Extractables and Leachables Analysis from Laboratory Tubing

Applications | 2019 | SCIEXInstrumentation
LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
Industries
Pharma & Biopharma
Manufacturer
SCIEX

Significance of the Topic


The analysis of extractables and leachables (E&L) is critical for ensuring the safety and integrity of pharmaceutical products that contact polymeric materials, governed by regulatory guidance such as USP chapters 1663 and 1664. It addresses non-volatile organic compounds that may compromise drug quality and patient safety.

Objectives and Study Overview


This study presents a comprehensive LC-MS/MS workflow employing SWATH data-independent acquisition combined with a curated E&L-focused library to improve the detection and identification of non-volatile extractables and leachables from laboratory tubing extracts.

Methodology and Instrumentation


  • Sample Preparation: Laboratory tubing soaked in acetone to extract potential leachables.
  • Chromatography: ExionLC AD HPLC with PDA detection, Phenomenex Luna Omega C18 column, gradient elution at 0.4 mL/min.
  • Mass Spectrometry: SCIEX X500R QTOF operated in SWATH mode with ten m/z windows spanning 100–1300 m/z.
  • Data Processing: SCIEX OS Analytics software using sample vs control comparison, targeted XIC list (~300 m/z) and non-targeted peak finding.
  • Library Search: Custom NIST MS/MS subset (~3,000 spectra) and ChemSpider integration for unknowns via formula generation and in-silico fragmentation.


Main Results and Discussion


  • SWATH acquisition provided MS/MS for coeluting, low-abundance analytes, exemplified by cyclohexylamine detection at m/z 100.076.
  • Unmatched peaks, such as a component at RT 5.67 (C17H35N2O2), were characterized using ChemSpider and in-silico MS/MS matching.
  • Polymer degradants, including nylon-based fragments, were tentatively identified based on molecular formula assignments.
  • Optical detection (PDA) verified key phthalate standards but did not reveal additional non-ionizable leachables.


Benefits and Practical Applications


The proposed workflow offers:
  • Comprehensive, unbiased fragmentation reducing missed analytes.
  • Regulatory compliance with USP chapters 1663/1664.
  • Expandable, user-customizable E&L libraries.
  • Integration of MS and optical detection for broader compound coverage.


Future Trends and Potential Uses


  • Expansion of specialized E&L spectral libraries through collaborative contributions.
  • Machine learning approaches for automated fragment assignment and identification.
  • Advanced polymer degradant elucidation using predictive modeling.
  • Real-time monitoring and automation of DIA workflows for high-throughput screening.


Conclusion


The SWATH-based high-resolution LC-MS/MS workflow on the SCIEX X500R QTOF platform delivers robust detection and confident identification of extractable and leachable compounds, supported by customizable libraries and complementary optical detection, establishing a reliable method for pharmaceutical material safety assessment.

Used Instrumentation


  • SCIEX X500R QTOF mass spectrometer with SWATH acquisition.
  • ExionLC AD HPLC system with PDA 200–400 nm detector.
  • Phenomenex Luna Omega C18 column (1.6 µm, 150×2.1 mm).
  • SCIEX OS Analytics software and custom NIST MS/MS library subset.


References


  1. Improved Data Quality Using Variable Q1 Window Widths in SWATH Acquisition - Data Independent Acquisition on TripleTOF and X-Series QTOF Systems. SCIEX Technical note RUO-MKT-02-2879-B.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Quantitative and Qualitative Analysis of Cell Culture Medium Using SWATH® Acquisition
Improved Metabolite Identification Using Data Independent Analysis for Metabolomics
Chemovar Typing of Cannabis Strains with MarkerView™ Software and the SCIEX X500R QTOF System
Next Generation Quality Control in Pharma Applications