TO-15 65 Component Mix on Rxi®-1ms (30 m)

Applications |  | RestekInstrumentation
GC/MSD, Thermal desorption, GC/SQ, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies, Restek, Nutech

Importance of the Topic


Volatile organic compounds are critical environmental and industrial analytes. Reliable detection and quantification of a broad mix of VOCs at low concentrations support regulatory compliance, indoor air quality assessment, and process monitoring.

Objectives and Study Overview


This work evaluates the performance of the Rxi-1ms GC column combined with cryogenic preconcentration for the separation and identification of sixty‐five target VOCs. The aim is to demonstrate rapid, high-resolution analysis suitable for TO-15 applications.

Methodology and Instrumentation


The following configuration was used:
  • Column: Rxi-1ms, 30 m × 0.32 mm ID, 1 µm film thickness
  • Preconcentrator: Nutech 8900DS cryofocusing at –160 °C
  • Trap materials: Glass beads (Trap 1) and Tenax (Trap 2)
  • GC Conditions: Oven 35 °C (1 min), ramp to 230 °C at 11 °C/min; helium carrier at 2.0 mL/min
  • Injection: Direct, 250 cc sample at 10 ppbv, desorb at 190 °C for 30 s
  • MS Detection: HP 5973 quadrupole, EI mode, scan 35–250 amu at 3.32 scans/s

Main Results and Discussion


All sixty-five compounds eluted between 1.37 and 12.38 minutes with baseline separation for the majority of analytes. Early eluters such as propylene and Freon® 12 are resolved within 1.4 minutes, while heavier aromatics and chlorinated species appear after 6 minutes. Two coelutions of minor concern share an ion but are resolved by unique qualifier ions. The fast ramp and optimized film thickness deliver a total run time under 13 minutes without sacrificing peak shape or resolution.

Benefits and Practical Applications


Key advantages include:
  • High throughput analysis for regulatory VOC panels
  • Robust separation across a wide volatility range
  • Enhanced sensitivity from cryogenic preconcentration
  • Compatibility with standard TO-15 sampling protocols

Future Trends and Opportunities


Advances may involve integrating two-dimensional GC or high-resolution MS for even greater compound discrimination. Development of automated sampling systems and miniaturized preconcentrators will further streamline field and laboratory workflows.

Conclusion


The Rxi-1ms column coupled with cryofocusing and quadrupole MS provides a rapid, sensitive, and reproducible method for comprehensive VOC analysis. The approach meets TO-15 requirements and supports diverse environmental and industrial applications.

References


  • Restek Corporation Application Note 34436: TO-15 65 Component Mix on Rxi-1ms, 30 m column

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

TO-15 65 Component Mix on Rxi®-624Sil MS (30 m)
TO-15 65 Component Mix on Rxi®-624Sil MS (30 m)
|Agilent Technologies|Applications
TO-15 65 Component Mix on Rxi®-5Sil MS (30 m, 1.5 mL/min)
TO-15 65 Component Mix on Rtx®-VMS (30 m, 2.0 mL/min)
TO-15 65 Component Mix on Rtx®-VMS (30 m, 1.2 mL/min)