LC/MS, LC/MS/MS, LC/QTRAP
IndustriesClinical Research
ManufacturerSCIEX
Significance of the Topic
Glucocorticoids such as budesonide play a critical role in anti-inflammatory therapy for asthma, allergic rhinitis, and dermatological conditions. Due to its low oral bioavailability, achieving reliable quantitation of budesonide at low picogram-per-milliliter levels in plasma is essential for accurate pharmacokinetic and bioequivalence assessments.
Objectives and Study Overview
This application note describes the development and validation of a sensitive LC-MS/MS assay using the SCIEX Triple Quad™ 5500+ – QTRAP® Ready system to quantify budesonide in human plasma across a concentration range of 2 to 1024 pg/mL.
Methodology and Experimental Procedures
- Sample Preparation
- Human plasma samples (200 µL) spiked with budesonide standards and deuterated internal standard (budesonide-D8).
- SPE cleanup on Phenomenex Strata-X RP cartridges: conditioning with methanol and water, sample load, wash with 5% methanol, and elution with 100% methanol.
- Drying under vacuum and reconstitution in 40% acetonitrile for LC injection.
- Chromatography
- Column: C18 (100 × 2.1 mm, 1.7 µm) at 50 °C.
- Mobile phase A: 5 mM ammonium bicarbonate; B: acetonitrile; flow rate 0.5 mL/min.
- Gradient: 40% B at 0 min to 95% B at 2 min, held to 3 min, returned to 40% B by 4 min.
- Injection volume: 10 µL.
- Mass Spectrometry
- SCIEX Triple Quad™ 5500+ – QTRAP® Ready in positive electrospray ionization.
- MRM transitions: m/z 431.3→323.2 and 431.3→147.1 for budesonide epimers; 439.3→323.2 for the internal standard.
- Source settings: curtain gas 30, temperature 600 °C, GS1 = 60, GS2 = 80, ion voltage 3000 V, dwell time 75 ms.
- Data Processing
- Acquisition in Analyst® 1.7.1; quantitation via SCIEX OS-Q with 1/x²-weighted linear regression.
Main Results and Discussion
- Sensitivity and Linearity
- LLOQ of 2 pg/mL achieved without significant matrix interference.
- Calibration curve linear from 2 to 1024 pg/mL (r = 0.99).
- Precision and Accuracy
- %CV ranged from 1.9% to 12.2%; accuracy between 95% and 106% across the calibration range.
- Robustness
- SPE sample cleanup effectively minimized matrix effects.
- Fast run time (<4 min) and stable performance with Turbo V™ source.
Benefits and Practical Applications
- High sensitivity supports low-level pharmacokinetic profiling.
- Wide dynamic range accommodates both therapeutic and trace-level measurements.
- Rapid sample preparation and chromatographic run increase laboratory throughput.
Future Trends and Opportunities
- Adoption of high-resolution mass spectrometry for enhanced metabolite confirmation.
- Automation of SPE workflows to improve consistency and throughput.
- Expansion toward multiplexed assays for respiratory biomarkers.
Conclusion
This validated LC-MS/MS method delivers selective, sensitive, and high-throughput quantitation of budesonide in human plasma, fulfilling stringent bioanalytical requirements for pharmacokinetic and clinical studies.
Reference
- Collins BA, Dima MA, Ivosev G, Zhong F (2013) A high dynamic range pulse counting detection system for mass spectrometry. Rapid Commun Mass Spectrom. 28(2):209–216.