low‐level nitrite nitrogen.
  • Method aligns with official drinking water quality standards for routine QC.

Future Trends and Potential Applications


Advancements may include coupling IC with mass spectrometry for broader anion profiling, integration into automated online monitoring systems for real-time water quality surveillance, miniaturized suppressor designs for field use, and application to emerging ionic contaminants in environmental samples.

Conclusion


A Shimadzu ICDS-40A based ion chromatography method reliably quantifies key anions in tap water, meeting stringent MHLW requirements, including trace nitrite nitrogen analysis, and offers rapid, high-precision performance for water quality laboratories.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Ion Chromatography Solutions for Environmental Analysis
Analysis of inorganic anions in tap water according to EPA Method 300.1 using Ion Chromatography
Analysis of Total Nitrogen and Total Phosphorus in Environmental Water According to ASTM D8001 by Ion Chromatography
Simultaneous Analysis of Anions and Cations Using Ion Chromatograph Dual Channel System